Characterization of the impact to PET quantification and image quality of an anterior array surface coil for PET/MR imaging

被引:22
|
作者
Wollenweber, Scott D. [1 ]
Delso, Gaspar [1 ,2 ]
Deller, Timothy [1 ]
Goldhaber, David [1 ]
Huellner, Martin [2 ]
Veit-Haibach, Patrick [2 ]
机构
[1] GE Healthcare, Waukesha, WI 53186 USA
[2] Univ Zurich Hosp, CH-8091 Zurich, Switzerland
关键词
Positron-emission tomography; Magnetic resonance imaging; Attenuation correction; ATTENUATION CORRECTION; TIME;
D O I
10.1007/s10334-013-0388-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Object The aim of this study was to determine the impact to PET quantification, image quality and possible diagnostic impact of an anterior surface array used in a combined PET/MR imaging system. Materials and methods An extended oval phantom and 15 whole-body FDG PET/CT subjects were re-imaged for one bed position following placement of an anterior array coil at a clinically realistic position. The CT scan, used for PET attenuation correction, did not include the coil. Comparison, including liver SUVmean, was performed between the coil present and absent images using two methods of PET reconstruction. Due to the time delay between PET scans, a model was used to account for average physiologic time change of SUV. Results On phantom data, neglecting the coil caused a mean bias of -8.2 % for non-TOF/PSF reconstruction, and -7.3 % with TOF/PSF. On clinical data, the liver SUV neglecting the coil presence fell by -6.1 % (+/- 6.5 %) for non-TOF/PSF reconstruction; respectively-5.2 %(+/- 5.3 %) with TOF/PSF. All FDG-avid features seen with TOF/PSF were also seen with non-TOF/PSF reconstruction. Conclusion Neglecting coil attenuation for this anterior array coil results in a small but significant reduction in liver SUVmean but was not found to change the clinical interpretation of the PET images.
引用
收藏
页码:149 / 159
页数:11
相关论文
共 50 条
  • [1] Characterization of the impact to PET quantification and image quality of an anterior array surface coil for PET/MR imaging
    Scott D. Wollenweber
    Gaspar Delso
    Timothy Deller
    David Goldhaber
    Martin Hüllner
    Patrick Veit-Haibach
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2014, 27 : 149 - 159
  • [2] Impact of flexible body surface coil and patient table on PET quantification and image quality in integrated PET/MR
    Fuerst, S.
    Souvatzoglou, M.
    Martinez-Moeller, A.
    Schwaiger, M.
    Nekolla, S. G.
    Ziegler, S. I.
    NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2014, 53 (03): : 79 - 87
  • [3] Do carotid MR surface coils affect PET quantification in PET/MR imaging?
    Martin J Willemink
    Mootaz Eldib
    Tim Leiner
    Zahi A Fayad
    Venkatesh Mani
    EJNMMI Physics, 2 (Suppl 1)
  • [4] Brain MRI Coil Attenuation Map Processing for the GE SIGNA PET/MR: Impact on PET Image Quantification and Uniformity
    Manavaki, Roido
    Hong, Young T.
    Fryer, Tim D.
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [5] Impact of Point-Spread Function Modeling on PET Image Quality in Integrated PET/MR Hybrid Imaging
    Aklan, Bassim
    Oehmigen, Mark
    Beiderwellen, Karsten
    Ruhlmann, Marcus
    Paulus, Daniel H.
    Jakoby, Bjoern W.
    Ritt, Philipp
    Quick, Harald H.
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57 (01) : 78 - 84
  • [6] PET Image Quality Improvement for Simultaneous PET/MRI with a Lightweight MRI Surface Coil
    Deller, Timothy W.
    Mathew, Nicholas K.
    Hurley, Samuel A.
    Bobb, Chad M.
    McMillan, Alan B.
    RADIOLOGY, 2021, 298 (01) : 166 - 172
  • [7] Impact of MRI surface coils on quantification in integrated PET/MR
    Fuerst, Sebastian
    Souvatzoglou, Michael
    Martinez-Moeller, Axel
    Nekolla, Stephan
    Schwaiger, Markus
    Ziegler, Sibylle
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [8] Evaluation of dedicated PET/MR breast coil in integrated PET/MR system for simultaneous PET/MR breast imaging
    Cho, Ihn Ho
    Kong, Eun-jung
    Chun, KyunghAh
    Jeong, Ju Hye
    Kwon, SangDon
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [9] The effect of patient positioning aids on PET quantification in PET/MR imaging
    Mantlik, Frederic
    Hofmann, Matthias
    Werner, Matthias K.
    Sauter, Alexander
    Kupferschlaeger, Juergen
    Schoelkopf, Bernhard
    Pichler, Bernd J.
    Beyer, Thomas
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 (05) : 920 - 929
  • [10] PET/MR imaging of sarcomas: effect of PET quantification by classification of tissue
    Claes Nøhr Ladefoged
    Adam Espe Hansen
    Kim Francis Andersen
    Annika Loft
    Liselotte Højgaard
    Andreas Kjær
    Flemming Andersen
    EJNMMI Physics, 1 (Suppl 1)