Artifacts of vascular stents in 3D MR angiography: experimental studies.

被引:0
|
作者
Baum, F [1 ]
Vosshenrich, R [1 ]
Fischer, U [1 ]
Castillo, E [1 ]
Grabbe, E [1 ]
机构
[1] Georg August Univ Klinikum, Rontgendiagnost Abt 1, D-37075 Gottingen, Germany
关键词
MR angiography MR; contrast material MR; vascular grafts;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this phantom study was to evaluate the visualization of vascular stents by using contrast-enhanced (CE) 3D MR angiography (MRA). Methods: The measurements were performed on a Magnetom Vision operating at 1.5 T with 25 mT/m gradients by using a head coil; a 3D FISP sequence (TR/TE/FA 4.6 ms/1.8 ms/30 degrees) was used. A phantom was designed with a length of 20 cm and consisted of a plastic cylinder filled with hydrogel. Tubes were used to install 7 different stents (Strecker, Boston Scientific; Cragg, Mintec; Walt, Schneider; Memotherm, Angiomed; St-Come, Trigon-MTS; Sinus, Optimed: Palmaz, Johnson & Johnson) with different metal components. The tubes were perfused with a solution of water and Gd-DTPA. Four radiologists evaluated the image quality. The measurements of the endoluminal stent diameters were compared with the findings obtained by digital subtraction angiography. The signal intensities were measured before (inflow), within (stent) and after (outflow) the different stent types. Results: The endoluminal diameter of the Cragg stent and the strecker graft were easily detectable. The signal loss within the lumen of the Wall stent was approx. 10% in relation to the inflow and outflow measurements. The result was an only slightly reduced image quality. The other grafts (St-Come, Sinus, Palmaz, Memotherm) showed significant signal losses ranging From 65% to 96%. A correct diagnosis was not possible within these graft lumina. Conclusion: Contrast-enhanced MR angiography can be used to evaluate vascular stents. A prerequisite is the application of particular commercially available grafts.
引用
收藏
页码:278 / 281
页数:4
相关论文
共 50 条
  • [21] Dynamic 3D MR angiography of intra- and extracranial vascular malformations at 3T: A technical note
    Ziyeh, S
    Strecker, R
    Berlis, A
    Weber, J
    Klisch, J
    Mader, I
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2005, 26 (03) : 630 - 634
  • [22] Prospective navigator gating with a dual acceptance window technique to reduce respiratory motion artifacts in 3D MR coronary angiography
    Yiping P. Du
    The International Journal of Cardiovascular Imaging, 2003, 19 : 157 - 162
  • [23] Prospective navigator gating with a dual acceptance window technique to reduce respiratory motion artifacts in 3D MR coronary angiography
    Du, YPP
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2003, 19 (02): : 157 - 162
  • [24] Metallic artifacts of coronary and iliac arteries stents in mr angiography and contrast-enhanced CT
    Amano, Y
    Ishihara, M
    Hayashi, H
    Gemma, K
    Kawamata, H
    Amano, M
    Kumazaki, T
    CLINICAL IMAGING, 1999, 23 (02) : 85 - 89
  • [25] Gadolinium-enhanced 3D time-of-flight MR angiography - Experimental and clinical evaluation
    Yano, T
    Kodama, T
    Suzuki, Y
    Watanabe, K
    ACTA RADIOLOGICA, 1997, 38 (01) : 47 - 54
  • [26] Automatic 3D vascular tree construction in CT angiography
    Chen, ZK
    Molloi, S
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2003, 27 (06) : 469 - 479
  • [27] 3D image-based navigators for coronary MR angiography
    Addy, Nii Okai
    Ingle, R. Reeve
    Luo, Jieying
    Baron, Corey A.
    Yang, Phillip C.
    Hu, Bob S.
    Nishimura, Dwight G.
    MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (05) : 1874 - 1883
  • [28] EVALUATION OF COARCTATION REPAIR WITH GATED MR ANGIOGRAPHY AND 3D RECONSTRUCTION
    BANK, ER
    PARKS, WJ
    SHEPPARD, SK
    PETTIGREW, RI
    RADIOLOGY, 1992, 185 : 308 - 308
  • [29] Lymph mode imaging by ultrarapid 3D MR angiography.
    Georgi, M
    Gaa, J
    Teubner, J
    Bolte, R
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1999, 170 (02): : 218 - 221
  • [30] 3D MR coronary angiography: optimization of the technique and preliminary results
    Prakken, Niek Hendrik Jan
    Vonken, Evert-Jan P. A.
    Velthuis, Birgitta K.
    Doevendans, Pieter A. F. M.
    Cramer, Maarten-Jan M.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2006, 22 (3-4): : 477 - 487