Systematic assessment of multi-echo dynamic susceptibility contrast MRI using a digital reference object

被引:13
|
作者
Stokes, Ashley M. [1 ]
Semmineh, Natenael B. [1 ]
Nespodzany, Ashley [1 ]
Bell, Laura C. [1 ]
Quarles, C. Chad [1 ]
机构
[1] Barrow Neurol Inst, Div Neuroimaging Res, 350 W Thomas Rd, Phoenix, AZ 85013 USA
关键词
brain tumor hemodynamics; digital reference object (DRO); dynamic susceptibility contrast (DSC); multi-echo; perfusion; CEREBRAL BLOOD-VOLUME; SINGULAR-VALUE DECOMPOSITION; DSC-MRI; LEAKAGE CORRECTION; AGENT EXTRAVASATION; PERFUSION; GRADE; MODEL; FLOW; T-1;
D O I
10.1002/mrm.27914
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Brain tumor dynamic susceptibility contrast (DSC) MRI is adversely impacted by T-1 and T2* contrast agent leakage effects that result in inaccurate hemodynamic metrics. While multi-echo acquisitions remove T-1 leakage effects, there is no consensus on the optimal set of acquisition parameters. Using a computational approach, we systematically evaluated a wide range of acquisition strategies to determine the optimal multi-echo DSC-MRI perfusion protocol. Methods Using a population-based DSC-MRI digital reference object (DRO), we assessed the influence of preload dosing (no preload and full dose preload), field strength (1.5 and 3T), pulse sequence parameters (echo time, repetition time, and flip angle), and leakage correction on relative cerebral blood volume (rCBV) and flow (rCBF) accuracy. We also compared multi-echo DSC-MRI protocols with standard single-echo protocols. Results Multi-echo DSC-MRI is highly consistent across all protocols, and multi-echo rCBV (with or without use of a preload dose) had higher accuracy than single-echo rCBV. Regression analysis showed that choice of repetition time and flip angle had minimal impact on multi-echo rCBV and rCBV, indicating the potential for significant flexibility in acquisition parameters. The echo time combination had minimal impact on rCBV, though longer echo times should be avoided, particularly at higher field strengths. Leakage correction improved rCBV accuracy in all cases. Multi-echo rCBF was less biased than single-echo rCBF, although rCBF accuracy was reduced overall relative to rCBV. Conclusions Multi-echo acquisitions were more robust than single-echo, essentially decoupling both repetition time and flip angle from rCBV accuracy. Multi-echo acquisitions obviate the need for preload dosing, although leakage correction to remove residual T2* leakage effects remains compulsory for high rCBV accuracy.
引用
收藏
页码:109 / 123
页数:15
相关论文
共 50 条
  • [1] A Dynamic Susceptibility Contrast MRI Digital Reference Object for Testing Software with Leakage Correction
    Chen, H.
    Jen, M.
    Hou, P.
    Stafford, R.
    Liu, H.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [2] Rapid simultaneous estimation of relaxation rates using multi-echo, multi-contrast MRI
    Keeling, Elizabeth G.
    Sisco, Nicholas J.
    Mcelvogue, Molly M.
    Borazanci, Aimee
    Dortch, Richard D.
    Stokes, Ashley M.
    [J]. MAGNETIC RESONANCE IMAGING, 2024, 112 : 116 - 127
  • [3] Optimization of Acquisition and Analysis Methods for Clinical Dynamic Susceptibility Contrast MRI Using a Population-Based Digital Reference Object
    Semmineh, N. B.
    Bell, L. C.
    Stokes, A. M.
    Hu, L. S.
    Boxerman, J. L.
    Quarles, C. C.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2018, 39 (11) : 1981 - 1988
  • [4] A dynamic susceptibility contrast MRI digital reference object for testing software with leakage correction: Effect of background simulation
    Chen, Henry Szu-Meng
    Jen, Mu-Lan
    Hou, Pin
    Stafford, R. Jason
    Liu, Ho-Ling
    [J]. MEDICAL PHYSICS, 2021, 48 (10) : 6051 - 6059
  • [5] Identifying systematic errors in quantitative dynamic-susceptibility contrast perfusion imaging by high-resolution multi-echo parallel EPI
    Jochimsen, Thies H.
    Newbould, Rexford D.
    Skare, Stefan T.
    Clayton, David B.
    Albers, Gregory W.
    Moseley, Michael E.
    Bammer, Roland
    [J]. NMR IN BIOMEDICINE, 2007, 20 (04) : 429 - 438
  • [6] A Population-Based Digital Reference Object (DRO) for Optimizing Dynamic Susceptibility Contrast (DSC)-MRI Methods for Clinical Trials
    Semmineh, Natenael B.
    Stokes, Ashley M.
    Bell, Laura C.
    Boxerman, Jerrold L.
    Quarles, C. Chad
    [J]. TOMOGRAPHY, 2017, 3 (01) : 41 - 49
  • [7] Improved Susceptibility Weighted Imaging Method Using Multi-Echo Acquisition
    Oh, Sung Suk
    Oh, Se-Hong
    Nam, Yoonho
    Han, Dongyeob
    Stafford, Randall B.
    Hwang, Jinyoung
    Kim, Dong-Hyun
    Park, HyunWook
    Lee, Jongho
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (02) : 452 - 458
  • [8] A novel noise removal using homomorphic normalization for multi-echo knee MRI
    Cui, Xuenan
    Kim, Hakil
    Hong, Seongwook
    Kwack, Kyu-Sung
    [J]. IEICE ELECTRONICS EXPRESS, 2011, 8 (08): : 604 - 611
  • [9] Robust selective weighted field mapping using multi-echo gradient echo-based MRI
    Fatnassi, Chemseddine
    Zaidi, Habib
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (21):
  • [10] Measurement of cerebral perfusion with dual-echo multi-slice quantitative dynamic susceptibility contrast MRI
    Vonken, EJPA
    van Osch, MJP
    Bakker, CJG
    Viergever, MA
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1999, 10 (02): : 109 - 117