Large area mercuric iodide x-ray imager

被引:10
|
作者
Zentai, G [1 ]
Partain, L [1 ]
Pavlyuchkova, R [1 ]
Virshup, G [1 ]
Zuck, A [1 ]
Melekhov, L [1 ]
Dagan, O [1 ]
Vilensky, A [1 ]
Gilboa, H [1 ]
机构
[1] Varian Med Syst, Ginzton Technol Ctr, Mt View, CA 94043 USA
关键词
imaging; X-ray radiology; polycrystalline; mercuric iodide; imaging detectors; Flat-Panel imaging arrays;
D O I
10.1117/12.465604
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Single crystals of mercuric iodide have been studied for many years for nuclear detectors. We have investigated the use of x-ray photoconductive polycrystalline mercuric iodide coatings on amorphous silicon flat panel thin film transistor (TFT) arrays as x-ray detectors for radiographic and fluoroscopic applications in medical imaging. The mercuric iodide coatings were vacuum deposited by Physical Vapor Deposition (PVD). This coating technology is capable of being scaled up to sizes required in common medical imaging applications. Coatings were deposited on 4" X 4" TFT arrays for imaging performance evaluation and also on conductive-coated glass substrates for measurements of x-ray sensitivity, dark current and image lag. The TFT arrays used included pixel pitch dimensions of both 100 and 139 microns. Coating thickness between 150 microns and 250 microns were tested in the 25 kVp-100 kVp x-ray energy range utilizing exposures typical for both fluoroscopic, and radiographic imaging. X-ray sensitivities measured for the mercuric iodide samples and coated TFT detectors were superior to any published results for competitive materials (up to 7100 ke/mR/pixel for 100 micron pixels). It is believed that this higher sensitivity, can result in fluoroscopic imaging signal levels high enough to overshadow electronic noise. Image lag characteristics appear adequate for fluoroscopic rates. Resolution tests on resolution target phantoms showed that resolution is limited to the Nyquist frequency for the 139 micron (resolution similar to3.6 lp/mm) pixel detectors. The ability to operate at low voltages (similar to100 V) gives adequate dark currents for most applications and allows low voltage electronics designs. Mercuric Iodide coated TFT arrays were found to be outstanding candidates for direct digital radiographic detectors for both static and dynamic (fluoroscopic) applications. Their high x-ray sensitivity, high resolution, low dark current, low voltage operation, and good lag characteristics provide a unique combination of desirable imaging performance parameters.
引用
收藏
页码:592 / 600
页数:3
相关论文
共 50 条
  • [21] Large area HgI2 polycrystalline films for X-Ray imager
    Xu, G.
    Li, J. Y.
    Nan, R. H.
    Zhou, W. L.
    Gu, Z.
    Zhang, L.
    Ma, X. M.
    Cao, X. P.
    JOURNAL OF OPTOELECTRONICS AND ADVANCED MATERIALS, 2016, 18 (9-10): : 842 - 846
  • [22] Evaluation of a large-area amorphous-silicon X-ray imager
    Jensen, T
    Gray, JN
    REVIEW OF PROGRESS IN QUANTITATIVE NONDESTRUCTIVE EVALUATION, VOLS 20A AND 20B, 2001, 557 : 1860 - 1867
  • [23] Mercuric iodide and lead iodide x-ray detectors for radiographic and fluoroscopic medical imaging
    Zentai, G
    Partain, L
    Pavlyuchkova, R
    Proano, C
    Virshup, G
    Melekhov, L
    Zuck, A
    Breen, BN
    Dagan, O
    Vilensky, A
    Schieber, M
    Gilboa, H
    Bennet, P
    Shah, K
    Dmitriev, Y
    Thomas, J
    Yaffe, M
    Hunter, D
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 77 - 91
  • [24] Stability Study on Ceramic Mercuric Iodide (Red) X-Ray Sensor
    Singh, Kulvinder
    Saxena, Manoj
    16TH INTERNATIONAL WORKSHOP ON PHYSICS OF SEMICONDUCTOR DEVICES, 2012, 8549
  • [25] Recent advances in γ- and x-ray spectrometry by means of mercuric iodide detectors
    Lab PHASE, Strasbourg, France
    Nucl Instrum Methods Phys Res Sect A, 1-2 (173-178):
  • [26] High flux X-ray response of composite mercuric iodide detectors
    Schieber, M
    Zuck, A
    Melekhov, L
    Shatunovsky, R
    Hermon, H
    Turchetta, R
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS, 1999, 3768 : 296 - 309
  • [27] Nondestructive imaging with mercuric iodide thick film X-ray detectors
    Schieber, M
    Hermon, H
    Zuck, A
    Vilensky, A
    Melekhov, L
    Lukach, M
    Meerson, E
    Saado, H
    Shtekel, E
    Reisman, B
    Zentai, G
    Partain, L
    Seppi, E
    Pavlyuchkova, R
    Virshup, G
    Street, R
    Ready, S
    ADVANCED NONDESTRUCTIVE EVALUATION FOR STRUCTURAL AND BIOLOGICAL HEALTH MONITORING, 2001, 4335 : 43 - 51
  • [28] Deposition of thick films of polycrystalline mercuric iodide X-ray detectors
    Hermon, H
    Schieber, M
    Zuck, A
    Vilensky, A
    Melekhov, L
    Shtekel, E
    Green, A
    Dagan, O
    Ready, SE
    Street, RA
    Seppi, E
    Pavlyuchkova, R
    Virshup, G
    Zentai, G
    Partain, L
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 133 - 139
  • [29] CAPABILITIES OF MERCURIC IODIDE AS A ROOM-TEMPERATURE X-RAY DETECTOR
    SLAPA, M
    HUTH, GC
    SEIBT, W
    SCHIEBER, MM
    RANDTKE, PT
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1976, 23 (01) : 102 - 111
  • [30] Quantitative evaluation of mercuric iodide and selenium for x-ray imaging device
    Cha, BY
    Park, JK
    Kang, SS
    Shin, JW
    Kim, JY
    Lee, HW
    Nam, SH
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4496 - 4500