In vivo quality assurance of volumetric modulated arc therapy for ano-rectal cancer with thermoluminescent dosimetry and image-guidance

被引:9
|
作者
Dipasquale, Giovanna [1 ]
Nouet, Philippe [1 ]
Rouzaud, Michel [1 ]
Dubouloz, Angele [1 ]
Miralbell, Raymond [1 ]
Zilli, Thomas [1 ]
机构
[1] Univ Hosp Geneva, Dept Radiat Oncol, CH-1211 Geneva 14, Switzerland
关键词
Anal cancer; Rectal cancer; VMAT; In vivo dosimetry; CBCT; RADIATION-THERAPY; IMRT; RADIOTHERAPY; VALIDATION; IMPLEMENTATION; RAPIDARC; GR-200; PHOTON;
D O I
10.1016/j.radonc.2014.04.014
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Objective: To assess in vivo dose distribution using cone-beam computed tomography scans (CBCTs) and thermoluminescent dosimeters (TLDs) in patients with anal or rectal cancer treated with volumetric modulated arc therapy (VMAT). Methods: Intracavitary (IC) in vivo dosimetry (IVD) was performed in 11 patients using adapted endorectal probes containing TLDs, with extra measurements at the perianal skin (PS) for anal margin tumors. Measured doses were compared to calculated ones obtained from image fusion of CBCT with CT treatments plans. Results: A total of 55 IC and 6 PS measurements were analyzed. IC TLD median planned and measured does were 1.81 Gy (range, 0.25-2.02 Gy) and 1.82 Gy (range, 0.19-2.12 Gy), respectively. In comparison to the planned doses all IC TLD dose measurements differed by a median dose of 0.02 Gy (range, -0.11/+0.19 Gy, p = 0.102) (median difference of 1.1%, range -6.1%/+10.6%). Overall, 95% of IC measurements were within +/- 7.7% of the expected percentage doses and only 1 value was above +10%. For PS measurements, only one was not within +/- 7.7% of expected values (i.e., -8.9%). Conclusions: Image guidance using CBCT for IVD with TLDs is helpful to validate the delivered doses in patients treated with VMAT for ano-rectal tumors. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
下载
收藏
页码:406 / 411
页数:6
相关论文
共 50 条
  • [31] Patient Specific Quality Assurance for Volumetric Modulated Arc Therapy (RapidArc) Using COMPASS 3D Dosimetry System
    Arun, G.
    Subramanian, V. Shanmuga
    Swamy, S. Thirumalai
    Kathirvel, M.
    Ramalingam, K.
    Kala, S.
    Kumar, S. Ashok
    Karthikeyan, K.
    Anantharaman, A.
    Kumar, N. Jagdish
    Arivarasan, I.
    Babaiah, M.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [32] Verification of patient specific quality assurance system for volumetric modulated arc therapy (VMAT)
    Amoabeng, Kwame Anokye
    Marthinsen, Anne Beate Langeland
    Hasford, Francis
    Tagoe, Samuel Nii Adu
    Anaafi, Evelyn
    HEALTH AND TECHNOLOGY, 2022, 12 (04) : 779 - 786
  • [33] A Systematic Approach to Patient-Specific Quality Assurance for Volumetric Modulated Arc Therapy
    SONG Yu-lin
    Ceferino Obcemea
    Boris Mueller
    Borys Mychalczak
    Chinese Journal of Biomedical Engineering, 2015, 24 (02) : 58 - 65
  • [34] Daily Based Quality Assurance of Volumetric Modulated Arc Therapy for the Full Session of Treatment
    Mohammad Mahfujur Rahman
    Chan Hyeong Kim
    Seonghoon Kim
    Journal of the Korean Physical Society, 2018, 73 : 990 - 1000
  • [35] Quality assurance of volumetric modulated arc therapy: Evaluation and comparison of different dosimetric systems
    Masi, L.
    Casamassima, F.
    Doro, R.
    Francescon, P.
    MEDICAL PHYSICS, 2011, 38 (02) : 612 - 621
  • [36] Daily Based Quality Assurance of Volumetric Modulated Arc Therapy for the Full Session of Treatment
    Rahman, Mohammad Mahfujur
    Kim, Chan Hyeong
    Kim, Seonghoon
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2018, 73 (07) : 990 - 1000
  • [37] Verification of patient specific quality assurance system for volumetric modulated arc therapy (VMAT)
    Kwame Anokye Amoabeng
    Anne Beate Langeland Marthinsen
    Francis Hasford
    Samuel Nii Adu Tagoe
    Evelyn Anaafi
    Health and Technology, 2022, 12 : 779 - 786
  • [38] Evaluation of Electronic Portal Imaging Device in the Quality Assurance of Volumetric Modulated Arc Therapy
    Liu, Z.
    Chan, P.
    Li, S.
    Micaily, B.
    Miyamoto, C.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [39] Fully automatic volumetric modulated arc therapy plan generation for rectal cancer
    Song, Ying
    Wang, Qiang
    Jiang, Xiaoqin
    Liu, Sha
    Zhang, Yingjie
    Bai, Sen
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 (03) : 531 - 536
  • [40] RECTAL TOXICITY IN MODERATE HYPOFRACTIONATION PROSTATE CANCER WITH VOLUMETRIC MODULATED ARC THERAPY
    Ricchetti, Francesco
    Fersino, Sergio
    Mazzola, Rosario
    Levra, Niccolo Giaj
    Fiorentino, Alba
    Naccarato, Stefania
    Sicignano, Gianluisa
    Ruggieri, Ruggero
    Cavalleri, Stefano
    Alongi, Filippo
    ANTICANCER RESEARCH, 2015, 35 (06) : 3626 - 3627