Twenty-Seven Years Experience With Transvenous Pacemaker Implantation in Children Weighing <10 kg

被引:26
|
作者
Konta, Laura [1 ]
Chubb, Mark Henry [1 ,2 ]
Bostock, Julian [1 ]
Rogers, Jan [1 ]
Rosenthal, Eric [1 ]
机构
[1] Evelina London Childrens Hosp, Dept Paediat Cardiol, London SE1 7EH, England
[2] Kings Coll London, Div Imaging Sci & Biomed Engn, London WC2R 2LS, England
来源
关键词
infant; defibrillators; implantable; subclavian vein; biological pacemakers; newborn; CONGENITAL HEART; VENOUS OBSTRUCTION; PEDIATRIC-PATIENTS; SINGLE-CHAMBER; PACING LEAD; INFANTS; CARDIOMYOPATHY; COMPLICATION; NEWBORNS; BLOCK;
D O I
10.1161/CIRCEP.115.003422
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background- Epicardial pacemaker implantation is the favored approach in children weighing < 10 kg in many units. The high incidence of premature failure and fractures with earlier epicardial leads led our unit to undertake transvenous pacemaker implantation in neonates and infants from 1987. To date there have been no long-term follow-up reports of what is for many a controversial strategy. Methods and Results- Between 1987 and 2003, 37 neonates and infants-median age 6.7 months (1 day to 3 years) and median weight 4.6 kg (2.7-10 kg)-had a permanent transvenous pacing system implanted. Pacing leads were placed into the right ventricular apex/outflow tract through a subclavian vein puncture with a redundant loop in the atrium. Three patients were lost to follow-up, 4 patients died from complications of cardiac surgery, and 2 patients had their system removed. At long-term follow-up in 28 patients at a median of 17.2 (range, 11.2-27.4) years, 10 patients have a single chamber ventricular pacemaker, 14 a dual chamber pacemaker, 3 a biventricular pacemaker, and 1 has a single chamber implantable cardioverter defibrillator. Subclavian vein patency was assessed in 26 patients. The overall subclavian vein occlusion rate was 10 of 13 (77%) < 5 kg and 2 of 13 (15%) > 5 kg during long-term follow-up. After a median of 14.3 (range, 13.4-17.6) years of pacing, 7 patients continue with their original lead. Conclusions- Transvenous pacing in infants < 10 kg results in encouraging short- and long-term clinical outcomes. Subclavian vein occlusion remains an important complication, occurring predominantly in those weighing < 5 kg.
引用
收藏
页数:8
相关论文
共 33 条
  • [31] Low incidence of complications after cephalic vein cutdown for pacemaker lead implantation in children weighing less than 10 kilograms: A single-center experience with long-term follow-up
    Kircanski, Bratislav
    Vasic, Dragan
    Savic, Dragutin
    Stojanov, Petar
    HEART RHYTHM, 2015, 12 (08) : 1820 - 1826
  • [32] To Study the Feasibility of Transcatheter Closure of Large Hypertensive PDA (Type B And C) & PDAs with Unusual Morphology Using the LIFETECH Konar-MF™ Device in Children Weighing &lt; 10 kg
    Ramaswamy, Rajesh
    Chandrasekharan, Ramyashri
    PEDIATRIC CARDIOLOGY, 2022, 43 (08) : 2028 - 2028
  • [33] Testicular descent and fixation through a scrotal stria incision for the treatment of palpable cryptorchidism in children aged &lt;3 years: 10-year experience with 1034 cases
    Chen, Liu
    Huang, Wen-Hua
    Wang, Yun-Jin
    Zhang, Qi-Liang
    Zhou, Chao-Ming
    Cui, Xu
    Zhang, Jian-Qin