Continuous central venous and pulmonary artery oxygen saturation monitoring in the critically ill

被引:290
|
作者
Reinhart, K
Kuhn, HJ
Hartog, C
Bredle, DL
机构
[1] Univ Jena, Dept Anesthesia & Intens Care Med, D-07747 Jena, Germany
[2] Univ Wisconsin, Dept Kinesiol, Eau Claire, WI 54701 USA
关键词
central venous oxygen saturation; mixed venous oxygen saturation; fiberoptic; oximetry;
D O I
10.1007/s00134-004-2337-y
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective. To compare the course of continuously measured mixed and central venous O-2 saturations in high-risk patients and to evaluate the impact of various factors that might interfere with reflection spectrophotometry. Design and setting. Prospective, descriptive study in the interdisciplinary ICU of a university hospital. Patients. 32 critically ill patients with triple-lumen central vein catheters, including 29 patients requiring pulmonary artery catheterization. Interventions. The accuracy of fiberoptic measurements was assessed by comparison to reference co-oximeter results at regular intervals. We examined the effect on measurement accuracy of physiological variables including hematocrit, hemoglobin, pH, temperature, and the administration of various solutions via central venous catheter. Continuous parallel measurements of SvO(2) and ScvO(2) were performed in patients with each type of catheters over a total observation time of 1097 h. Results. ScvO(2) values were more accurate and stable than in vitro oximeter measurements (r=0.96 from 150 samples, mean difference 0.15%, average drift 0.10%/day) and was not significantly affected by synchronous infusion therapy or by changes in hematocrit, hemoglobin, pH, or temperature. ScvO(2) values closely paralleled SvO(2), whether measured in vitro (r=0.88 from 150 samples) or in vivo (r=0.81 from 395,128 samples) but averaged about 7+/-4 saturation percentage higher. ScvO(2) changed in parallel in 90% of the 1,498 instances in which SvO(2) changed more than 5% (over an average of 43 min). Conclusions. Continuous fiberoptic measurement of central vein O-2 saturation has potential to be a reliable and convenient tool which could rapidly warn of acute change in the oxygen supply/demand ratio of critically ill patients.
引用
收藏
页码:1572 / 1578
页数:7
相关论文
共 50 条
  • [1] Continuous central venous and pulmonary artery oxygen saturation monitoring in the critically ill
    Konrad Reinhart
    Hans-Jörg Kuhn
    Christiane Hartog
    Donald L. Bredle
    [J]. Intensive Care Medicine, 2004, 30 : 1572 - 1578
  • [2] Continuous central venous saturation monitoring in critically ill patients
    D Chiumello
    M Cressoni
    A Marino
    E Gallazzi
    C Mietto
    V Berto
    M Chierichetti
    [J]. Critical Care, 15 (Suppl 1):
  • [3] Continuous monitoring of venous oxygen saturation in critically-ill infants
    Ko, WJ
    Chang, CI
    Chiu, IS
    Chu, SH
    [J]. JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 1996, 95 (03) : 258 - 262
  • [4] CONTINUOUS MONITORING OF MIXED VENOUS OXYGEN-SATURATION IN CRITICALLY ILL PATIENTS
    BAELE, PL
    MCMICHAN, JC
    MARSH, HM
    SILL, JC
    SOUTHORN, PA
    [J]. ANESTHESIA AND ANALGESIA, 1982, 61 (06): : 513 - 517
  • [5] Femoral venous oxygen saturation and central venous oxygen saturation in critically ill patients
    Zhang, Xiaohong
    Wang, Jiandong
    Dong, Yun
    Chen, Youdai
    [J]. JOURNAL OF CRITICAL CARE, 2015, 30 (04) : 768 - 772
  • [6] MONITORING OF CENTRAL VENOUS OXYGEN-SATURATION VERSUS MIXED VENOUS OXYGEN-SATURATION IN CRITICALLY ILL PATIENTS
    MARTIN, C
    AUFFRAY, JP
    BADETTI, C
    PERRIN, G
    PAPAZIAN, L
    GOUIN, F
    [J]. INTENSIVE CARE MEDICINE, 1992, 18 (02) : 101 - 104
  • [7] FEMORAL VENOUS OXYGEN SATURATION AS A SUBSTITUTE OF CENTRAL VENOUS OXYGEN SATURATION IN CRITICALLY ILL PATIENTS
    Marti, Y. N.
    Milena, L. B.
    de Azevedo, R. P.
    Machado, F. R.
    [J]. INTENSIVE CARE MEDICINE, 2013, 39 : S485 - S485
  • [8] UTILITY OF MONITORING CONTINUOUS MIXED VENOUS OXYGEN-SATURATION IN CRITICALLY ILL PATIENTS
    CHELLURI, L
    JASTREMSKI, MS
    BAILLY, RT
    BENEY, KM
    [J]. CHEST, 1986, 89 (06) : S479 - S479
  • [9] Central venous and mixed venous oxygen saturation in critically ill patients
    Ladakis, C
    Myrianthefs, P
    Karabinis, A
    Karatzas, G
    Dosios, T
    Fildissis, G
    Gogas, J
    Baltopoulos, G
    [J]. RESPIRATION, 2001, 68 (03) : 279 - 285
  • [10] CENTRAL VENOUS SATURATION AS A SURROGATE FOR MIXED VENOUS SATURATION DURING MONITORING CRITICALLY ILL CHILDREN
    Hijazi, O. M.
    Jfieh, A.
    Taweel, H.
    Martin, R.
    Loh, W. -L.
    Ismail, S.
    Kabbani, M.
    [J]. INTENSIVE CARE MEDICINE, 2013, 39 : S42 - S43