Directional ablation in radiofrequency ablation using a multi -tine electrode functioning in multipolar mode: An in-silico study using a finite set of states

被引:7
|
作者
Dhiman, Manoj [1 ]
Kumawat, Aakash Kumar [1 ]
Repaka, Ramjee [1 ]
机构
[1] Indian Inst Technol Ropar, Dept Mech Engn, Rupnagar, Punjab, India
关键词
Radiofrequency ablation; Directional ablation; Temperature-controlled RFA; Multi-tine electrode; Multipolar mode RFA; THERMAL-ABLATION; HEPATOCELLULAR-CARCINOMA; ARTIFICIAL ASCITES; LIVER-TUMORS; TISSUE; COAGULATION; PERFUSION; HYDRODISSECTION; COMPLICATIONS; VALIDATION;
D O I
10.1016/j.compbiomed.2020.104007
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: To analyse the feasibility of directional ablation using a multi-tine electrode. Methods: A multi-tine electrode capable of operating in multipolar mode has been used to study the directional ablation. In addition to the basic design, similar to commercially available FDA approved multi-tine electrode, tines have been insulated from each other inside the probe base and tip using a thin insulating material of thickness 0.25 mm. A cylindrical single-compartment model of size 6 cm x 6 cm has been used to model normal liver tissue. The temperature-controlled radiofrequency ablation has been employed to maintain the tine-tips at different temperatures. Electro-thermal simulations have been performed by using a commercial multi-physics software package based on finite element methods. To make this study feasible a new approach to predict the ablations have been proposed and used in this study. Results: Asymmetric ablation zone with up to 5 mm difference in ablation boundary between the intended and non-intended direction has been observed along the transverse direction. Reduction in ablation up to 5 mm along the axial direction in comparison to the monopolar mode has also been observed. Conclusion: Multi-tine electrode modified to operate in multipolar mode can create directional ablations of different shapes and can be used to target position and shape specific tumours.
引用
收藏
页数:15
相关论文
共 46 条
  • [1] Study on Radiofrequency Ablation of Lung and Kidney Tumors With and Without Nanoparticles Using Multi-Tined Electrode
    Ahammed, Md Ezaz
    Peravali, Mrunal Swaroop
    Naik, Santosh S.
    Yadav, Ajay Kumar
    Laxminidhi, T.
    Journal of Engineering and Science in Medical Diagnostics and Therapy, 2024, 7 (03)
  • [2] Lesion Characteristics Using a Novel Multi-Electrode Radiofrequency Balloon Ablation Catheter
    Lee, Byron K.
    Al-Ahmad, Amin
    Rippy, Marian
    Reddy, Vivek
    Narayan, Girish
    Kuck, Karl-Heinz
    Crozier, Ian
    Daly, Matt
    Melton, Iain
    Walters, Tomas E.
    Lee, Randall J.
    Badhwar, Nitish
    CIRCULATION, 2016, 134
  • [3] Long ablation cycles result in increased depth of multi-electrode ablation lesions using pulsed radiofrequency energy
    McRury, I
    Schlichting, A
    Wilson, C
    Diamond, S
    XXI CONGRESS OF THE EUROPEAN SOCIETY OF CARDIOLOGY, 1999, : 851 - 855
  • [4] Optimizing Electrode Placement Using Finite-Element Models in Radiofrequency Ablation Treatment Planning
    Chen, Chun-Cheng R.
    Miga, Michael I.
    Galloway, Robert L., Jr.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2009, 56 (02) : 237 - 245
  • [5] Radiofrequency ablation of chondroblastoma using a multi-tined expandable electrode system: initial results
    Bernhard Tins
    Victor Cassar-Pullicino
    Iain McCall
    Paul Cool
    David Williams
    David Mangham
    European Radiology, 2006, 16 : 804 - 810
  • [6] Radiofrequency ablation of chondroblastoma using a multi-tined expandable electrode system: initial results
    Tins, B
    Cassar-Pullicino, V
    McCall, I
    Cool, P
    Williams, D
    Mangham, D
    EUROPEAN RADIOLOGY, 2006, 16 (04) : 804 - 810
  • [7] Multipolar radiofrequency ablation using 4-6 applicators simultaneously: A study in the ex vivo bovine liver
    Stoffner, Rudolf
    Kremser, Christian
    Schullian, Peter
    Haidu, Marion
    Widmann, Gerlig
    Bale, Reto J.
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (10) : 2568 - 2575
  • [8] Radiofrequency ablation using injectable cooled electrode: the effects of lidocaine injection in ex vivo study
    Park, Sun-Young
    Chung, Hye Won
    Baek, Jung Hwan
    Lee, Jeong Hyun
    Lee, Min Hee
    Lee, Joon Woo
    ACTA RADIOLOGICA, 2020, 61 (02) : 219 - 226
  • [9] Transluminal radiofrequency ablation using new stent-type electrode: An experimental pilot study
    Rhim, H
    Song, HY
    Shin, JH
    Seo, TS
    Koh, BH
    Cho, OK
    RADIOLOGY, 2002, 225 : 637 - 637
  • [10] Radiofrequency catheter ablation of frequent premature ventricular contractions using ARRAY multi-electrode balloon catheter
    Dubner, Sergio
    Hadid, Claudio
    Azocar, Damian
    Labadet, Carlos
    Valsecchi, Cecilia
    Dominguez, Agustin
    CARDIOLOGY JOURNAL, 2016, 23 (01) : 17 - 22