Dosimetric Effect of Rotational Setup Errors in Single-Isocenter Volumetric-Modulated Arc Therapy of Multiple Brain Metastases

被引:19
|
作者
Selvan, Kasirajan Tamil [1 ,2 ]
Padma, Ganesan [1 ]
Revathy, Murugesan Kesavan [1 ]
Raj, N. Arunai Nambi [3 ]
Senthilnathan, K. [2 ]
Babu, P. Ramesh [2 ]
机构
[1] Apollo Canc Hosp, Dept Oncol, Hyderabad, Telangana, India
[2] Vellore Inst Technol Univ, Sch Adv Sci, Dept Phys, Vellore, Tamil Nadu, India
[3] Vellore Inst Technol Univ, Ctr Biomat Cellular & Mol Theranost, Vellore, Tamil Nadu, India
关键词
Rotational errors; stereotactic radiosurgery; volumetric modulated arc therapy; STEREOTACTIC RADIOSURGERY; RADIOTHERAPY; ACCURACY; RAPIDARC;
D O I
10.4103/jmp.JMP_103_18
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Aim: The aim of this study is to investigate the dosimetric effects of rotational uncertainties in patient positioning on target coverage in stereotactic radiosurgery (SRS) of multiple brain metastases using single isocenter volumetric modulated arc therapy (VMAT). Materials and Methods: Ten SRS cases with multiple brain metastases(2-8) planned with single isocenter non-coplanar VMAT technique were investigated in this study. Pitch, yaw and roll displacements of 1, 3 and 5 degrees were simulated about isocenter along each axes and nine plans were obtained for each case. Gross tumor volume (GTV) coverage obtained on simulated plans were compared with the original plan on four metrics of mean dose (Dmean), minimum dose to GTV (Dmin), dose to 95% of GTV (D95) and the volume covered by the 95% of the prescribed dose (V95). Results: At 1 degrees rotation about 4% of the PTV had V-95 and D95 values <95%. The minimum dose obtained across all PTV was 85.1% and there was not much change in the mean dose values obtained. The PTV volume which had V95 and D95 values <95% were in the range of 0.05 - 0.07cc and at a radial distance of 6.2 cm - 7.2 cm. At 3 degrees rotation almost 50% of the PTV had V95 and D95 values <95%. The minimum dose obtained across all PTV was 48.3% and the mean dose reduced to as low as 78.8%. At 5 degrees rotation almost 74% of the PTV had V95and D95 values <95%. The minimum dose obtained across all PTV was 21.2% and the mean dose reduced to as low as 49.2%. Conclusion: Our results indicate that correcting rotational uncertainties is critical in single-isocenter, multi-target SRS. For rotational deviations, radial distance of the target from isocenter along the respective axis has a strong influence on target coverage. For rotational setup deviation at a given radial distance larger targets tend to have lesser geometric miss compared to smaller targets. Mathematical model for spherical targets can be used to estimate V95 for given rotational errors.
引用
收藏
页码:84 / 90
页数:7
相关论文
共 50 条
  • [31] Effect of setup error in the single-isocenter technique on stereotactic radiosurgery for multiple brain metastases
    Nakano, Hisashi
    Tanabe, Satoshi
    Utsunomiya, Satoru
    Yamada, Takumi
    Sasamoto, Ryuta
    Nakano, Toshimichi
    Saito, Hirotake
    Takizawa, Takeshi
    Sakai, Hironori
    Ohta, Atsushi
    Abe, Eisuke
    Kaidu, Motoki
    Aoyama, Hidefumi
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (12): : 155 - 165
  • [32] Evaluation of Multileaf Collimator Width and Beam Quality Effect On Single-Isocenter Multitarget Radiosurgery Using Volumetric-Modulated Arc Therapy
    Laoui, S.
    Roa, D.
    Kuo, J.
    Baldytchev, M.
    Gonzales, J.
    Dietrich, S.
    Ramsinghani, N.
    Daroui, P.
    Farol, H.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E554 - E554
  • [33] Dosimetric effect of rotational setup errors in stereotactic radiosurgery with HyperArc for single and multiple brain metastases
    Sagawa, Tomohiro
    Ohira, Shingo
    Ueda, Yoshihiro
    Akino, Yuichi
    Mizuno, Hirokazu
    Matsumoto, Masao
    Miyazaki, Masayoshi
    Teshima, Teruki
    Koizumi, Masahiko
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (10): : 84 - 91
  • [34] Plan Quality and Delivery Time Comparison of Single-Isocenter Volumetric Modulated Arc Therapy Versus Stereotactic Radiosurgery for Brain Metastases
    Henke, L. E.
    Kashani, R.
    Abraham, C. D., Sr.
    Huang, J.
    Tsien, C.
    Dowling, J.
    Rich, K.
    Knutson, N.
    Robinson, C. G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E521 - E522
  • [35] Dosimetric Effect of Jaw Tracking in Volumetric-Modulated Arc Therapy
    Thongsawad, Sangutid
    Khamfongkhruea, Chirasak
    Tannanonta, Chirapha
    [J]. JOURNAL OF MEDICAL PHYSICS, 2018, 43 (01) : 52 - 57
  • [36] Single Isocenter Radiosurgery for the Simultaneous Treatment of Multiple Brain Metastases: Volumetric Modulated Arc Therapy Or Dynamic Conformal Arc?
    Maurer, J.
    Sintay, B.
    Manning, M.
    Wiant, D.
    Liu, H.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3580 - 3580
  • [37] Dosimetric Comparison between Single-Isocenter RapidArc and Multi-Isocenter Circular Cone Arc for Radiosurgery to Multiple Brain Metastases
    Dai, P.
    Zhang, P.
    Luo, L.
    Deng, G.
    Huang, Q.
    Cai, L.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E508 - E508
  • [38] Optimization of isocenter position for multiple brain metastases single-isocenter stereotactic radiosurgery to minimize dosimetric variations due to rotational uncertainty
    Shen, Zhenjiong
    wang, Hao
    Shao, Yan
    Duan, Yanhua
    Gu, Hengle
    Chen, Hua
    Feng, Aihui
    Huang, Ying
    Xu, Zhiyong
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 111
  • [39] Dosimetric Impact of Rotational Setup Errors for Cervical Cancer Patients With Postoperative Volumetric Modulated Arc Therapy
    Tsujii, K.
    Ueda, Y.
    Isono, M.
    Miyazaki, M.
    Teshima, T.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E603 - E603
  • [40] Longitudinal Grouping of Target Volumes for Volumetric-Modulated Arc Therapy of Multiple Brain Metastases
    Xu, Yingjie
    Miao, Junjie
    Liu, Qingfeng
    Huang, Peng
    Ma, Pan
    Chen, Xinyuan
    Men, Kuo
    Xiao, Jianping
    Dai, Jianrong
    [J]. FRONTIERS IN ONCOLOGY, 2021, 11