Continuous glucose monitoring and control in a rotating wall perfused bioreactor

被引:9
|
作者
Xu, YH
Sun, JJ
Mathew, G
Jeevarajan, AS
Anderson, MM
机构
[1] NASA, Lyndon B Johnson Space Ctr, Biol Syst Off, Houston, TX 77058 USA
[2] Wyle Labs Life Sci Syst & Serv, Houston, TX 77058 USA
关键词
glucose control; in-line glucose sensor; rotating; bioreactor; BHK-21; cells; cell culture;
D O I
10.1002/bit.20141
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
A glucose control system consisting of a single in-line glucose sensor, concentrated glucose solution, and computer hardware and software were developed. The system was applied to continuously control glucose concentrations of a perfusion medium in a rotating wall perfused vessel (RWPV) bioreactor culturing BHK-21 cells. The custom-made glucose sensor was based on a hydrogen peroxide electrode. The sensor continuously and accurately measured the glucose concentration of GTSF-2 medium in the RWPV bioreactor during cell culture. Three sets of two-point calibrations were applied to the glucose sensor during the 55-day cell culture. The system first controlled the glucose concentration in perfusing medium between 4.2 and 5.6 mM for 36 days and then at different glucose levels for 19 days. A stock solution with a high glucose concentration (266 mM) was used as the glucose injection solution. The standard error of prediction (SEP) for glucose measurement by the sensor, compared to measurement by the Beckman glucose analyzer, was +/-0.4 mM for 55 days. (C) 2004 Wiley Periodical, Inc.
引用
收藏
页码:473 / 477
页数:5
相关论文
共 50 条
  • [31] Variable structure control for a continuous bioreactor
    Tham, HJ
    Hussain, MA
    Ramachandran, KB
    IEEE 2000 TENCON PROCEEDINGS, VOLS I-III: INTELLIGENT SYSTEMS AND TECHNOLOGIES FOR THE NEW MILLENNIUM, 2000, : 433 - 436
  • [32] Calibration of the continuous glucose monitoring system for transient glucose monitoring
    Lesperance, Leann M.
    Spektor, Anna
    McLeod, Kenneth J.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2007, 9 (02) : 183 - 190
  • [33] MONITORING THE KINETICS OF GLUCOSE TRANSFER ACROSS A BIOREACTOR MEMBRANE WITH A GLUCOSE SENSOR
    VELHO, G
    REACH, G
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1989, 12 (08): : 539 - 543
  • [34] Continuous glucose monitoring in diabetes
    Moe, Samantha S.
    Kolber, Michael R.
    Falk, Jamie
    CANADIAN FAMILY PHYSICIAN, 2023, 69 (06) : 406 - 406
  • [35] Continuous glucose monitoring and pumps
    Cohen, Neale
    AUSTRALIAN FAMILY PHYSICIAN, 2015, 44 (05) : 284 - 287
  • [36] The Future of Continuous Glucose Monitoring
    Garg, Satish K.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2009, 11 : S1 - S3
  • [37] Continuous Glucose Monitoring in 2018
    Bode, Bruce W.
    Battelino, Tadej
    DIABETES TECHNOLOGY & THERAPEUTICS, 2019, 21 : S13 - S31
  • [38] Reimbursement for Continuous Glucose Monitoring
    Heinemann, Lutz
    DeVries, J. Hans
    DIABETES TECHNOLOGY & THERAPEUTICS, 2016, 18 : 48 - 52
  • [39] Continuous Glucose Monitoring in 2014
    Bode, Bruce W.
    Battelino, Tadej
    DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 : S12 - S20
  • [40] Continuous Glucose Monitoring in Gastroparesis
    Zeeshan Ramzan
    Frank Duffy
    Javier Gomez
    Robert S. Fisher
    Henry P. Parkman
    Digestive Diseases and Sciences, 2011, 56