Technical Report: TG-142 compliant and comprehensive quality assurance tests for respiratory gating

被引:18
|
作者
Woods, Kyle [1 ]
Rong, Yi [2 ]
机构
[1] Ohio State Univ, Dept Radiat Oncol, Columbus, OH 43210 USA
[2] Univ Calif Davis, Dept Radiat Oncol, Ctr Comprehens Canc, Sacramento, CA 95817 USA
关键词
respiratory gating; temporal accuracy; time delays; real-time position management; calibration of surrogate; GATED RADIATION-THERAPY; RADIOTHERAPY; MOTION; DELIVERY; ONCOLOGY; SYSTEM; DELAYS;
D O I
10.1118/1.4932363
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To develop and establish a comprehensive gating commissioning and quality assurance procedure in compliance with TG-142. Methods: Eight Varian TrueBeam Linacs were used for this study. Gating commissioning included an end-to-end test and baseline establishment. The end-to-end test was performed using a CIRS dynamic thoracic phantom with a moving cylinder inside the lung, which was used for carrying both optically simulated luminescence detectors (OSLDs) and Gafchromic EBT2 films while the target is moving, for a point dose check and 2D profile check. In addition, baselines were established for beam-on temporal delay and calibration of the surrogate, for both megavoltage (MV) and kilovoltage (kV) beams. A motion simulation device (MotionSim) was used to provide periodic motion on a platform, in synchronizing with a surrogate motion. The overall accuracy and uncertainties were analyzed and compared. Results: The OSLD readings were within 5% compared to the planned dose (within measurement uncertainty) for both phase and amplitude gated deliveries. Film results showed less than 3% agreement to the predicted dose with a standard sinusoid motion. The gate-on temporal accuracy was averaged at 139 +/- 10 ms for MV beams and 92 +/- 11 ms for kV beams. The temporal delay of the surrogate motion depends on the motion speed and was averaged at 54.6 +/- 3.1 ms for slow, 24.9 +/- 2.9 ms for intermediate, and 23.0 +/- 20.1 ms for fast speed. Conclusions: A comprehensive gating commissioning procedure was introduced for verifying the output accuracy and establishing the temporal accuracy baselines with respiratory gating. The baselines are needed for routine quality assurance tests, as suggested by TG-142. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:6488 / 6497
页数:10
相关论文
共 40 条
  • [21] Commissioning and Comprehensive Quality Assurance of commercial 3D Treatment Planning System using IAEA Technical Report Series — 430
    S. V. Jamema
    R. R. Upreti
    S. Sharma
    D. D. Deshpande
    Australasian Physics & Engineering Sciences in Medicine, 2008, 31 : 207 - 215
  • [22] Report of AAPM TG 135: Quality assurance for robotic radiosurgery (vol 38, pg 2914, 2011)
    Dieterich, Sonja
    Cavedon, Carlo
    Chuang, Cynthia F.
    Cohen, Alan B.
    Garrett, Jeffrey A.
    Lee, Charles L.
    Lowenstein, Jessica R.
    d'Souza, Maximian F.
    Taylor, David D., Jr.
    Wu, Xiaodong
    Yu, Cheng
    MEDICAL PHYSICS, 2011, 38 (09) : 5264 - 5264
  • [23] Commissioning and Comprehensive Quality Assurance of commercial 3D Treatment Planning System using IAEA Technical Report Series-430
    Jamema, S. V.
    Upreti, R. R.
    Sharma, S.
    Deshpande, D. D.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2008, 31 (03) : 207 - 215
  • [24] A transit portal dosimetry method for respiratory gating quality assurance with a dynamic 3D printed tumor phantom
    Tan, Hong Qi
    Koh, Calvin Wei Yang
    Tan, Lloyd Kuan Rui
    Lew, Kah Seng
    Chua, Clifford Ghee Ann
    Ang, Khong Wei
    Lee, James Cheow Lei
    Park, Sung Yong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (05):
  • [25] Quality assurance device for four-dimensional IMRT or SBRT and respiratory gating using patient-specific intrafraction motion kernels
    Nelms, Benjamin E.
    Ehler, Eric
    Bragg, Henry
    Tome, Wolfgang A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2007, 8 (04): : 152 - 168
  • [26] Technical Note: A novel quality assurance test to identify gantry angle inaccuracies in respiratory-gated VMAT treatments
    Hubley, Emily
    Briscoe, Michael
    Ploquin, Nicolas
    Pierce, Greg
    MEDICAL PHYSICS, 2017, 44 (10) : 5075 - 5080
  • [27] Comprehensive Quality Assurance Test of Second Generation Toshiba Aquilion Large Bore CT Simulator Based On AAPM TG-66 Recommendations
    Zhang, D.
    MEDICAL PHYSICS, 2015, 42 (06) : 3246 - 3246
  • [28] The untapped potential of physical activity monitoring for quality assurance of field-based walking tests in clinical respiratory trials
    Orme, Mark W.
    Pina, Ilaria
    Singh, Sally J.
    CHRONIC RESPIRATORY DISEASE, 2022, 19
  • [29] Quality assurance for image-guided radiation therapy utilizing CT-based technologies: A report of the AAPM TG-179
    Bissonnette, Jean-Pierre
    Balter, Peter A.
    Dong, Lei
    Langen, Katja M.
    Lovelock, D. Michael
    Miften, Moyed
    Moseley, Douglas J.
    Pouliot, Jean
    Sonke, Jan-Jakob
    Yoo, Sua
    MEDICAL PHYSICS, 2012, 39 (04) : 1946 - 1963
  • [30] Physical and Technical Quality Assurance in German Breast Cancer Screening: Progress Report of the Reference Center Muenster after Three Years
    Sommer, A.
    Girnus, R.
    Wendt, B.
    Czwoydzinski, J.
    Wuestenbecker, C.
    Heindel, W.
    Lenzen, H.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2009, 181 (05): : 447 - 453