Model-Based Image Updating for Brain Shift in Deep Brain Stimulation Electrode Placement Surgery

被引:5
|
作者
Li, Chen [1 ]
Fan, Xiaoyao [1 ]
Hong, Jennifer [2 ,3 ]
Roberts, David W. [1 ,2 ,3 ,4 ]
Aronson, Joshua P. [2 ,3 ]
Paulsen, Keith D. [1 ,2 ,4 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[2] Dartmouth Coll, Geisel Sch Med, Hanover, NH 03755 USA
[3] Dartmouth Hitchcock Med Ctr, Sect Neurosurg, Lebanon, NH 03766 USA
[4] Norris Cotton Canc Ctr, Lebanon, NH 03756 USA
关键词
Surgery; Brain modeling; Strain; Mathematical model; Biological system modeling; Satellite broadcasting; Fluids; Biomechanics; brain modeling; computational modeling; deep brain stimulation; image-guided intervention; FUNCTIONAL NEUROSURGERY; REGISTRATION; DEFORMATION; LOCATION; PLATFORM;
D O I
10.1109/TBME.2020.2990669
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: The efficacy of deep brain stimulation (DBS) depends on accurate placement of electrodes. Although stereotactic frames enable co-registration of image-based surgical planning and the operative field, the accuracy of electrode placement can be degraded by intra-operative brain shift. In this study, we adapted a biomechanical model to estimate whole brain displacements from which we deformed preoperative CT (preCT) to generate an updated CT (uCT) that compensates for brain shift. Methods: We drove the deformation model using displacement data derived from deformation in the frontal cortical surface that occurred during the DBS intervention. We evaluated 15 patients, retrospectively, who underwent bilateral DBS surgery, and assessed the accuracy of uCT in terms of target registration error (TRE) relative to a CT acquired post-placement (postCT). We further divided subjects into large (Group L) and small (Group S) deformation groups based on a TRE threshold of 1.6mm. Anterior commissure (AC), posterior commissure (PC) and pineal gland (PG) were identified on preCT and postCT and used to quantify TREs in preCT and uCT. Results: In the group of large brain deformation, average TREs for uCT were 1.11 +/- 0.13 and 1.07 +/- 0.38 mm at AC and PC, respectively, compared to 1.85 +/- 0.17 and 0.92 +/- 0.52 mm for preCT. The model updating approach improved AC localization but did not alter TREs at PC. Conclusion: This preliminary study suggests that our image updating method may compensate for brain shift around surgical targets of importance during DBS surgery, although further investigation is warranted before conclusive evidence will be available. Significance: With further development and evaluation, our model-based image updating method using intraoperative sparse data may compensate for brain shift in DBS surgery efficiently, and have utility in updating targeting coordinates.
引用
收藏
页码:3542 / 3552
页数:11
相关论文
共 50 条
  • [21] Brain shift during deep brain stimulation surgery for Parkinson's disease
    Halpern, Casey H.
    Danish, Shabbar F.
    Baltuch, Gordon H.
    Jaggi, Jurg L.
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2008, 86 (01) : 37 - 43
  • [22] Model-based correction for brain shift in deep brain stimulation burr hole procedures: a comparison using interventional magnetic resonance imaging
    Luo, Ma
    Narasimhan, Saramati
    Martin, Alastair J.
    Larson, Paul S.
    Miga, Michael I.
    MEDICAL IMAGING 2018: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2018, 10576
  • [23] Interventional MR Imaging for Deep-Brain Stimulation Electrode Placement
    Chansakul, Thanissara
    Chen, Paul N., Jr.
    Lee, Thomas C.
    Tierney, Travis
    RADIOLOGY, 2016, 281 (03) : 940 - 946
  • [24] Pre- to post-operative CT image registration to estimate cortical shift for image updating in deep brain stimulation
    Li, Chen
    Fan, Xiaoyao
    Aronson, Joshua
    Paulsen, Keith D.
    MEDICAL IMAGING 2018: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2018, 10576
  • [25] Deep Brain Stimulation: Eye Movements Reveal Anomalous Effects of Electrode Placement and Stimulation
    Antoniades, Chrystalina A.
    Buttery, Philip
    FitzGerald, James J.
    Barker, Roger A.
    Carpenter, Roger H. S.
    Watts, Colin
    PLOS ONE, 2012, 7 (03):
  • [26] Model-based closed-loop control of thalamic deep brain stimulation
    Tian, Yupeng
    Saradhi, Srikar
    Bello, Edward
    Johnson, Matthew D.
    D'Eleuterio, Gabriele
    Popovic, Milos R.
    Lankarany, Milad
    FRONTIERS IN NETWORK PHYSIOLOGY, 2024, 4
  • [27] Cellular effects of deep brain stimulation: Model-based analysis of activation and inhibition
    McIntyre, CC
    Grill, WM
    Sherman, DL
    Thakor, NV
    JOURNAL OF NEUROPHYSIOLOGY, 2004, 91 (04) : 1457 - 1469
  • [28] Model-Based Analysis of Pathway Recruitment During Subthalamic Deep Brain Stimulation
    Bower, Kelsey L.
    Noecker, Angela M.
    Frankemolle-Gilbert, Anneke M.
    Mcintyre, Cameron C.
    NEUROMODULATION, 2024, 27 (03): : 455 - 463
  • [29] Final Electrode Placement Changes from Planned Target Based on Microelectrode Recording Findings in Awake Deep Brain Stimulation Surgery
    Davis, Gregory
    Arguello, Jacquelin
    Patterson, Terrence
    Levine, Zachary
    JOURNAL OF NEUROSURGERY, 2019, 131 (01)
  • [30] Electrode Placement Accuracy in Robot-Assisted Asleep Deep Brain Stimulation
    David VanSickle
    Victoria Volk
    Patricia Freeman
    Jamie Henry
    Meghan Baldwin
    Clare K. Fitzpatrick
    Annals of Biomedical Engineering, 2019, 47 : 1212 - 1222