Regional pressure volume curves by electrical impedance tomography in a model of acute lung injury

被引:83
|
作者
Kunst, PWA [1 ]
Böhm, SH
de Anda, GV
Amato, MBP
Lachmann, B
Postmus, PE
de Vries, PMJM
机构
[1] Vrije Univ Amsterdam, Acad Hosp, Cardiovasc Res Inst, Dept Pulm Med, Amsterdam, Netherlands
[2] Erasmus Univ, Dept Anesthesiol, NL-3000 DR Rotterdam, Netherlands
[3] Univ Sao Paulo, Hosp Clin, Intens Care Unit, Sao Paulo, Brazil
关键词
acute lung injury; electrical impedance; pressure; tomography; volume;
D O I
10.1097/00003246-200001000-00029
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: A new noninvasive method, electrical impedance tomography (EIT), was used to make pressure-impedance (PI) curves in a lung lavage model of acute lung injury in pigs. The lower inflection point (LIP) and the upper deflection point (UDP) were determined from these curves and from the traditional pressure-volume (PV) curves to determine whether the PI curves resemble the traditional PV curves. Furthermore, regional differences in the mentioned determinants were investigated. Design: Prospective, experimental study. Setting: Animal research laboratory. Interventions: In nine anesthetized pigs, repeated lung lavage was performed until a Pa0(2) <80 torr was reached. Thereafter, an inspiratory PV curve was made using a constant flow of oxygen, During the intervention, EIT measurements were performed. Measurements and Main Results: In this study, the LIPEIT was within 2 cm H2O of the LIPPV. Furthermore, it was possible to visualize regional PI curves by EIT. No significant difference was found between the LIPPV (21.3 +/- 3.0 cm H2O) and the LIPEIT of the total lung (21.5 +/- 3.0 cm H2O) or the anterior parts of the lung (21.5 +/- 2.9 cm H2O). A significantly higher LIP (29.5 +/- 4.9 cm H2O) was found in the posterior parts of the lung. A UDPPV could be found in three animals only, whereas in all animals a UDPEIT could be determined from the anterior part of the lung. Conclusions Using EIT, determination of LIP and UDP from the regional PI curves is possible. The obtained information from the regional PI curves may help in understanding alveolar recruitment. The use of this new bedside technique for clinical decision making remains to be examined.
引用
收藏
页码:178 / 183
页数:6
相关论文
共 50 条
  • [31] Electrical impedance tomography in monitoring experimental lung injury
    Frerichs, I
    Hahn, G
    Schroder, T
    Hellige, G
    INTENSIVE CARE MEDICINE, 1998, 24 (08) : 829 - 836
  • [32] Tidal recruitment assessed by electrical impedance tomography and computed tomography in a porcine model of lung injury
    Muders, Thomas
    Luepschen, Henning
    Zinserling, Joerg
    Greschus, Susanne
    Fimmers, Rolf
    Guenther, Ulf
    Buchwald, Miriam
    Grigutsch, Daniel
    Leonhardt, Steffen
    Putensen, Christian
    Wrigge, Hermann
    CRITICAL CARE MEDICINE, 2012, 40 (03) : 903 - 911
  • [33] ELECTRICAL IMPEDANCE TOMOGRAPHY AND RESPIRATORY MECHANICS CHANGE IN TWEEN LAVAGE INDUCED ACUTE LUNG INJURY PORCINE MODEL
    Yu-Ling, Dai
    Hung, Chang
    Chung-Kan, Peng
    Kun-Lun, Huang
    RESPIROLOGY, 2018, 23 : 256 - 256
  • [34] Elastic pressure-volume curves in acute lung injury and acute respiratory distress syndrome
    Jonson, B
    INTENSIVE CARE MEDICINE, 2005, 31 (02) : 205 - 212
  • [35] Elastic pressure-volume curves in acute lung injury and acute respiratory distress syndrome
    Björn Jonson
    Intensive Care Medicine, 2005, 31 : 205 - 212
  • [36] Use of Electrical Impedance Tomography (EIT) to Estimate Global and Regional Lung Recruitment Volume (VREC) Induced by Positive End-Expiratory Pressure (PEEP): An Experiment in Pigs with Lung Injury
    Wang, Yu-Mei
    Sun, Xiu-Mei
    Zhou, Yi-Min
    Chen, Jing-Ran
    Cheng, Kun-Ming
    Li, Hong-Liang
    Yang, Yan-Lin
    Zhang, Linlin
    Zhou, Jian-Xin
    MEDICAL SCIENCE MONITOR, 2020, 26
  • [37] Electrical impedance tomography - a method for monitoring regional lung function
    Frerichs, I
    TECHNISCHES MESSEN, 2004, 71 (09): : 519 - 526
  • [38] Monitoring of regional lung ventilation using electrical impedance tomography
    Vasques, Francesco
    Sanderson, Barnaby
    Barrett, Nicholas A.
    Camporota, Luigi
    MINERVA ANESTESIOLOGICA, 2019, 85 (11) : 1231 - 1241
  • [39] Comparison of regional lung recruitment in electrical impedance tomograms and CT scans in experimental acute lung injury
    T Meier
    H Luepschen
    J Karsten
    T Leibecke
    M Großherr
    S Leonhardt
    Critical Care, 10 (Suppl 1):
  • [40] PRESSURE-VOLUME CURVES IN ACUTE PULMONARY INJURY
    Albaiceta, Guillermo M.
    MEDICINA INTENSIVA, 2009, 33 (05) : 243 - 250