Ion source with LaB6 hollow cathode for a diagnostic neutral beam injector

被引:16
|
作者
Deichuli, PP
Abdrashitov, GF
Ivanov, AA
Kolmogorov, VV
Mishagin, VV
Shul'zhenko, GI
Stupishin, NV
Beals, D
Granetz, R
机构
[1] Budker Inst Nucl Phys, Novosibirsk 630090, Russia
[2] MIT, Ctr Plasma Fus, Cambridge, MA 02139 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2006年 / 77卷 / 03期
关键词
D O I
10.1063/1.2171754
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this article ion source capable of providing 8 A, 54 keV proton beam in maximum 3 s pulses is described. A general description of the diagnostic injector based on this ion source is also given. The ion beam is extracted and accelerated by a four-electrode multiaperture ion-optical system. In the ion source, hydrogen (deuterium) plasma is generated by an arc discharge plasma box with a hot LaB6 cathode. The plasma jet diverging from a small anode orifice enters the plasma expansion volume with a peripheral multipole magnetic field and after partial reflection from this field forms a uniform plasma emitter of 160 mm in diameter in the plasma grid plane. The plasma box operates with an arc current of 350-700 A, providing more than 75% of full energy specie in the extracted ion beam. The ion beam is subsequently neutralized in a hydrogen target with similar to 50% efficiency and is focused 4 m downstream from the ion source that is provided by a spherical shape of the grids. The beam angular divergence is 0.6 degrees. The ion source is used for plasma diagnostics at Alcator C-mod tokamak. (c) 2006 American Institute of Physics.
引用
收藏
页数:4
相关论文
共 50 条
  • [21] Measurement of LaB6 hollow cathode emitter region temperature and plasma characteristics
    Yuya Oshio
    Shogo Masuyama
    Hiroki Watanabe
    Ikkoh Funaki
    Journal of Electric Propulsion, 4 (1):
  • [22] Plasma plume diagnostics of a LaB6 hollow cathode with triple Langmuir probes
    Becatti, G.
    Pedrini, D.
    Kasoji, B.
    Giusti, N.
    Pieri, L.
    Tellini, C.
    Andreussi, T.
    Paganucci, F.
    Andrenucci, M.
    JOURNAL OF INSTRUMENTATION, 2019, 14
  • [23] 500-A LaB6 Hollow cathode for high power electric thrusters
    Becatti, Giulia
    Goebel, Dan M.
    VACUUM, 2022, 198
  • [24] NEW LAB6 THERMIONIC CATHODE
    HOSOKI, S
    OHTSUKA, M
    OKANO, H
    JOURNAL OF ELECTRON MICROSCOPY, 1975, 24 (03): : 180 - 180
  • [25] Influence of propellant oxygen concentration on LaB6 hollow cathode for electric propulsion
    Yang, Chen
    Sun, Jianghong
    Meng, Tianhang
    Li, Chenying
    Ning, Zhongxi
    Zhang, Yan
    Wang, Yanan
    Yu, Daren
    VACUUM, 2025, 237
  • [26] CHARACTERISTICS OF A LAB6 HOLLOW-CATHODE IN THE GLOW-ARC TRANSITION
    TONEGAWA, A
    TAKAYAMA, K
    PHYSICAL REVIEW A, 1990, 42 (08): : 4898 - 4907
  • [27] The energy spread of a LaB6 cathode operated in the virtual source mode
    Wells, T.
    El-Gomati, M.
    ELECTRON MICROSCOPY AND ANALYSIS GROUP CONFERENCE 2013 (EMAG2013), 2014, 522
  • [28] LaB6 cathode workfunction evaluation
    Katsap, Victor
    2013 IEEE 14TH INTERNATIONAL VACUUM ELECTRONICS CONFERENCE (IVEC), 2013,
  • [29] Numerical Simulations of the Partially Ionized Gas in a 100-A LaB6 Hollow Cathode
    Mikellides, Ioannis G.
    Goebel, Dan M.
    Jorns, Benjamin A.
    Polk, James E.
    Guerrero, Pablo
    IEEE TRANSACTIONS ON PLASMA SCIENCE, 2015, 43 (01) : 173 - 184
  • [30] Discharge and plasma plume characterization of a 100 A-class LaB6 hollow cathode
    Mazouffre, S.
    Joussot, R.
    Vincent, B.
    Tsikata, S.
    JOURNAL OF APPLIED PHYSICS, 2021, 130 (17)