共 50 条
Evaluation of Thin-slice Coronal Single-shot Turbo Spin-echo Diffusion-weighted Imaging of the Hand: A Comparison with Conventional Echo-planar Diffusion-weighted Imaging
被引:5
|作者:
Ogawa, Masaki
[1
]
Nakagawa, Motoo
[1
]
Arai, Nobuyuki
[2
]
Kan, Hirohito
[2
]
Ohba, Shota
[1
]
Shibata, Shunsuke
[1
]
Maki, Hiroyuki
[1
]
Shibamoto, Yuta
[1
]
机构:
[1] Nagoya City Univ, Dept Radiol, Grad Sch Med Sci, Nagoya, Aichi, Japan
[2] Nagoya City Univ Hosp, Dept Radiol, Nagoya, Aichi, Japan
关键词:
musculoskeletal lesion;
magnetic resonance imaging;
non-echo-planar imaging technique;
turbo spin-echo diffusion-weighted imaging;
EPI;
D O I:
10.2463/mrms.tn-2019-0115
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
We prospectively evaluated thin-slice coronal turbo spin-echo (TSE) diffusion-weighted imaging (DWI) covering a larger area with the recently-developed techniques on a 3T MRI scanner, compared with echoplanar imaging (EPI)-DWI in patients undergoing routine hand MRI. Visual score assessment and apparent diffusion coefficient (ADC) measurement were performed for patients with suspected hand tumors. TSE-DWI was superior to EPI-DWI, with less image distortion. The visual score and ADC comparison assessments proved that the image noise of TSE-DWI was acceptable.
引用
收藏
页码:382 / 388
页数:7
相关论文