Automated functional imaging in atrioventricular delay time optimisation in patients with dual chamber pacemakers

被引:0
|
作者
Arik, Osman Ziya [1 ]
Kalkan, Gulhan Yuksel [2 ]
Sahin, Durmus Yildiray [2 ]
Tekin, Kamuran [3 ]
Elbasan, Zafer [2 ]
Gur, Mustafa [2 ]
Kuloglu, Osman [2 ]
Gencaslan, Murat [2 ]
Turkoglu, Caner [2 ]
Ucak, Haci Ali [2 ]
Cayli, Murat [2 ]
机构
[1] Gumushane State Hosp, Gumushane, Turkey
[2] Adana Numune Educ & Res Hosp, Adana, Turkey
[3] Batman State Hosp, Batman, Turkey
关键词
pacing; transthoracic echocardiography; strain imaging; ventricular function; B-type natriuretic peptide; COMPLETE HEART-BLOCK; IMPEDANCE CARDIOGRAPHY; VENTRICULAR-FUNCTION; NATRIURETIC-PEPTIDE; DDD PACEMAKER; INTERVAL;
D O I
10.5603/KP.a2014.0045
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Optimisation of atrioventricular (AV) delay time has positive effects on left ventricular (LV) functions in patients with a DDD pacemaker. However, the method used for optimisation is still debated. Aim: To evaluate the effect of different AV delay times on various LV systolic performances by using automated functional imaging (AFI) in patients with a DDD pacemaker and preserved LV systolic function. Methods: The study population consisted of 40 patients with a DDD pacemaker implanted for third degree AV block and preserved LV systolic function (19 men; mean age 64.3 +/- 10.9 years). During each pacing period, blood samples were taken for the measurement of B-type natriuretic peptide (BNP) levels, and telemetric and echocardiographic evaluations were performed to all patients. Also peak systolic global longitudinal strain (PSGLS) was calculated using the AFI method. Results: No significant differences except for LV outflow tract-velocity time integral (LVOT-VTI) were observed in pulse wave Doppler parameters with different AV delay times. PSGLS were better at 150 and 200 ms AV delay times compared to 100 ms (p < 0.001 for 100-150 ms and 100-200 ms). Similarly, LVOT-VTI values were significantly higher at 150 and 200 ms AV delay times compared to 100 ms (for 100-150 ms, p = 0.017 and for 100-200 ms, p = 0.013). Also there was a significant reduction in BNP levels at 150 ms and 200 ms compared to 100 ms AV delay time (for 100-150 ms, p = 0.001, and for 100-200 ms, p < 0.001). Conclusions: In patients with an implanted DDD pacemaker and preserved LV systolic function, increasing AV delay time has beneficial effects on LV systolic performance in the acute phase, as shown by the AFI method in our study.
引用
收藏
页码:640 / 645
页数:6
相关论文
共 50 条
  • [1] Automated Functional Imaging in Atrioventricular Delay Time Optimization in Patients with Dual Chamber Pacemakers
    Arik, Osman Ziya
    Kalkan, Gulhan Yuksel
    Sahin, Durmus Yildiray
    Elbasan, Zafer
    Gur, Mustafa
    Kuloglu, Osman
    Tekin, Kamuran
    Gencaslan, Murat
    Turkoglu, Caner
    Ucak, Haci Ali
    Cayli, Murat
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (18) : C7 - C7
  • [2] OPTIMAL ATRIOVENTRICULAR INTERVAL IN PATIENTS WITH DUAL CHAMBER PACEMAKERS
    MEHTA, D
    GILMOUR, S
    LAU, CP
    WARD, D
    CAMM, J
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (02): : 437 - 437
  • [3] IS THE OPTIMAL ATRIOVENTRICULAR DELAY THE SAME AT REST AND DURING EXERCISE - AN ASSESSMENT IN PATIENTS WITH DUAL CHAMBER PACEMAKERS
    MEHTA, D
    GILMOUR, S
    LAU, CP
    MEHTA, N
    WARD, DE
    CAMM, AJ
    [J]. BRITISH HEART JOURNAL, 1988, 59 (01): : 93 - 93
  • [4] NONINVASIVE ASSESSMENT OF OPTIMAL ATRIOVENTRICULAR DELAY AT REST AND DURING EXERCISE IN PATIENTS WITH DUAL CHAMBER PACEMAKERS
    MEHTA, D
    GILMOUR, S
    CAMM, AJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (04): : 989 - 989
  • [5] IMPORTANCE OF AN INDIVIDUALLY PROGRAMMED ATRIOVENTRICULAR DELAY AT REST AND ON WORK CAPACITY IN PATIENTS WITH DUAL-CHAMBER PACEMAKERS
    FRIELINGSDORF, J
    GERBER, AE
    DUR, P
    VUILLIOMENET, A
    BERTEL, O
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (01): : 37 - 45
  • [6] Echocardiography for Determining the Optimal Atrioventricular Interval in Patients with Dual Chamber Pacemakers
    Sun, Pin
    Zhao, Qing
    Wang, Zhi-Bin
    Li, Jian-Sheng
    Wo, Jin-Shan
    Cai, Shang-Lang
    [J]. CLINICAL CARDIOLOGY, 2009, 32 (08) : 439 - 441
  • [7] NONINVASIVE ASSESSMENT OF OPTIMAL ATRIOVENTRICULAR INTERVAL IN PATIENTS WITH DUAL CHAMBER PACEMAKERS
    MEHTA, D
    GILMOUR, S
    WARD, D
    CAMM, J
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03): : 715 - 715
  • [8] EARLY OPTIMIZATION OF ATRIOVENTRICULAR DELAY IN DUAL CHAMBER PACEMAKERS (DDD) COULD BE A RESOURCE FOR BEST HEMODYNAMICS?
    Statescu, C.
    Sascau, R. A.
    Antoniu, F.
    Georgescu, C. Arsenescu
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2009, 20 : S62 - S62
  • [9] A perspective on atrioventricular delay optimization in patients with a dual chamber pacemaker
    Porciani, MC
    Corbucci, G
    Fantini, F
    Musilli, N
    Sabini, A
    Mlchelucci, A
    Colella, A
    Pieragnoli, P
    Demarchi, G
    Padeletti, L
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2004, 27 (03): : 333 - 338
  • [10] Effect of programmed atrioventricular delay on the patterns of mitral and pulmonary vein flow profiles and their durations in patients with dual-chamber pacemakers
    Pai, SM
    Jacobson, AK
    Pai, RG
    [J]. ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 1999, 16 (03): : 231 - 238