Aortic Insufficiency During HeartMate 3 Left Ventricular Assist Device Support

被引:20
|
作者
Imamura, Teruhiko [1 ,2 ]
Narang, Nikhil [3 ]
Kim, Gene [1 ]
Nitta, Daisuke [1 ]
Fujino, Takeo [1 ]
Nguyen, Ann [1 ]
Grinstein, Jonathan [1 ]
Rodgers, Daniel [1 ]
Ota, Takeyoshi [4 ]
Jeevanandam, Valluvan [4 ]
Sayer, Gabriel [5 ]
Uriel, Nir [5 ]
机构
[1] Univ Chicago, Med Ctr, Dept Med, Chicago, IL 60637 USA
[2] Univ Toyama, Dept Med 2, 2630 Sugitani, Toyama, Toyama 9300194, Japan
[3] Div Advocate Christ Med Ctr, Oak Lawn, IL USA
[4] Univ Chicago, Med Ctr, Dept Surg, Chicago, IL 60637 USA
[5] Columbia Univ, Cardiol Div, Med Ctr, New York, NY USA
关键词
heart failure; hemodynamic; regurgitation; QUANTIFICATION;
D O I
10.1016/j.cardfail.2020.05.013
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Aortic insufficiency (AI) is associated with morbidity and mortality in patients with continuous-flow left ventricular assist devices (LVADs), whereas its impact on the HeartMate 3 LVAD cohorts remains uninvestigated. We aimed to investigate the clinical impact of AI on patients with HeartMate 3 LVADs. Methods and Results: Consecutive 61 patients (median age 54 years; 67% male) implanted with HeartMate 3 LVAD between 2015 and 2019 were enrolled and underwent echocardiography at 3 months after LVAD implantation. AI severity was quantified by the novel Doppler echocardiographic method obtained at the outflow cannula and the calculated regurgitation fraction of 30% or greater (moderate or greater) was defined as significant. At 3 months after implant, 12 patients (20%) had significant AI. They had a higher incidence of death or heart failure readmissions compared with those without significant AI during a 1-year observational period (70% vs 24%, P =.003) with an adjusted hazard ratio of 2.76 (95% confidence interval 1.03-7.88). Conclusions: In patients with HeartMate 3 LVAD support, significant AI remains both prevalent and a clinically significant downstream complication.
引用
收藏
页码:863 / 869
页数:7
相关论文
共 50 条
  • [21] Recovery Plug for HeartMate 3 Left Ventricular Assist Device
    Mulzer, J.
    Potapov, E.
    Kaufmann, F.
    Mueller, M.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2018, 37 (04): : S480 - S480
  • [22] Recovery plug for HeartMate 3 left ventricular assist device
    Mulzer, Johanna
    Faerber, Gloria
    Kaufmann, Friedrich
    Potapov, Evgenij V.
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2019, 157 (02): : E35 - E37
  • [23] Left Ventricular Assist Device-Acquired Aortic Insufficiency
    Akashi, Hirokazu
    [J]. CIRCULATION JOURNAL, 2015, 79 (01) : 43 - 44
  • [24] The Hemodynamic Effects of Aortic Regurgitation in Patients Supported by a HeartMate 3 Left Ventricular Assist Device
    Rubinstein, Gal
    Moeller, Cathrine m.
    Lotan, Dor
    Slomovich, Sharon
    Fernandez-valledor, Andrea
    Ranard, Lauren s.
    Leahy, Nicole e.
    Ladanyi, Annamaria
    Oren, Daniel
    Clerkin, Kevin j.
    Raikhelkar, Jayant k.
    Topkara, Veli k.
    Fried, Justin a.
    Vahl, Torsten
    Colombo, Paolo c.
    Kaku, Yuji
    Takeda, Koji
    Naka, Yoshifumi
    Yuzefpolskaya, Melana
    Sayer, Gabriel t.
    Uriel, Nir
    [J]. JOURNAL OF CARDIAC FAILURE, 2024, 30 (01) : 95 - 99
  • [25] Progression of aortic valve insufficiency during centrifugal versus axial flow left ventricular assist device support
    Contreras, Fabian Jimenez
    Pla, Michelle Mendiola
    Schroder, Jacob
    Bryner, Benjamin
    Agarwal, Richa
    Russell, Stuart D.
    Mirza, Jacqueline
    Daneshmand, Mani A.
    Milano, Carmelo
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2022, 61 (05) : 1188 - 1196
  • [26] Transcatheter Aortic Valve Replacement Management of Aortic Insufficiency in Patients with Left Ventricular Assist Device Support
    Kar, B.
    Prathipati, P.
    Macedo, F. Y.
    Nathan, S.
    Radovancevic, R.
    Loyalka, P.
    Gregoric, I. D.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2017, 36 (04): : S438 - S438
  • [27] HeartMate 3 Left Ventricular Assist Device Implant and Repair of a Left Ventricular Pseudoaneurysm
    Meknat, Aryan
    Kanwar, Manreet K.
    Lee, Candice Y.
    Tsukashita, Masaki
    [J]. CIRCULATION-HEART FAILURE, 2024, 17 (01) : E010925
  • [28] Titanium Fastener Utilization During HeartMate 3 Left Ventricular Assist Device Implantation
    Johnson, Carl A., Jr.
    Melvin, Amber L.
    Robinson, Davida A.
    Amirjamshidi, Hossein
    Knight, Peter A.
    Gosev, Igor
    [J]. INNOVATIONS-TECHNOLOGY AND TECHNIQUES IN CARDIOTHORACIC AND VASCULAR SURGERY, 2018, 13 (05) : 361 - 364
  • [29] Assessment of Severity and Implication of Aortic Insufficiency During Left Ventricular Assist Device Supports
    Imamura, Teruhiko
    Narang, Nikhil
    [J]. ASAIO JOURNAL, 2021, 67 (05) : E103 - E103
  • [30] Intermittent Aortic Insufficiency as an Aid to Diagnosing Obstruction in a HeartMate II Continuous-Flow Left Ventricular Assist Device
    Pauwaa, Sunil
    Raghuvir, Rashmi
    Kurien, Sudha
    Tatooles, Antone J.
    Pappas, Pat S.
    Bhat, Geetha
    [J]. ASAIO JOURNAL, 2011, 57 (03) : 244 - 246