Hepatitis B virus (HBV) genotype determination by the COBAS® AmpliPrep/COBAS® TaqMan® HBV Test, v2.0 in serum and plasma matrices

被引:17
|
作者
Goedel, Sebastian [1 ]
Rullkoetter, Matthias [1 ]
Weisshaar, Sabrina [1 ]
Mietag, Carola [1 ]
Leying, Hermann [1 ]
Boehl, Florian [1 ]
机构
[1] Roche Diagnost Ltd, CH-6343 Rotkreuz, Switzerland
关键词
HBV genotypes; Quantification; COBAS; AmpliPrep; TaqMan; Fully automated; ANTIVIRAL THERAPY; QUANTITATIVE PCR; INFECTION; DNA; STANDARD; ASSAY;
D O I
10.1016/j.jcv.2009.05.021
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Background: Viral load quantification is established in the clinical management of chronic Hepatitis B virus (HBV) infection for assessing efficacies and resistance developments in anti-viral drug treatment. Objectives: The fully automated COBAS (R) AmpliPrep/COBAS (R) TaqMan (R) HBV Test, v2.0 was evaluated for the linear measuring range and the inclusivity of HBV genotype determination in EDTA plasma and serum samples. Study design: Two kit lots of the test were used to determine the linear measuring range as well as linearity and limit of detection applying different concentration levels of specimens representing HBV genotypes A to H along with a pre-core mutant and the WHO Standard. Results: The COBAS (R) AmpliPrep/COBAS (R) TaqMan (R) HBV Test, v2.0 displayed a linear measuring range of seven log(10) steps from 20 IU/mL(lower limit of quantification) to 2.3E+08 IU/mL(upper limit of quantification) yielding similar results for EDTA plasma and serum. Inclusivity was shown by reliable quantification of HBV genotypes A to H at different concentration levels. The >= 95% hit rate LOD was 15 IU/ml for genotypes C, D, F, C, the pre-core mutant and the WHO Standard and 20 IU/ml. for genotypes A, B, E and H matching the test's lower limit of quantification. 95% PROBIT analysis yielded concentrations of 8.9 IU/mL for the WHO Standard and of 6.0-16.4 IU/mL for the genotypes. Conclusions: The COBAS (R) AmpliPrep/COBAS (R) TaqMan (R) HBV Test, v2.0 provides genotype inclusivity for accurate viral load monitoring in serum and EDTA plasma samples and supports clinical routine in the management of HBV infection. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:232 / 236
页数:5
相关论文
共 50 条
  • [21] Comparison of the cobas Ampliprep™/Cobas TaqMan™ HBV test versus the Cobas Amplicor HBV monitor for HBV-DNA detection and quantification during antiviral therapy
    Allice, Tiziano
    Cerutti, Francesco
    Pittaluga, Fabrizia
    Varetto, Silvia
    Gabella, Silvia
    Marzano, Alfredo
    Franchello, Alessandro
    Ghisetti, Valeria
    NEW MICROBIOLOGICA, 2008, 31 (01): : 27 - 35
  • [22] Cobas Ampliprep/Cobas TaqMan HIV-1 v2.0 Assay: Consequences at the Cohort Level
    Taylor, Ninon
    Grabmeier-Pfistershammer, Katharina
    Egle, Alexander
    Greil, Richard
    Rieger, Armin
    Ledergerber, Bruno
    Oberkofler, Hannes
    PLOS ONE, 2013, 8 (08):
  • [23] 国产荧光定量PCR与COBAS AmpliPrep/COBAS TaqMan v 2.0检测HBV DNA结果比较
    倪俊明
    孙梅
    吴旭平
    临床检验杂志, 2014, 32 (05) : 398 - 399
  • [24] Evaluation of the COBAS AmpliPrep-total nucleic acid isolation-COBAS TaqMan hepatitis B virus (HBV) quantitative test and comparison to the VERSANT HBV DNA 3.0 assay
    Ronsin, C
    Pillet, A
    Bali, C
    Denoyel, GA
    JOURNAL OF CLINICAL MICROBIOLOGY, 2006, 44 (04) : 1390 - 1399
  • [25] Differences between quantification of genotype 3 hepatitis C virus RNA by Versions 1.0 and 2.0 of the COBAS AmpliPrep/COBAS TaqMan HCV Test
    Pierce, Virginia M.
    Eversley, Jacqueline S.
    Tran, Thuy K.
    Rosenberg, Eric S.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (07) : 956 - 961
  • [26] Performance evaluation of the DAAN HCV assay for quantification of hepatitis C virus RNA and its comparison with COBAS AmpliPrep/COBAS TaqMan HCV Quantitative Test, v2.0
    He, Yuting
    Wang, Yichong
    Chen, Xuefang
    Huang, Hao
    Deng, Jiankai
    Chen, Peisong
    Bin, Huang
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2020, 34 (07)
  • [27] COBAS® AmpliPrep/COBAS® Taqman® HCV Quantitative Test, Version 2.0: An In Vitro Test for Hepatitis C Virus RNA Quantification
    Emma D. Deeks
    Molecular Diagnosis & Therapy, 2015, 19 : 1 - 7
  • [28] PERFORMANCE CHARACTERISTICS OF THE COBAS AMPLIPREP/COBAS TAQMAN HCV QUANTITATIVE TEST, V2.0 FOR HCV-RNA DETECTION AND QUANTIFICATION
    Vermehren, J.
    Susser, S.
    Schutten, M.
    Fueller, C.
    Perner, D.
    Diepstraten-Pas, S.
    Molenkamp, R.
    Gohl, P.
    Colucci, G.
    Zitzer, H.
    Sarrazin, C.
    JOURNAL OF HEPATOLOGY, 2012, 56 : S364 - S364
  • [29] An in-house real-time polymerase chain reaction: standardisation and comparison with the Cobas Amplicor HBV monitor and Cobas AmpliPrep/Cobas TaqMan HBV tests for the quantification of hepatitis B virus DNA
    de Torres Santos, Ana Paula
    Levi, Jose Eduardo
    Lemos, Marcilio Figueiredo
    Calux, Samira Julien
    Oba, Isabel Takano
    Moreira, Regina Celia
    MEMORIAS DO INSTITUTO OSWALDO CRUZ, 2016, 111 (02): : 134 - 140
  • [30] Development of a Second Version of the Cobas AmpliPrep/Cobas TaqMan Hepatitis C Virus Quantitative Test with Improved Genotype Inclusivity
    Vermehren, Johannes
    Colucci, Giuseppe
    Gohl, Peter
    Hamdi, Nabila
    Abdelaziz, Ahmed Ihab
    Karey, Ursula
    Thamke, Diana
    Zitzer, Heike
    Zeuzem, Stefan
    Sarrazin, Christoph
    JOURNAL OF CLINICAL MICROBIOLOGY, 2011, 49 (09) : 3309 - 3315