Effects of an intra-ventricular assist device on the stroke volume of failing ventricle: Analysis of a mock circulatory system

被引:4
|
作者
Zhu, Shidong [1 ]
Luo, Lin [1 ]
Yang, Bibo [2 ]
Ni, Kai [1 ]
Zhou, Qian [1 ]
Li, Xinghui [1 ]
Wang, Xiaohao [1 ]
机构
[1] Tsinghua Univ, Div Adv Mfg, Grad Sch Shenzhen, Shenzhen 518055, Guangdong, Peoples R China
[2] Capital Med Univ, Beijing Anzhen Hosp, Cardiac Surg Ctr, Beijing 100029, Peoples R China
关键词
Left ventricular assist device; mock circulatory system; pulsatile pump; stroke volume; PISTON BLOOD PUMPS; SUPPORT-SYSTEMS; HEART-FAILURE; FLOW; PULSATILE; COUNTERPULSATION; PERFORMANCE; MODEL;
D O I
10.3233/THC-174752
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
BACKGROUND: A novel intra-ventricular assist device (iVAD) was established as a new pulsatile assist device to address various disadvantages, such as bulky configuration and reduced arterial pulsatility, observed in conventional ventricular assist devices. OBJECTIVE: Analyzed the native left ventricular stroke volume (SV) after iVAD support in vitro. METHODS: The SV of iVAD was examined in a home-designed mock circulatory system (MCS) at different heart rates and drive pressures and the SV of a failure ventricle was examined with iVAD at 75, 90, 120 bpm and 120-180 mmHg drive pressure after iVAD support. Data pertaining to native left ventricular SV before and after iVAD support were compared. RESULTS: The native ventricular SV was improved by iVAD when its drive pressure (DP) was slightly greater than that of the mock system. Conversely, the native ventricular SV was decreased when DP was much greater than that (150 mmHg) of MCS. A high DP had a significant effect on SV. CONCLUSIONS: The proposed device improved the dysfunctional native left ventricular SV when DP of iVAD was slightly greater than that of MCS. However, iVAD reduced the SV when the drive pressure was greater than that of MCS.
引用
收藏
页码:S471 / S479
页数:9
相关论文
共 50 条
  • [41] Ventricular assist device support for failing systemic ventricle in adult patient with congenitally corrected transposition of the great arteries and dextrocardia
    Jurinjak, S. Sandra Jaksic
    Udovicic, M.
    Starcevic, B.
    Stipcevic, M.
    Ivanovic, V.
    Susnjar, D.
    Falak, H.
    Vincelj, J.
    Planinc, M.
    Blazekovic, R.
    Bradic, N.
    Sutlic, Z.
    EUROPEAN JOURNAL OF HEART FAILURE, 2018, 20 : 191 - 192
  • [42] Berlin heart ventricular assist device as a long-term bridge to transplantation in a Fontan patient with failing single ventricle
    Halaweish, I.
    Ohye, R. G.
    Si, M. S.
    PEDIATRIC TRANSPLANTATION, 2015, 19 (08) : E193 - E195
  • [43] Right ventricle pressure-volume loops for monitoring right ventricular function in left ventricular assist device patient
    Carlson, Samuel F.
    Smith, Nathan J.
    Brown, Colton
    Joyce, Lyle D.
    Joyce, David L.
    ARTIFICIAL ORGANS, 2022, 46 (03) : 509 - 512
  • [44] EFFECTS OF INTRA-VENTRICULAR ANGIOTENSIN-II MEDIATED BY SYMPATHETIC NERVOUS-SYSTEM
    FALCON, JC
    PHILLIPS, MI
    HOFFMAN, WE
    BRODY, MJ
    AMERICAN JOURNAL OF PHYSIOLOGY, 1978, 235 (04): : H392 - H399
  • [45] ASSESSMENT OF THE EFFECTS OF A CONTINUOUS FLOW LEFT VENTRICULAR ASSIST DEVICE ON AORTIC VALVE FUNCTION IN A MOCK CIRCULATION
    Tuzun, E.
    Rutten, M. C. M.
    van de Vosse, F. N.
    de Mol, B. A.
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2009, 32 (07): : 424 - 424
  • [46] Perioperative Outcomes in Patients With Failing Single-Ventricle Physiology Undergoing Ventricular Assist Device Placement: A Single Institutional Experience
    Gorbea, Mikel
    Duarte, Ingrid Moreno
    Reisch, Joan S.
    Hernandez, Jennifer
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2022, 36 (12) : 4347 - 4356
  • [47] PHONOCARDIOGRAPHIC STUDY OF SOUNDS PRODUCED BY A CIRCULATORY ASSIST DEVICE - REPORT OF A PATIENT WITH A RUPTURED VENTRICULAR SYSTEM
    CLEMENTS, SD
    WHITAKER, A
    LOGUE, RB
    HURST, JW
    ARCHIVES OF INTERNAL MEDICINE, 1977, 137 (11) : 1619 - 1620
  • [48] 3D printing for preoperative planning and surgical simulation of ventricular assist device implantation in a failing systemic right ventricle
    Miller, Jacob R.
    Singh, Gautam K.
    Woodard, Pamela K.
    Eghtesady, Pirooz
    Anwar, Shafkat
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2020, 14 (06) : E172 - E174
  • [49] A Review of Physiologic Considerations and Challenges in Pediatric Patients With Failing Single-Ventricle Physiology Undergoing Ventricular Assist Device Placement
    Gorbea, Mikel
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2022, 36 (06) : 1756 - 1770
  • [50] Placement of a Continuous-Flow Ventricular Assist Device in the Failing Ventricle of an Adult Patient with Complex Cyanotic Congenital Heart Disease
    Morris, Cullen D.
    Gregoric, Igor D.
    Cooley, Denton A.
    Cohn, William E.
    Loyalka, Pranav
    Frazier, O. H.
    HEART SURGERY FORUM, 2008, 11 (03): : E143 - E144