Depth of interaction resolution measurements for a high resolution PET detector using position sensitive avalanche photodiodes

被引:127
|
作者
Yang, Yongfeng
Dokhale, Purushottam A.
Silverman, Robert W.
Shah, Kanai S.
McClish, Mickel A.
Farrell, Richard
Entine, Gerald
Cherry, Simon R.
机构
[1] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[2] Radiat Monitoring Devices Inc, Watertown, MA 02172 USA
[3] Univ Calif Los Angeles, Dept Mol & Med Pharmacol, Los Angeles, CA 90095 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2006年 / 51卷 / 09期
关键词
D O I
10.1088/0031-9155/51/9/001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We explore dual-ended read out of LSO arrays with two position sensitive avalanche photodiodes (PSAPDs) as a high resolution, high efficiency depth-encoding detector for PET applications. Flood histograms, energy resolution and depth of interaction (DOI) resolution were measured for unpolished LSO arrays with individual crystal sizes of 1.0, 1.3 and 1.5 mm, and for a polished LSO array with 1.3 mm pixels. The thickness of the crystal arrays was 20 mm. Good flood histograms were obtained for all four arrays, and crystals in all four arrays can be clearly resolved. Although the amplitude of each PSAPD signal decreases as the interaction depth moves further from the PSAPD, the sum of the two PSAPD signals is essentially constant with irradiation depth for all four arrays. The energy resolutions were similar for all four arrays, ranging from 14.7% to 15.4%. A DOI resolution of 3-4 mm (including the width of the irradiation band which is similar to 2 mm) was obtained for all the unpolished arrays. The best DOI resolution was achieved with the unpolished 1mm array (average 3.5 mm). The DOI resolution for the 1.3 mm and 1.5 mm unpolished arrays was 3.7 and 4.0 mm respectively. For the polished array, the DOI resolution was only 16.5 mm. Summing the DOI profiles across all crystals for the 1 mm array only degraded the DOI resolution from 3.5 mm to 3.9 mm, indicating that it may not be necessary to calibrate the DOI response separately for each crystal within an array. The DOI response of individual crystals in the array confirms this finding. These results provide a detailed characterization of the DOI response of these PSAPD-based PET detectors which will be important in the design and calibration of a PET scanner making use of this detector approach.
引用
收藏
页码:2131 / 2142
页数:12
相关论文
共 50 条
  • [21] Array of avalanche photodiodes as a position-sensitive X-ray detector
    Kishimoto, S
    Yoda, Y
    Seto, M
    Kitao, S
    Kobayashi, Y
    Haruki, R
    Harami, T
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 513 (1-2): : 193 - 196
  • [22] HIGH-RESOLUTION VIBRATION MEASUREMENTS USING A POSITION-SENSITIVE PHOTOSENSOR
    GRUNWALD, G
    SCHOTT, H
    SCHMIDT, B
    SENSORS AND ACTUATORS A-PHYSICAL, 1991, 26 (1-3) : 467 - 472
  • [23] High-resolution detection of 100 keV electrons using avalanche photodiodes
    Ogasawara, K.
    Hirahara, M.
    Miyake, W.
    Kasahara, S.
    Takashima, I.
    Asamura, K.
    Saito, Y.
    Mukai, T.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 594 (01): : 50 - 55
  • [24] Performance measurements of a depth-encoding PET detector module based on position-sensitive avalanche photodiode read-out
    Dokhale, PA
    Silverman, RW
    Shah, KS
    Grazioso, R
    Farrell, R
    Glodo, J
    McClish, MA
    Entine, G
    Tran, VH
    Cherry, SR
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (18): : 4293 - 4304
  • [25] HIGH TIME RESOLUTION X-RAY MEASUREMENTS WITH AN AVALANCHE PHOTODIODE DETECTOR
    KISHIMOTO, S
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01): : 824 - 827
  • [26] Depth-of-interaction-encoding high-resolution PET detector with sub-nanosecond timing
    Burr, KC
    Ivan, A
    Castleberry, DE
    Thompson, RA
    LeBlanc, JW
    Shah, KS
    Farrell, R
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3880 - 3884
  • [27] Resolution improvement using tapered fiber for a high resolution PET block detector
    Yamamoto, S
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) : 1074 - 1077
  • [28] Prototype of high resolution PET using resistive electrode position sensitive CdTe detectors
    Kikuchi, Y.
    Ishii, K.
    Terakawa, A.
    Matsuyama, S.
    Yamazaki, H.
    Hatazawa, J.
    Kotani, K.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2669 - +
  • [29] Development of a High-Resolution and Depth-of-Interaction Capable Detector for Time-of-Flight PET
    Krishnamoorthy, Srilalan
    Wiener, Rony I.
    Kaul, Madhuri
    Panetta, Joseph
    Karp, Joel S.
    Surti, Suleman
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2457 - 2461
  • [30] Resolution improvement using tapered fiber for a high resolution PET block detector
    Yamamoto, S
    1997 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1 & 2, 1998, : 1442 - 1445