CT Metal Artifact Reduction in the Spine: Can an Iterative Reconstruction Technique Improve Visualization?

被引:54
|
作者
Kotsenas, A. L. [1 ]
Michalak, G. J. [1 ]
DeLone, D. R. [1 ]
Diehn, F. E. [1 ]
Grant, K. [2 ]
Halaweish, A. F. [2 ]
Krauss, A. [3 ]
Raupach, R. [3 ]
Schmidt, B. [3 ]
McCollough, C. H. [1 ]
Fletcher, J. G. [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[2] Siemens Med Solut, Malvern, PA USA
[3] Siemens Healthcare, Forchheim, Germany
关键词
DUAL-ENERGY CT; COMPUTED-TOMOGRAPHY; VIVO EVALUATION; IMPLANTS;
D O I
10.3174/ajnr.A4416
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: Metal-related artifacts from spine instrumentation can obscure relevant anatomy and pathology. We evaluated the ability of CT images reconstructed with and without iterative metal artifact reduction to visualize critical anatomic structures in postoperative spines and assessed the potential for implementation into clinical practice. MATERIALS AND METHODS: We archived CT projection data in patients with instrumented spinal fusion. CT images were reconstructed by using weighted filtered back-projection and iterative metal artifact reduction. Two neuroradiologists evaluated images in the region of spinal hardware and assigned a score for the visualization of critical anatomic structures by using soft-tissue and bone windows (critical structures totally obscured, n = 0; anatomic recognition with high diagnostic confidence, n = 5). Using bone windows, we measured the length of the most pronounced linear artifacts. For each patient, neuroradiologists made recommendations regarding the optimal use of iterative metal artifact reduction and its impact on diagnostic confidence. RESULTS: Sixty-eight patients met the inclusion criteria. Visualization of critical soft-tissue anatomic structures was significantly improved by using iterative metal artifact reduction compared with weighted filtered back-projection (median,1 +/- 1.5 versus 3 +/- 1.3, P <.001), with improvement in the worst visualized anatomic structure in 88% (60/68) of patients. There was not significant improvement in visualization of critical osseous structures. Linear metal artifacts were reduced from 29 to 11 mm (P <.001). In 87% of patients, neuroradiologists recommended reconstructing iterative metal artifact reduction images instead of weighted filtered back-projection images, with definite improvement in diagnostic confidence in 32% (22/68). CONCLUSIONS: Iterative metal artifact reduction improves visualization of critical soft-tissue structures in patients with spinal hardware. Routine generation of these images in addition to routine weighted filtered back-projection is recommended.
引用
收藏
页码:2184 / 2190
页数:7
相关论文
共 50 条
  • [1] The λ-MLEM algorithm:: An iterative reconstruction technique for metal artifact reduction in CT images
    Oehler, May
    Buzug, Thorsten M.
    ADVANCES IN MEDICAL ENGINEERING, 2007, 114 : 42 - +
  • [2] Iterative CT reconstruction of real data with metal artifact reduction
    Hamelin, Benoit
    Goussard, Yves
    Gendron, David
    Dussault, Jean-Pierre
    Cloutier, Guy
    Beaudoin, Gilles
    Soulez, Gilles
    2008 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, VOLS 1-4, 2008, : 1453 - +
  • [3] Unintended Consequences: Review of New Artifacts Introduced by Iterative Reconstruction CT Metal Artifact Reduction in Spine Imaging
    Wayer, D. R.
    Kim, N. Y.
    Otto, B. J.
    Grayev, A. M.
    Kuner, A. D.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (11) : 1973 - 1975
  • [4] Metal pin artifact reduction with masked iterative nano-CT reconstruction
    Li, Mengzhou
    Cooley, Victoria
    Nikitin, Viktor
    Stock, Stuart R.
    Wang, Ge
    DEVELOPMENTS IN X-RAY TOMOGRAPHY XIV, 2022, 12242
  • [5] Iterative deblurring for CT metal artifact reduction
    Wang, G
    Snyder, DL
    OSullivan, JA
    Vannier, MW
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1996, 15 (05) : 657 - 664
  • [6] ITERATIVE DEBLURRING OF CT IMAGE-RESTORATION, METAL ARTIFACT REDUCTION AND LOCAL RECONSTRUCTION
    WANG, G
    SNYDER, DL
    VANNIER, MW
    RADIOLOGY, 1995, 197 : 291 - 291
  • [7] Imaging of Arthroplasties: Improved Image Quality and Lesion Detection With Iterative Metal Artifact Reduction, a New CT Metal Artifact Reduction Technique
    Subhas, Naveen
    Polster, Joshua M.
    Obuchowski, Nancy A.
    Primak, Andrew N.
    Dong, Frank F.
    Herts, Brian R.
    Iannotti, Joseph P.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 207 (02) : 378 - 385
  • [8] Iterative metal artifact reduction: Evaluation and optimization of technique
    Naveen Subhas
    Andrew N. Primak
    Nancy A. Obuchowski
    Amit Gupta
    Joshua M. Polster
    Andreas Krauss
    Joseph P. Iannotti
    Skeletal Radiology, 2014, 43 : 1729 - 1735
  • [9] Iterative metal artifact reduction: Evaluation and optimization of technique
    Subhas, Naveen
    Primak, Andrew N.
    Obuchowski, Nancy A.
    Gupta, Amit
    Polster, Joshua M.
    Krauss, Andreas
    Iannotti, Joseph P.
    SKELETAL RADIOLOGY, 2014, 43 (12) : 1729 - 1735
  • [10] Metal Artifact Reduction of CT Scans to Improve PET/CT
    van der Vos, Charlotte S.
    Arens, Anne I. J.
    Hamill, James J.
    Hofmann, Christian
    Panin, Vladimir Y.
    Meeuwis, Antoi P. W.
    Visser, Eric P.
    de Geus-Oei, Lioe-Fee
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58 (11) : 1867 - 1872