Ideation and assessment of a nitric oxide delivery system for spontaneously breathing subjects

被引:13
|
作者
Gianni, Stefano [1 ,2 ]
Morais, Caio C. A. [1 ,2 ]
Larson, Grant [1 ,2 ]
Pinciroli, Riccardo [1 ,2 ]
Carroll, Ryan [2 ,3 ]
Yu, Binglan [1 ,2 ]
Zapol, Warren M. [1 ,2 ]
Berra, Lorenzo [1 ,2 ]
机构
[1] Massachusetts Gen Hosp, Dept Anaesthesia Crit Care & Pain Med, Boston, MA 02114 USA
[2] Harvard Med Sch, Boston, MA 02115 USA
[3] Massachusetts Gen Hosp, Dept Pediat, Boston, MA 02114 USA
来源
关键词
Nitric oxide; Nitrogen dioxide; Delivery system; METHEMOGLOBIN; OXIMETRY;
D O I
10.1016/j.niox.2020.08.004
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Background: There is an increasing interest in safely delivering high dose of inhaled nitric oxide (NO) as an antimicrobial and antiviral therapeutics for spontaneously breathing patients. A novel NO delivery system is described. Methods: We developed a gas delivery system that utilizes standard respiratory circuit connectors, a reservoir bag, and a scavenging chamber containing calcium hydroxide. The performance of the system was tested using a mechanical lung, assessing the NO concentration delivered at varying inspiratory flows. Safety was assessed in vitro and in vivo by measuring nitrogen dioxide (NO2) levels in the delivered NO gas. Lastly, we measured the inspired and expired NO and NO2 of this system in 5 healthy subjects during a 15-min administration of high dose NO (160 parts-per-million, ppm) using our delivery system. Results: The system demonstrated stable delivery of prescribed NO levels at various inspiratory flow rates (0-50 L/min). The reservoir bag and a high flow of entering air minimized the oscillation of NO concentrations during inspiration on average 4.6 ppm for each 10 L/min increment in lung inspiratory flow. The calcium hydroxide scavenger reduced the inhaled NO2 concentration on average 0.9 ppm (95% CI -1.58, 0.22; p = .01). We performed 49 NO administrations of 160 ppm in 5 subjects. The average concentration of inspired NO was 164.8 +/- 10.74 ppm, with inspired NO2 levels of 0.7 +/- 0.13 ppm. The subjects did not experience any adverse events; transcutaneous methemoglobin concentrations increased from 1.05 +/- 0.58 to 2.26 +/- 0.47%. Conclusions: The system we developed to administer high-dose NO for inhalation is easy to build, reliable, was well tolerated in healthy subjects.
引用
收藏
页码:29 / 35
页数:7
相关论文
共 50 条
  • [21] Novel Delivery System for the Bioregulatory Agent Nitric Oxide
    Liu, Harvey A.
    Balkus, Kenneth J., Jr.
    CHEMISTRY OF MATERIALS, 2009, 21 (21) : 5032 - 5041
  • [22] Nitric oxide delivery system for cell culture studies
    Wang, C
    Deen, WM
    ANNALS OF BIOMEDICAL ENGINEERING, 2003, 31 (01) : 65 - 79
  • [23] Nitric Oxide Delivery System for Cell Culture Studies
    Chen Wang
    William M. Deen
    Annals of Biomedical Engineering, 2003, 31 : 65 - 79
  • [24] NOVEL SYSTEM DESIGN FOR DELIVERY OF GASEOUS NITRIC OXIDE
    Gardner, K.
    Drenkard, E.
    Yonker, L. M.
    Weigand, S.
    Hurley, B.
    Carroll, R.
    PEDIATRIC PULMONOLOGY, 2020, 55 : S183 - S183
  • [25] EVALUATION OF A UNIVERSAL NITRIC-OXIDE DELIVERY SYSTEM USING PREDILUTION OF NITRIC-OXIDE
    FOUBERT, L
    MAREELS, K
    LATIMER, R
    HERREGODS, L
    DENBLAUWEN, N
    ROLLY, G
    BRITISH JOURNAL OF ANAESTHESIA, 1995, 74 : 1 - 1
  • [26] Aerosol delivery to spontaneously breathing neonates: Spacer or nebulizer?
    Uhlig, T
    Eber, E
    Devadason, SG
    Pemberton, P
    Badawi, N
    LeSouef, PN
    Wildhaber, JH
    PEDIATRIC ASTHMA ALLERGY & IMMUNOLOGY, 1997, 11 (02) : 111 - 117
  • [27] Aerosol Delivery To Spontaneously Breathing Pediatric Patients With A Tracheostomy
    Willis, L. D.
    Berlinski, A.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2010, 181
  • [28] Increased levels of exhaled nitric oxide during nasal and oral breathing in subjects with seasonal rhinitis
    Martin, U
    Bryden, K
    Devoy, M
    Howarth, P
    JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1996, 97 (03) : 768 - 772
  • [29] Assessment of Volume Status in Spontaneously Breathing Patients
    Ibrahim, Ahmad
    ElGamal, Mohamed
    Eid, Anna
    Shelley, Kirk
    Alian, Aymen
    ANESTHESIA AND ANALGESIA, 2020, 130
  • [30] ASSESSMENT OF VENTILATION IN THE SPONTANEOUSLY BREATHING ANESTHETIZED PATIENT
    OCONNOR, SA
    WILKINSON, DJ
    CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1986, 7 (02): : 199 - 200