Artifacts of vascular stents in 3D MR angiography: experimental studies.

被引:0
|
作者
Baum, F [1 ]
Vosshenrich, R [1 ]
Fischer, U [1 ]
Castillo, E [1 ]
Grabbe, E [1 ]
机构
[1] Georg August Univ Klinikum, Rontgendiagnost Abt 1, D-37075 Gottingen, Germany
关键词
MR angiography MR; contrast material MR; vascular grafts;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this phantom study was to evaluate the visualization of vascular stents by using contrast-enhanced (CE) 3D MR angiography (MRA). Methods: The measurements were performed on a Magnetom Vision operating at 1.5 T with 25 mT/m gradients by using a head coil; a 3D FISP sequence (TR/TE/FA 4.6 ms/1.8 ms/30 degrees) was used. A phantom was designed with a length of 20 cm and consisted of a plastic cylinder filled with hydrogel. Tubes were used to install 7 different stents (Strecker, Boston Scientific; Cragg, Mintec; Walt, Schneider; Memotherm, Angiomed; St-Come, Trigon-MTS; Sinus, Optimed: Palmaz, Johnson & Johnson) with different metal components. The tubes were perfused with a solution of water and Gd-DTPA. Four radiologists evaluated the image quality. The measurements of the endoluminal stent diameters were compared with the findings obtained by digital subtraction angiography. The signal intensities were measured before (inflow), within (stent) and after (outflow) the different stent types. Results: The endoluminal diameter of the Cragg stent and the strecker graft were easily detectable. The signal loss within the lumen of the Wall stent was approx. 10% in relation to the inflow and outflow measurements. The result was an only slightly reduced image quality. The other grafts (St-Come, Sinus, Palmaz, Memotherm) showed significant signal losses ranging From 65% to 96%. A correct diagnosis was not possible within these graft lumina. Conclusion: Contrast-enhanced MR angiography can be used to evaluate vascular stents. A prerequisite is the application of particular commercially available grafts.
引用
收藏
页码:278 / 281
页数:4
相关论文
共 50 条
  • [1] Artifacts in 3D rotational angiography. An experimental study
    Hagen, G
    Lindgren, PG
    Jangland, L
    Magnusson, P
    Magnusson, A
    ACTA RADIOLOGICA, 2005, 46 (01) : 32 - 36
  • [2] Fast reduction of undersampling artifacts in radial MR angiography with 3D total variation on graphics hardware
    Florian Knoll
    Markus Unger
    Clemens Diwoky
    Christian Clason
    Thomas Pock
    Rudolf Stollberger
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2010, 23 : 103 - 114
  • [3] Fast reduction of undersampling artifacts in radial MR angiography with 3D total variation on graphics hardware
    Knoll, Florian
    Unger, Markus
    Diwoky, Clemens
    Clason, Christian
    Pock, Thomas
    Stollberger, Rudolf
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2010, 23 (02) : 103 - 114
  • [4] Experimental animal vascular studies.
    Braun, H
    DERMATOLOGISCHE WOCHENSCHRIFT, 1942, 114 (1/26): : 1008 - 1012
  • [5] Delineation of simulated vascular stenosis with Gd-DTPA-enhanced 3D gradient echo MR angiography: An experimental study
    Mitsuzaki, K
    Yamashita, Y
    Onomichi, M
    Tsuchigame, T
    Takahashi, M
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2000, 24 (01) : 77 - 82
  • [6] Postprocessing techniques for 3D MR angiography: Comparison with conventional angiography
    Davis, CP
    Wildermuth, S
    Hany, TF
    Pfammatter, T
    Debatin, JF
    RADIOLOGY, 1996, 201 : 641 - 641
  • [7] 3D Contrast-enhanced MR angiography
    Zhang, HongLei
    Maki, Jeffrey H.
    Prince, Martin R.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (01) : 13 - 25
  • [8] 3D TOF turbo MR angiography for intracranial arteries: Phantom and clinical studies
    Shigematsu, Y
    Korogi, Y
    Hirai, T
    Okuda, T
    Sugahara, T
    Liang, LX
    Takahashi, M
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1999, 10 (06): : 939 - 944
  • [9] Intracranial vascular stenosis and occlusion: comparison of 3D time-of-flight and 3D phase-contrast MR angiography
    Oelerich, M
    Lentschig, MG
    Zunker, P
    Reimer, P
    Rummeny, EJ
    Schuierer, G
    NEURORADIOLOGY, 1998, 40 (09) : 567 - 573
  • [10] Intracranial vascular stenosis and occlusion: comparison of 3D time-of-flight and 3D phase-contrast MR angiography
    M. Oelerich
    M. G. Lentschig
    P. Zunker
    P. Reimer
    E. J. Rummeny
    G. Schuierer
    Neuroradiology, 1998, 40 : 567 - 573