Intraoperative X-Ray Detection and MRI-Based Quantification of Brain Shift Effects Subsequent to Implantation of the First Electrode in Bilateral Implantation of Deep Brain Stimulation Electrodes

被引:31
|
作者
Hunsche, Stefan [1 ]
Sauner, Dieter [3 ]
Maarouf, Mohammad [1 ]
Poggenborg, Joerg [2 ]
Lackner, Klaus [2 ]
Sturm, Volker [1 ]
Treuer, Harald [1 ]
机构
[1] Univ Cologne, Dept Stereotact & Funct Neurosurg, DE-50937 Cologne, Germany
[2] Univ Cologne, Dept Radiol, DE-50937 Cologne, Germany
[3] Municipal Clin Duisburg, Dept Radiol & Neuroradiol, Duisburg, Germany
关键词
Stereotactic neurosurgery; Brain shift; X-ray; Intraoperative MRI; REGISTRATION; DEFORMATION;
D O I
10.1159/000235804
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Objective: After implantation of the first electrode in bilateral deep brain stimulation (DBS) lead implantation, brain shift effects in the target region and along the implantation trajectory of the second electrode are quantified with intra-operative magnetic resonance imaging (MRI). We investigated intra-operative X-ray imaging for its feasibility in indirect detection of brain shift. Methods: In 25 patients who underwent bilateral DBS lead implantation, X-ray and MRI were performed before and after implantation of the first electrode. Two parameters of brain shift were assessed with nonrigid free-form deformation field analysis of the MRI data: global brain shift along the anterior and posterior commissure (AC-PC) line and specific brain shift along the implantation trajectory of the second electrode. Pre- and intraoperative X-ray images were geometrically and intensity corrected for detection of significant signal changes through intracranial air accumulation during implantation of the first electrode. Results: After implantation of the first electrode, brain shift greater than 1 mm (maximum 1.3 mm) was observed at the AC and brain shift greater than 2 mm (maximum 2.5 mm) was observed along the planned implantation trajectory of the second electrode. In 1 patient, the implantation trajectory of the second electrode went through a sulcus after cortical brain shift. In 9 patients, intracranial air volume between 0.1 and 38.5 ml was observed with MRI after implantation of the first electrode. Significant X-ray absorption changes were induced by an intracranial air volume of greater than 8 ml. Conclusion: In bilateral DBS implantation, brain shift effects can cause misallocation of the second electrode with the risk of adverse or no stimulation effects as well as unnecessary cortical damage. A lack of X-ray signal changes caused by intracranial air invasion during DBS lead implantation indicates a lack of clinically relevant brain shift. Copyright (C) 2009 S. Karger AG, Basel
引用
收藏
页码:322 / 329
页数:8
相关论文
共 25 条
  • [1] Brain shift: an error factor during implantation of deep brain stimulation electrodes
    Miyagi, Yasushi
    Shima, Fumio
    Sasaki, Tomio
    JOURNAL OF NEUROSURGERY, 2007, 107 (05) : 989 - 997
  • [2] Frameless Stereotaxis for Subthalamic Nucleus Deep Brain Stimulation: An Innovative Method for the Direct Visualization of Electrode Implantation by Intraoperative X-ray Control
    Mazzone, Paolo
    Stefani, Alessandro
    Viselli, Fabio
    Scarnati, Eugenio
    BRAIN SCIENCES, 2018, 8 (05):
  • [3] Simultaneous bilateral MER-guided stereotactic implantation of deep brain stimulation electrodes
    Fonoff, E. T.
    Contreras, W. O.
    Ghilardi, Lopez M.
    Cury, R. G.
    Martinez, R. C. R.
    Azevedo, A.
    Barbosa, E. R.
    Navarro, J.
    Teixeira, M. J.
    MOVEMENT DISORDERS, 2014, 29 : S439 - S440
  • [4] Early Deformation of Deep Brain Stimulation Electrodes Following Surgical Implantation: Intracranial, Brain, and Electrode Mechanics
    Chapelle, Frederic
    Manciet, Lucie
    Pereira, Bruno
    Sontheimer, Anna
    Coste, Jerome
    El Ouadih, Youssef
    Cimpeanu, Ruxandra
    Gouot, Dimitri
    Lapusta, Yuri
    Claise, Beatrice
    Sautou, Valerie
    Bouattour, Yassine
    Marques, Ana
    Wohrer, Adrien
    Lemaire, Jean-Jacques
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2021, 9
  • [5] Deep Brain Stimulation of Heschl Gyrus: Implantation Technique, Intraoperative Localization, and Effects of Stimulation
    Donovan, Chris
    Sweet, Jennifer
    Eccher, Matthew
    Megerian, Cliff
    Semaan, Maroun
    Murray, Gail
    Miller, Jonathan
    NEUROSURGERY, 2015, 77 (06) : 940 - 947
  • [6] Brain Shift and Pneumocephalus Assessment During Frame-Based Deep Brain Stimulation Implantation With Intraoperative Magnetic Resonance Imaging
    Matias, Caio M.
    Frizon, Leonardo A.
    Asfahan, Fadi
    Uribe, Juan D.
    Machado, Andre G.
    OPERATIVE NEUROSURGERY, 2018, 14 (06) : 668 - 674
  • [7] Direction and Predictive Factors for the Shift of Brain Structure During Deep Brain Stimulation Electrode Implantation for Advanced Parkinson's Disease
    Obuchi, Toshiki
    Katayama, Yoichi
    Kobayashi, Kazutaka
    Oshima, Hideki
    Fukaya, Chikashi
    Yamamoto, Takamitsu
    NEUROMODULATION, 2008, 11 (04): : 302 - 310
  • [8] Biplanar X-Ray Methods for Stereotactic Intraoperative Localization in Deep Brain Stimulation Surgery
    Sedrak, Mark
    Sabelman, Eric
    Pezeshkian, Patrick
    Duncan, John
    Bernstein, Ivan
    Bruce, Diana
    Tse, Victor
    Khandhar, Suketu
    Call, Elena
    Heit, Gary
    Alaminos-Bouza, Armando
    OPERATIVE NEUROSURGERY, 2020, 19 (03) : 302 - 312
  • [9] Lead migration of deep brain stimulation electrodes - Role of the plain skull X-ray
    Samuel, M
    Bain, PG
    Jarosz, J
    Aziz, TZ
    Brand, G
    Selway, R
    MOVEMENT DISORDERS, 2005, 20 : S159 - S159
  • [10] Asymptomatic transient MRI signal changes after unilateral deep brain stimulation electrode implantation for movement disorder
    Ryu, SI
    Romanelli, P
    Heit, G
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2004, 82 (2-3) : 65 - 69