Hemodynamic evaluation of stentless mitral valve replacement

被引:4
|
作者
Walther, T [1 ]
Falk, V [1 ]
Lehmann, S [1 ]
Walther, C [1 ]
Doll, N [1 ]
Bucerius, J [1 ]
Gummert, J [1 ]
Mohr, FW [1 ]
机构
[1] Univ Leipzig, Klin Herzchirurg, Herzzentrum, D-04289 Leipzig, Germany
关键词
echocardiography; mitral valve replacement; stentless mitral valve; xenograft;
D O I
10.1046/j.1525-1594.2002.06986.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to evaluate the perioperative and postoperative hemodynamic function of recently introduced stentless mitral valves (SMVs). Forty-two selected patients who had received the SMV since 1997 had preoperative, postoperative, and follow-up transthoracic echocardiographic (TTE) examinations. Transesophageal echocardiography (TEE) was applied routinely during the operation as well as postoperatively if indicated. Adequate SMV function was seen in all patients perioperatively. Echocardiographic prediction of valve size and of papillary muscle distances was correct in 37 of 42 patients. Transvalvular blood flow velocities were 1.6+/-0.2 m/s postoperatively versus 1.8+/-0.3 m/s at 1 year and 1.7+/-0.3 m/s at 3-year follow-up. Mean transvalvular gradients were 4.3+/-1.6, 3.8+/-1.9, and 4.7+/-1.1 mm Hg, respectively. TEE and TTE are extremely useful to assist SMV implantation as well as for regular follow-up evaluation. The hemodynamic function after SMV implantation with preservation of the annulo-ventricular continuity is promising.
引用
收藏
页码:847 / 850
页数:4