A systematic study of the pulse height deficit in propane-filled gas ionization detectors

被引:9
|
作者
Weijers, TDM [1 ]
Ophel, TR
Timmers, H
Elliman, RG
机构
[1] Australian Natl Univ, Res Sch Phys Sci & Engn, Dept Elect Mat Engn, Canberra, ACT 0200, Australia
[2] Australian Natl Univ, Res Sch Phys Sci & Engn, Res Sch Phys Sci & Engn, Dept Nucl Phys, Canberra, ACT 0200, Australia
关键词
gas ionization detector; pulse height deficit; ion beam analysis; elastic recoil detection; detector response;
D O I
10.1016/S0168-9002(01)01944-1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The response of gas ionization detectors to heavy ions is species dependent. Relative to a linear calibration based on the response to light ions, heavy ions show a pulse height deficit. Detailed knowledge of this deficit is essential for accurate energy spectroscopy of heavy ions. Precise measurements of the response of a propane-filled gas ionization detector have been performed for a large selection of ions in the ranges Z = 6-79 and E = 0.4-2.6 MeV/amu. Pulse height deficits were determined with respect to the response to carbon ions, and found to be as large as 25% of the incident energy. The deficit was found to depend on ion energy and atomic number. Isotopic effects were also observed. Care was taken to obtain a reliable estimate of the energy loss of the ions in the detector window. This included measurements of the window thickness and shape, and measurements of heavy ion stopping powers for Mylar. Monte-Carlo simulations were carried Out using the code TRIM to evaluate the contribution of elastic collisions of the ions in the detector gas. It was found that after accounting for both window energy loss and elastic collisions, a significant residual pulse height deficit remains. An empirical formula was developed to predict the magnitude of the pulse height deficit, and aid in the accurate energy calibration of gas ionization detectors. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:676 / 688
页数:13
相关论文
共 25 条
  • [1] Origins of the residual pulse height deficit in propane-filled gas ionization detectors
    Weijers-Dall, TDM
    Timmers, H
    Elliman, RG
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 550 (1-2): : 139 - 144
  • [2] Accurate depth profiling through energy-dependent pulse height deficit compensation in gas ionization detectors
    Weijers, TDM
    Timmers, H
    Elliman, RG
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2002, 190 : 397 - 401
  • [3] Effect of adding Ar gas on the pulse height distribution of BF3-filled neutron detectors
    M. Padalakshmi
    A. M. Shaikh
    Pramana, 2008, 71 : 1177 - 1181
  • [4] Effect of adding Ar gas on the pulse height distribution of BF3-filled neutron detectors
    Padalakshmi, M.
    Shaikh, A. M.
    PRAMANA-JOURNAL OF PHYSICS, 2008, 71 (05): : 1177 - 1181
  • [5] A STATISTICAL STUDY OF GAS CHROMATOGRAPHIC SYSTEMS EMPLOYING FLAME IONIZATION DETECTORS
    FOLMER, OF
    HAASE, DJ
    ANALYTICA CHIMICA ACTA, 1969, 48 (01) : 63 - &
  • [6] SYSTEMATIC MEASUREMENTS OF PULSE-HEIGHT DEFECTS FOR HEAVY-IONS IN SURFACE-BARRIER DETECTORS
    OGIHARA, M
    NAGASHIMA, Y
    GALSTER, W
    MIKUMO, T
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 251 (02): : 313 - 320
  • [7] Basic study on pulse height distribution of DOI detectors constructed of stacked crystal elements
    Umehara, T
    Murayama, H
    Omura, T
    Ishibashi, H
    Kawai, H
    Inadama, N
    Kasahara, T
    Orita, N
    Tsuda, T
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1429 - 1433
  • [8] THE PRECISION OF AREA AND HEIGHT MEASUREMENTS WITH FLAME IONIZATION DETECTORS IN TEMPERATURE-PROGRAMMED CAPILLARY GAS-CHROMATOGRAPHY
    PARK, JH
    HUSSAM, A
    COUASNON, P
    CARR, PW
    MICROCHEMICAL JOURNAL, 1987, 35 (02) : 232 - 239
  • [9] Inter-pulse dynamics at high repetition rates by femtosecond ionization in gas-filled fibres
    Koettig, Felix
    Koehler, Johannes R.
    Tani, Francesco
    Russell, Philip St J.
    2019 CONFERENCE ON LASERS AND ELECTRO-OPTICS EUROPE & EUROPEAN QUANTUM ELECTRONICS CONFERENCE (CLEO/EUROPE-EQEC), 2019,
  • [10] SYSTEMATIC MEASUREMENTS OF PULSE-HEIGHT DEFECTS FOR HEAVY-IONS IN SURFACE-BARRIER DETECTORS - A COMMENT
    FINCH, EC
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 257 (02): : 381 - 383