Renal Near-Infrared Spectroscopy for Assessment of Renal Oxygenation in Adults Undergoing Cardiac Surgery: A Method Validation Study

被引:16
|
作者
Tholen, Maria [1 ]
Ricksten, Sven-Erik [1 ]
Lannemyr, Lukas [1 ]
机构
[1] Univ Gothenburg, Sahlgrenska Univ Hosp, Sahlgrenska Acad, Dept Anesthesiol & Intens Care Med, Gothenburg, Sweden
关键词
renal circulation; renal hypoxia; near-infrared spectroscopy; cardiac surgery;
D O I
10.1053/j.jvca.2020.04.044
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objective: To investigate the correlation between invasively measured renal venous oxygen saturation (SrvO(2)) and tissue oxygenation (rSO(2)) measured with near-infrared spectroscopy (NIRS) in adult patients undergoing cardiac surgery. Design: Prospective observational study. Setting: Single cardiac surgery center at a university hospital. Participants: Thirteen adult patients with skin- to- kidney distance <= 4 cm undergoing open cardiac surgery with cardiopulmonary bypass (CPB). Interventions: All patients received renal vein catheters for invasive measurement of SrvO(2), and NIRS electrodes for assessment of renal rSO(2) were placed over the kidney using ultrasound guidance. Measurements were made before CPB, during CPB at 3 different flow rates (2.4, 2.7, and 3.0 L/min/m(2)), and after CPB. Measurements and Main Results: Repeated- measures correlation analyses and Bland-Altman plots were used to study the correlation and agreement between rSO(2) and SrvO(2). For all measurement points, renal rSO(2) correlated with SrvO(2) (r(rm) = 0.61, p < 0.001), and the mean difference (bias) between rSO(2) and SrvO(2) was -2.71 +/- 7.22 (p = 0.002), with an error of 17.6%. When measurements during CPB and before and after CPB were studied separately, rSO(2) and SrvO(2) were correlated (r(rm) = 0.51, p < 0.007 and r(rm) = 0.73, p < 0.001, respectively). During CPB, renal rSO(2) predicted SrvO(2) with a bias of -3.41 +/- 7.76 (p = 0.009) and an error of 18.8%. Before and after CPB, the mean difference was -1.93 +/- 6.60 (p = 0.092), with an error of 16.2%. Conclusions: Renal rSO(2) is correlated to and predicts SrvO(2) with a small bias and acceptable agreement. Further studies are needed before renal NIRS can be used as a surrogate marker of renal oxygenation in clinical practice. (C) 2020 Elsevier Inc. All rights reserved.
引用
收藏
页码:3300 / 3305
页数:6
相关论文
共 50 条
  • [31] Randomized trial of near-infrared spectroscopy for personalized optimization of cerebral tissue oxygenation during cardiac surgery
    Rogers, C. A.
    Stoica, S.
    Ellis, L.
    Stokes, E. A.
    Wordsworth, S.
    Dabner, L.
    Clayton, G.
    Downes, R.
    Nicholson, E.
    Bennett, S.
    Angelini, G. D.
    Reeves, B. C.
    Murphy, G. J.
    BRITISH JOURNAL OF ANAESTHESIA, 2017, 119 (03) : 384 - 393
  • [33] APPLICATION OF NEAR-INFRARED SPECTROSCOPY FOR THE ASSESSMENT OF THE OXYGENATION LEVEL OF MYOGLOBIN AND HEMOGLOBIN IN CARDIAC-MUSCLE INVIVO
    THORNILEY, MS
    HOUSTON, R
    WICKRAMASINGHE, YABD
    ROLFE, P
    BIOCHEMICAL SOCIETY TRANSACTIONS, 1990, 18 (06) : 1195 - 1196
  • [34] Application of Near-Infrared Spectroscopy During Fetal Cardiac Surgery
    Reed, Casey A.
    Baker, R. Scott
    Lam, Christopher T.
    Hilshorst, Jerri L.
    Ferguson, Robert
    Lombardi, John
    Eghtesady, Pirooz
    JOURNAL OF SURGICAL RESEARCH, 2011, 171 (01) : 159 - 163
  • [35] Cerebral and Lower Limb Near-Infrared Spectroscopy in Adults on Extracorporeal Membrane Oxygenation
    Wong, Joshua K.
    Smith, Thomas N.
    Pitcher, Harrison T.
    Hirose, Hitoshi
    Cavarocchi, Nicholas C.
    ARTIFICIAL ORGANS, 2012, 36 (08) : 659 - 667
  • [36] Cerebral near-infrared spectroscopy in children undergoing heart surgery - Reply
    duPlessis, A
    Newburger, J
    Jonas, R
    Hickey, P
    Volpe, J
    ANNALS OF NEUROLOGY, 1996, 40 (05) : 818 - 819
  • [37] SPLANCHNIC-RENAL OXYGENATION MONITORING USING NEAR INFRARED SPECTROSCOPY (NIRS): IS IT A MARKER OF LOW CARDIAC OUTPUT STATE IN CHILDREN AFTER CARDIAC SURGERY?
    Bhalala, Utpal
    Mcqueen, Derrick
    Balaguru, Duraisamy
    Nishisaki, Akira
    Bird, Geoffrey
    Nathan, Meena
    Starr, Joanne
    CRITICAL CARE MEDICINE, 2009, 37 (12) : A97 - A97
  • [38] Evaluation of renal oxygenation by near-infrared spectroscopy during endoscopic injection of bulking agents in children with vesicoureteral reflux
    Çiğdem Arslan Alıcı
    Sema Şanal Baş
    Umut Alıcı
    Baran Tokar
    Pediatric Surgery International, 39
  • [39] Cerebral oxygenation and hemodynamic measurements during craniosynostosis surgery with near-infrared spectroscopy
    Martini, M.
    Rohrig, A.
    Wenghoefer, M.
    Schindler, E.
    Messing-Juenger, A. M.
    CHILDS NERVOUS SYSTEM, 2014, 30 (08) : 1367 - 1374
  • [40] Cerebral oxygenation and hemodynamic measurements during craniosynostosis surgery with near-infrared spectroscopy
    M. Martini
    A. Röhrig
    M. Wenghoefer
    E. Schindler
    A. M. Messing-Jünger
    Child's Nervous System, 2014, 30 : 1367 - 1374