Evaluating the spectroradiometric performance of an uncooled midwave infrared hyperspectral interferometer using a microbolometer array detector

被引:2
|
作者
Honniball, Casey, I [1 ]
Wright, Robert [1 ]
Lucey, Paul G. [1 ]
Khayat, Alain S. J. [2 ,3 ]
机构
[1] Univ Hawaii Manoa, Hawaii Inst Geophys & Planetol, Dept Earth & Planetary Sci, Honolulu, HI 96822 USA
[2] NASA, Goddard Space Flight Ctr, Greenbelt, MD USA
[3] Univ Maryland, Dept Astron, Ctr Res & Explorat Space Sci & Technol, College Pk, MD 20742 USA
关键词
hyperspectral; Sagnac; interferometer; uncooled; microbolometer; mid-wave infrared; THERMAL EMISSION; SPECTROMETER; SATELLITE; SMIFTS;
D O I
10.1117/1.OE.59.7.074103
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Improved technology and emerging interferometric techniques have allowed the use of uncooled microbolometers in the long-wave infrared (LWIR; 8 to 14 mu m) for hyperspectral imaging (HSI). The midwave infrared (MWIR; 3 to 5 mu m) presents several advantages with respect to the LWIR for Earth and planetary science. For example, important atmospheric trace gases on Earth such as CO2 and CH4 are not masked by other atmospheric constituents in the MWIR. However, HSI in the MWIR is more challenging at ambient Earth temperatures because less radiance is available to measure. We describe how hyperspectral images in the MWIR can be acquired with an instrument using an uncooled microbolometer married to a Sagnac interferometer. Standard characterization tests are used to benchmark the performance of the microbolometer instrument with a cryogenically cooled photon detector with the same optical design. At a spectral resolution of 100 cm(-1) (17 bands between 3 and 5 mu m), we measured a signal-to-noise ratio (SNR) of 100 at 30 degrees C with the microbolometer instrument and an SNR of 50 at 50 cm(-1) (33 bands). Results from this work show that coupling microbolometers with interferometers allows for quality measurements with adequate SNR for high-temperature science applications. (C) 2020 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:16
相关论文
共 50 条
  • [21] Research of High Sensitivity Uncooled Infrared Detector Array
    Zhang, Ping-chuan
    Zhang, Bo
    [J]. 3RD INTERNATIONAL PHOTONICS AND OPTOELECTRONICS MEETINGS (POEM 2010), 2011, 276
  • [22] Uncooled Microbolometer Infrared Focal Plane Array Without Substrate Temperature Stabilization
    Lv, Jian
    Que, Longcheng
    Wei, Linhai
    Zhou, Yun
    Liao, Baobin
    Jiang, Yadong
    [J]. IEEE SENSORS JOURNAL, 2014, 14 (05) : 1533 - 1544
  • [23] Infrared holography using a microbolometer array
    George, Nicholas
    Khare, Kedar
    Chi, Wanli
    [J]. APPLIED OPTICS, 2008, 47 (04) : A7 - A12
  • [24] Mid-infrared (longwave infrared) passive spectroscopic imaging with an uncooled microbolometer array sensor
    Morimoto, Yusuke
    Tahara, Shiori
    Yamashita, So
    Anabuki, Daichi
    Kitazaki, Tomoya
    Shimizu, Hiroko
    Nishiyama, Akira
    Wada, Kenji
    Ishimaru, Ichiro
    [J]. ELECTRO-OPTICAL AND INFRARED SYSTEMS: TECHNOLOGY AND APPLICATIONS XIX, 2022, 12271
  • [25] Performance analysis of low-cost uncooled microbolometer infrared detectors
    Zhang Junju
    Sun Lianjun
    Wang Shiyun
    Chang Benkang
    Qian Yunsheng
    Yu Chunyu
    [J]. OPTOELETRONIC MATERIALS AND DEVICES, PTS 1 AND 2, 2006, 6352
  • [26] A LWIR hyperspectral imager using a Sagnac interferometer and cooled HgCdTe detector array
    Lucey, Paul G.
    Wood, Mark
    Crites, Sarah T.
    Akagi, Jason
    [J]. ALGORITHMS AND TECHNOLOGIES FOR MULTISPECTRAL, HYPERSPECTRAL, AND ULTRASPECTRAL IMAGERY XVIII, 2012, 8390
  • [27] Quantitative thermal nondestructive evaluation using an uncooled microbolometer infrared camera
    Zalameda, JN
    Winfree, WP
    [J]. THERMOSENSE XXIV, 2002, 4710 : 610 - 617
  • [28] A low cost uncooled infrared microbolometer focal plane array using the CMOS n-well layer
    Tezcan, DS
    Eminoglu, S
    Akar, OS
    Akin, T
    [J]. 14TH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 2001, : 566 - 569
  • [29] Microbolometer detector array for satellite-based thermal infrared imaging
    Pope, TD
    Alain, C
    Bergeron, A
    Jerominek, H
    Zayer, I
    Bézy, JL
    [J]. 2004 INTERNATIONAL CONFERENCE ON MEMS, NANO AND SMART SYSTEMS, PROCEEDINGS, 2004, : 310 - 313
  • [30] Performance and results from a space borne, uncooled microbolometer array spectral radiometric imager
    Spinhirne, JD
    Scott, VS
    Lancaster, RS
    Manizade, K
    Palm, SP
    [J]. 2000 IEEE AEROSPACE CONFERENCE PROCEEDINGS, VOL 3, 2000, : 125 - 134