A simple digital delay for nuclear physics experiments

被引:2
|
作者
Marques, J. G. [1 ]
Cruz, C. [2 ]
机构
[1] Univ Lisbon, Inst Super Tecn, C2TN, P-2695066 Bobadela Lrs, Portugal
[2] Univ Lisbon, Inst Super Tecn, LATR, P-2695066 Bobadela Lrs, Portugal
关键词
Perturbed Angular Correlations; Experimental nuclear physics; Digital delay; Emitter-Coupled Logic; TDPAC SPECTROMETER; RESOLUTION; SYSTEM;
D O I
10.1016/j.nima.2014.01.060
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A simple high precision digital delay for nuclear physics experiments was developed using fast ECL electronics. The circuit uses an oscillator synchronized with the signal to be delayed and a presettable counter It is capable of delaying a negative NlM signal by 2 is with a precision better than 50 ps. The circuit was developed for use in slow-fast coincidence units for Perturbed Angular Correlation spectrometers but it is not limited to this application. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:50 / 56
页数:7
相关论文