Numerical analysis of mass transfer in double-pass parallel-plate dialyzers with external recycle

被引:17
|
作者
Yeh, H. M. [1 ]
机构
[1] Tamkang Univ, Dept Chem & Mat Engn, Tamsui 251, Taipei County, Taiwan
关键词
Dialysis; External recycle; Parallel plate; Double pass; Mass transfer; Performance improved; FLOW MEMBRANE MODULES; SOLVENT-EXTRACTION; NEOSEPTA-AFN; DIALYSIS; CHANNEL; ULTRAFILTRATION; EXCHANGERS; TRANSPORT; IONS;
D O I
10.1016/j.compchemeng.2008.12.006
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The effect of external recycle on the performance of dialysis in double-pass rectangular membrane modules was investigated. Theoretical analysis of mass transfer in parallel-flow dialyzers is analogous to heat transfer in parallel-flow heat exchangers. In contrast to a single-pass device without recycle, improvement in mass transfer of about 30% is achievable if parallel-flow dialysis is operated in a double-pass device of same size with external recycle which provides the increase of fluid velocity, resulting in reduction of mass-transfer resistance. (C) 2009 Elsevier Ltd. All rights reserved.
引用
下载
收藏
页码:815 / 821
页数:7
相关论文
共 50 条