Intraoperative X-Ray Detection and MRI-Based Quantification of Brain Shift Effects Subsequent to Implantation of the First Electrode in Bilateral Implantation of Deep Brain Stimulation Electrodes

被引:31
|
作者
Hunsche, Stefan [1 ]
Sauner, Dieter [3 ]
Maarouf, Mohammad [1 ]
Poggenborg, Joerg [2 ]
Lackner, Klaus [2 ]
Sturm, Volker [1 ]
Treuer, Harald [1 ]
机构
[1] Univ Cologne, Dept Stereotact & Funct Neurosurg, DE-50937 Cologne, Germany
[2] Univ Cologne, Dept Radiol, DE-50937 Cologne, Germany
[3] Municipal Clin Duisburg, Dept Radiol & Neuroradiol, Duisburg, Germany
关键词
Stereotactic neurosurgery; Brain shift; X-ray; Intraoperative MRI; REGISTRATION; DEFORMATION;
D O I
10.1159/000235804
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Objective: After implantation of the first electrode in bilateral deep brain stimulation (DBS) lead implantation, brain shift effects in the target region and along the implantation trajectory of the second electrode are quantified with intra-operative magnetic resonance imaging (MRI). We investigated intra-operative X-ray imaging for its feasibility in indirect detection of brain shift. Methods: In 25 patients who underwent bilateral DBS lead implantation, X-ray and MRI were performed before and after implantation of the first electrode. Two parameters of brain shift were assessed with nonrigid free-form deformation field analysis of the MRI data: global brain shift along the anterior and posterior commissure (AC-PC) line and specific brain shift along the implantation trajectory of the second electrode. Pre- and intraoperative X-ray images were geometrically and intensity corrected for detection of significant signal changes through intracranial air accumulation during implantation of the first electrode. Results: After implantation of the first electrode, brain shift greater than 1 mm (maximum 1.3 mm) was observed at the AC and brain shift greater than 2 mm (maximum 2.5 mm) was observed along the planned implantation trajectory of the second electrode. In 1 patient, the implantation trajectory of the second electrode went through a sulcus after cortical brain shift. In 9 patients, intracranial air volume between 0.1 and 38.5 ml was observed with MRI after implantation of the first electrode. Significant X-ray absorption changes were induced by an intracranial air volume of greater than 8 ml. Conclusion: In bilateral DBS implantation, brain shift effects can cause misallocation of the second electrode with the risk of adverse or no stimulation effects as well as unnecessary cortical damage. A lack of X-ray signal changes caused by intracranial air invasion during DBS lead implantation indicates a lack of clinically relevant brain shift. Copyright (C) 2009 S. Karger AG, Basel
引用
收藏
页码:322 / 329
页数:8
相关论文
共 25 条
  • [21] 3D X-ray based visualization of directional deep brain stimulation lead orientation
    Egger, Karl
    Rau, Alexander
    Urbach, Horst
    Reisert, Marco
    Reinacher, Peter C.
    JOURNAL OF NEURORADIOLOGY, 2022, 49 (03) : 293 - 297
  • [22] Frameless x-ray-based lead re-implantation after partial hardware removal of deep brain stimulation system with preservation of intracerebral trajectories
    Vesna Malinova
    Dariusz J. Jaskólski
    Rafal Wójcik
    Dorothee Mielke
    Veit Rohde
    Acta Neurochirurgica, 2021, 163 : 1873 - 1878
  • [23] Frameless x-ray-based lead re-implantation after partial hardware removal of deep brain stimulation system with preservation of intracerebral trajectories
    Malinova, Vesna
    Jaskolski, Dariusz J.
    Wojcik, Rafal
    Mielke, Dorothee
    Rohde, Veit
    ACTA NEUROCHIRURGICA, 2021, 163 (07) : 1873 - 1878
  • [24] Review of the targeting accuracy of frameless and frame-based robot-assisted deep brain stimulation electrode implantation in pediatric patients using the Neurolocate module
    David, Raenette
    Scala, Maria Rosaria
    Ellenbogen, Jonathan
    JOURNAL OF NEUROSURGERY-PEDIATRICS, 2024, 33 (03) : 207 - 213
  • [25] Enhanced brain tumor detection and classification using a deep image recognition generative adversarial network (DIR-GAN): a comparative study on MRI, X-ray, and FigShare datasets
    S. Karpakam
    N. Kumareshan
    Neural Computing and Applications, 2025, 37 (14) : 8731 - 8758