Neutron dose in and out of 18 MV photon fields

被引:8
|
作者
Ezzati, A. O. [1 ]
Studenski, M. T. [2 ]
机构
[1] Univ Tabriz, Fac Phys, Tabriz, Iran
[2] Univ Miami, Dept Radiat Oncol, Miami, FL USA
关键词
High energy photons; Neutron contamination dose; Out-of-field dose; Neutron capture gamma; Monte Carlo; RADIATION-THERAPY; ELEKTA LINACS; VARIAN; RADIOTHERAPY; ACCELERATORS; SPECTRA;
D O I
10.1016/j.apradiso.2017.01.046
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In radiation therapy, neutron contamination is an undesirable side effect of using high energy photons to treat patients. Neutron contamination requires adjustments to the shielding requirements of the linear accelerator vault and contributes to the risk of secondary malignancies in patients by delivering dose outside of the primary treatment field. Using MCNPX, an established Monte Carlo code, manufacturer blueprints, and the most up to date ICRP neutron dose conversion factors, the neutron spectra, neutron/photon dose ratio, and the neutron capture gamma ray dose were calculated at different depths and off axis distances in a tissue equivalent phantom. Results demonstrated that the neutron spectra and dose are dependent on field size, depth in the phantom, and off-axis distance. Simulations showed that because of the low neutron absorption cross section of the linear accelerator head materials, the contribution to overall patient dose from neutrons can be up to 1000 times the photon dose out of the treatment field and is also dependent on field size and depth. Beyond 45 cm off axis, the dependence of the neutron dose on field size is minimal. Neutron capture gamma ray dose is also field size dependent and is at a maximum at a depth of about 7 cm. It is important to remember that when treating with high energy photons, the dose from contamination neutrons must be considered as it is much greater than the photon dose.
引用
收藏
页码:186 / 192
页数:7
相关论文
共 50 条
  • [1] Neutron dose measurement in Intensity Modulated Radiotherapy of 18 MV photon beams
    Chern, SS
    Selvaraj, RN
    Sheetz, M
    Wu, A
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1267 - 1268
  • [2] NEUTRON DOSE DELIVERED WHEN USING 18 MV UNFLATTENED PHOTON BEAM
    Vondracek, V.
    Kralik, M.
    Turek, K.
    Navratil, M.
    Kubes, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S476 - S476
  • [3] NEUTRON DOSE WITH 18 MV IMRT
    Vondracek, V.
    Kralik, M.
    Turek, K.
    Navratil, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S189 - S189
  • [4] The out-of-field dose measurement of neutron and photon during radiotherapy with 20MV
    Kruszyna, M.
    Malicki, J.
    Adamczyk, S.
    Kowalik, A.
    Jackowiak, W.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S286 - S286
  • [5] Measurements of secondary neutron dose from 15 MV and 18 MV IMRT
    Howell, RM
    Ferenci, MS
    Hertel, NE
    Fullerton, GD
    Fox, T
    Davis, LW
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 115 (1-4) : 508 - 512
  • [6] Photon and photon-neutron experimental dosimetry in Grid therapy with 18 MV photon beams
    Tajiki, Sareh
    Gholami, Somayeh
    Esfahani, Mahbod
    Rastjoo, Ali
    Hakimi, Amir
    Beheshti, Amir K.
    Meigooni, Ali
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (04) : 466 - 472
  • [7] Calculation of effective dose from measurements of secondary neutron spectra and scattered photon dose from dynamic MLC IMRT for 6 MV, 15 MV, and 18 MV beam energies
    Howell, RM
    Hertel, NE
    Wang, ZL
    Hutchinson, J
    Fullerton, GD
    [J]. MEDICAL PHYSICS, 2006, 33 (02) : 360 - 368
  • [8] Monte Carlo calculation of photo-neutron dose produced by circular cones at 18 MV photon beams
    Hosseinzadeh, Elham
    Banaee, Nooshin
    Nedaie, Hassan Ali
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2018, 23 (01) : 39 - 46
  • [9] Monte Carlo Calculation of Photo-Neutron Dose Produced by Circular Cones at 18 MV Photon Beams
    Banaee, N.
    Hosseinzadeh, E.
    Nedaie, H.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2907 - 2908
  • [10] Monte Carlo calculation of Photo-Neutron dose produced by circular cones at 18 MV photon beams
    Banaee, N.
    Nedaie, H. A.
    Hosseinzade, E.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S986 - S986