Precise determination of hypochlorite in commercial bleaches with established traceability using automatic potentiometric titration

被引:10
|
作者
Gonzaga, Fabiano Barbieri [1 ]
Cordeiro, Lidiane Rodrigues [1 ]
机构
[1] Natl Inst Metrol Qual & Technol INMETRO, Chem Metrol Div, BR-25250020 Duque De Caxias, RJ, Brazil
关键词
Potentiometric titration; Hypochlorite determination; Bleach; COULOMETRIC TITRATION; CHLORINE BLEACH; PRODUCTS; WATER;
D O I
10.1007/s00769-014-1059-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A measurement procedure is proposed for precisely determining hypochlorite in commercial bleaches, with established traceability and a full description of its uncertainty sources. The procedure is based on automatic volumetric titration with potentiometric detection and was evaluated in the analysis of several commercial bleach samples, with Na-hypochlorite mass fractions in the range 0.002-0.027 g g(-1), with an average expanded uncertainty (k = 2) of 0.00027 g g(-1) and an average relative standard deviation between measurement triplicates of 0.36 %. Compared with a standard procedure based on manual volumetric titration with visual colorimetric detection, the improved procedure produced equivalent results with a lower average expanded uncertainty (0.00029 g g(-1) as compared to 0.0015 g g(-1) for the standard procedure). This was related to better precision of the volume measurement using an automatic buret and to better precision of the titration endpoint determination using potentiometric detection. Compared with other studies reported in the literature, the improved procedure presented a lower average relative standard deviation between measurement replicates (0.36 % as compared to 1.00-11.53 %).
引用
收藏
页码:283 / 287
页数:5
相关论文
共 50 条
  • [11] RAPID DETERMINATION OF ARTEMISININ BY IODOMETRY COMBINED WITH AUTOMATIC POTENTIOMETRIC TITRATION
    Fan, Yan-Ru
    Wang, Yue
    Huang, Yu
    Lu, Xiao-Quan
    OXIDATION COMMUNICATIONS, 2015, 38 (04): : 1731 - 1738
  • [12] Determination of vanadium(V) by direct automatic potentiometric titration with EDTA using a chemically modified electrode as a potentiometric sensor
    Quintar, SE
    Santagata, JP
    Cortinez, VA
    TALANTA, 2005, 67 (04) : 843 - 847
  • [13] DETERMINATION OF OVALBUMIN BY USING A POTENTIOMETRIC TITRATION TECHNIQUE
    KAWANO, K
    YOSHINAGA, T
    TANIMURA, T
    BUNSEKI KAGAKU, 1983, 32 (06) : 347 - 351
  • [14] POTENTIOMETRIC AUTOMATIC TITRATION OF ALUMINUM USING FLUORIDE SELECTIVE ELECTRODE
    BLAZ, T
    KUBIAK, W
    CHEMIA ANALITYCZNA, 1981, 26 (01): : 157 - 160
  • [15] Determination of gold in alloys using potentiometric titration with thiocarbamide
    Zhitenko, LP
    Rysev, AP
    Khomutova, EG
    Solomonova, AS
    INDUSTRIAL LABORATORY, 1998, 64 (04): : 217 - 219
  • [16] DETERMINATION OF PALLADIUM(II) WITH DITHIOOXAMIDE USING POTENTIOMETRIC TITRATION
    STANKOVIC, B
    STEFANOVIC, A
    DUGANDZIC, M
    MIKROCHIMICA ACTA, 1977, 1 (3-4) : 395 - 399
  • [17] CRITICAL COMPARISON OF EVALUATION METHODS OF POTENTIOMETRIC TITRATION DATA USING A MICROCOMPUTER-CONTROLLED AUTOMATIC POTENTIOMETRIC TITRATION SYSTEM
    HAZI, J
    DARUHAZI, L
    PINTER, O
    SZAKALY, Z
    MIKROCHIMICA ACTA, 1986, 2 (1-6) : 379 - 388
  • [18] Validation and Application of an Automatic Potentiometric Titration Method for Determination of Artesunate and its Tablets
    Fan, Yan-Ru
    Chen, Lu-Lu
    Huang, Yu
    LATIN AMERICAN JOURNAL OF PHARMACY, 2021, 40 (09): : 2064 - 2067
  • [19] Automatic potentiometric flow titration procedure for ascorbic acid determination in pharmaceutical formulations
    Paim, APS
    Almeida, CMNV
    Reis, BF
    Lapa, RAS
    Zagatto, EAG
    Lima, JLFC
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2002, 28 (06) : 1221 - 1225
  • [20] DETERMINATION OF ARSENIC BY METHOD OF AUTOMATIC POTENTIOMETRIC TITRATION OF NON-AQUEOUS SOLUTIONS
    KRESHKOV, AP
    MILAEV, SM
    KHANTURG.GI
    ZAVODSKAYA LABORATORIYA, 1974, 40 (08): : 926 - 928