Distal Radius Volar Locking Plate Design and Associated Vulnerability of the Flexor Pollicis Longus

被引:49
|
作者
Limthongthang, Roongsak [1 ]
Bachoura, Abdo
Jacoby, Sidney M.
Osterman, A. Lee
机构
[1] Mahidol Univ, Dept Orthopaed Surg, Fac Med, Siriraj Hosp, Bangkok 10700, Thailand
来源
关键词
Distal radius fracture; flexor tendon rupture; plate prominence; volar plating; watershed line; TENDON RUPTURES; FRACTURE; FIXATION; PLACEMENT;
D O I
10.1016/j.jhsa.2014.01.038
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Flexor pollicis longus (FPL) tendon rupture is a well-documented complication related to the use of distal radius volar locking plates (VLPs). The final common pathway of flexor tendon rupture appears to involve implants prominent at the watershed line. We hypothesized that significant differences in VLP prominence exist between various plate designs. Methods Ten fresh frozen specimens were dissected to identify the path of the FPL in relationship to the distal radius at the watershed line. Five VLP designs were fixed to each specimen based on their anatomic fit, and slid distally until the distal edge of the plate reached the watershed line. The position of each fixed plate was evaluated by fluoroscopy. We used a 3-dimensional laser scanner to create computer models. The total surface area of plate prominence volar to the watershed line and the prominent area beneath the FPL were measured in the axial plane using computer software. Results At the watershed line, the FPL was located at 54% of the maximal width of the radius, as measured from its volar-ulnar comer. There were no significant differences in the location of plate fixation on lateral view radiographs according to the classification of Soong et al. The mean total surface area of plate prominence was 36 mm(2). The mean prominent area beneath the FPL was 10 mm(2). Significant differences in plate prominence were noted for various designs. Conclusions Despite optimal plate placement, various VLP designs were observed to have prominent profiles volar to the watershed line, to varying extents. Copyright (C) 2014 by the American Society for Surgery of the Hand. All rights reserved.
引用
收藏
页码:852 / 860
页数:9
相关论文
共 50 条
  • [1] Flexor pollicis longus tendon rupture by sandwiched underlying volar locking plate and distal radius
    Takuya Uemura
    Tadashi Okano
    Ema Onode
    Takuya Yokoi
    Kosuke Shintani
    Mitsuhiro Okada
    Hiroaki Nakamura
    [J]. Journal of Medical Ultrasonics, 2018, 45 : 647 - 651
  • [2] Flexor pollicis longus tendon rupture by sandwiched underlying volar locking plate and distal radius
    Uemura, Takuya
    Okano, Tadashi
    Onode, Ema
    Yokoi, Takuya
    Shintani, Kosuke
    Okada, Mitsuhiro
    Nakamura, Hiroaki
    [J]. JOURNAL OF MEDICAL ULTRASONICS, 2018, 45 (04) : 647 - 651
  • [3] Positioning of a Volar Locking Plate with a Central Flexor Pollicis Longus Tendon Notch in Distal Radius Fractures
    Kaiser, Peter
    Gruber, Hannes
    Loth, Fanny
    Schmidle, Gernot
    Arora, Rohit
    Gabl, Markus
    [J]. JOURNAL OF WRIST SURGERY, 2019, 8 (06) : 482 - 488
  • [4] Rupture of Flexor Pollicis Longus Tendon: A Complication of Volar Locking Plating of the Distal Radius
    Rajeev, Aysha Sethunathan
    Sreverthana, Shanaka
    Harrison, John
    [J]. EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2010, 36 (04) : 385 - 387
  • [5] Rupture of Flexor Pollicis Longus Tendon: A Complication of Volar Locking Plating of the Distal Radius
    Aysha Sethunathan Rajeev
    Shanaka Sreverthana
    John Harrison
    [J]. European Journal of Trauma and Emergency Surgery, 2010, 36 : 385 - 387
  • [6] Flexor Pollicis Longus Dysfunction After Volar Plate Fixation of Distal Radius Fractures
    Chilelli, Brian J.
    Patel, Ronak M.
    Kalainov, David M.
    Peng, Jie
    Zhang, Li-Qun
    [J]. JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2013, 38A (09): : 1691 - 1697
  • [7] PREVENTION OF FLEXOR POLLICIS LONGUS TENDON RUPTURE AFTER VOLAR PLATE FIXATION OF DISTAL RADIUS FRACTURES
    Tada, Kaoru
    Ikeda, Kazuo
    Shigemoto, Kenji
    Suganuma, Seigo
    Tsuchiya, Hiroyuki
    [J]. JOURNAL OF HAND SURGERY-ASIAN-PACIFIC VOLUME, 2011, 16 (03): : 271 - 275
  • [8] Evaluation of Flexor Pollicis Longus Tendon Rupture after Treatment of Distal Radius Fracture with the Volar Plate
    Yamak, Kamil
    Karahan, Huseyin Gokhan
    Karatan, Berrak
    Kayali, Cemil
    Altay, Taskin
    [J]. JOURNAL OF WRIST SURGERY, 2020, 9 (03) : 219 - 224
  • [9] Ultrasonic Evaluation of the Flexor Pollicis Longus Tendon Following Volar Plate Fixation for Distal Radius Fractures
    Kara, Adnan
    Celik, Haluk
    Bankaoglu, Mujdat
    Oc, Yunus
    Bulbul, Murat
    Sugun, Tahir Sadik
    [J]. JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2016, 41 (03): : 374 - 380
  • [10] Effect of Distal Radius Volar Plate Position on Contact Pressure Between the Flexor Pollicis Longus Tendon and the Distal Plate Edge
    Tanaka, Yoshitaka
    Aoki, Mitsuhiro
    Izumi, Tomoki
    Fujimiya, Mineko
    Yamashita, Toshihiko
    Imai, Tomohito