A high gas pressure calibration system using a liquid-lubricated pressure balance

被引:4
|
作者
Iizumi, Hideaki [1 ]
Kajikawa, Hiroaki [1 ]
Kobata, Tokihiko [1 ]
机构
[1] AIST, Natl Metrol Inst Japan, Tsukuba, Ibaraki, Japan
关键词
Pressure standard; High gas pressure; Liquid-lubricated pressure balance; Calibration;
D O I
10.1016/j.measurement.2017.01.054
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
We developed a high gas pressure calibration system that uses a liquid-lubricated (LL) pressure balance as a standard for pressures up to 100 MPa. The LL pressure balance can generate gas or hydraulic pressure by changing the pressure medium. We first evaluate characteristics of the LL piston-cylinder (PC), such as its effective area and fall rate. The effective area of the LL PC is evaluated through comparison with a hydraulic pressure balance. When the LL PC uses a gas, the fall rate is affected by gas dissolution into liquid in the LL PC reservoir. The results of calibrating a pressure gauge under the developed system are consistent with those by an existing hydraulic pressure calibration system. (C) 2017 Elsevier Ltd. All rights reserved.
引用
收藏
页码:106 / 111
页数:6
相关论文
共 50 条
  • [31] High Frequency Liquid Sinusoidal Pressure Calibration Based on Laser Interferometry
    Li B.
    Yang J.
    Huang N.
    Huang R.
    Yin X.
    Shi Y.
    Gongcheng Kexue Yu Jishu/Advanced Engineering Sciences, 2020, 52 (05): : 216 - 222
  • [32] The pressure balance, the normal instrument for high pressure.
    Holborn, L
    ZEITSCHRIFT DES VEREINES DEUTSCHER INGENIEURE, 1923, 67 : 188 - 189
  • [33] Discussion of application of material balance for high-pressure gas reservoirs
    Walsh, M.P.
    SPE Journal, 1998, 3 (04): : 402 - 404
  • [34] Application of a general material balance for high-pressure gas reservoirs
    Fetkovich, Michael J.
    Reese, Dave E.
    Whitson, C.H.
    SPE Reprint Series, 1999, (52): : 48 - 58
  • [35] CFD simulation of gas–liquid flow in a high-pressure bubble column with a modified population balance model
    Bo Zhang
    Lingtong Kong
    Haibo Jin
    Guangxiang He
    Suohe Yang
    Xiaoyan Guo
    ChineseJournalofChemicalEngineering, 2018, 26 (06) : 1350 - 1358
  • [36] Dynamic modeling of liquid impulse pressure generator for calibration of pressure sensors
    Yang, Jun
    Fan, Shangchun
    Li, Bo
    Huang, Riheng
    Shi, Yusong
    Shi, Bo
    SENSORS AND ACTUATORS A-PHYSICAL, 2018, 279 : 120 - 131
  • [37] The pressure field in the gas-lubricated step slider bearing
    Penesis, I
    Shepherd, JJ
    Connell, HJ
    ANZIAM JOURNAL, 2004, 45 : 423 - 442
  • [38] The Influence of Hydraulic Fluids on Pressure Balance Calibration Results
    Giusca, Claudiu L.
    Severn, Ian
    Greenwood, John
    JOINT 21ST AIRAPT AND 45TH EHPRG INTERNATIONAL CONFERENCE ON HIGH PRESSURE SCIENCE AND TECHNOLOGY, 2008, 121
  • [39] HIGH-PRESSURE LIQUID CHROMATOGRAPHY IN COMPARISON TO GAS CHROMATOGRAPHY
    DEININGE.G
    ERDOL UND KOHLE ERDGAS PETROCHEMIE VEREINIGT MIT BRENNSTOFF-CHEMIE, 1971, 24 (11): : 677 - &
  • [40] Use of liquid chromatography packings in high pressure gas chromatography
    Shariff, SM
    Robson, MM
    Bartle, KD
    Myers, P
    Clifford, AA
    HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY, 1996, 19 (09): : 527 - 529