Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging

被引:37
|
作者
Savini, Giacomo [1 ]
Taroni, Leonardo [2 ]
Schiano-Lomoriello, Domenico [1 ]
Hoffer, Kenneth J. [3 ,4 ]
机构
[1] IRCCS GB, Bietti Fdn, Rome, Italy
[2] Univ Bologna, S Orsola Malpighi Univ Hosp, Bologna, Italy
[3] Univ Calif Los Angeles, Stein Eye Inst, Los Angeles, CA USA
[4] St Marys Eye Ctr, Santa Monica, CA USA
关键词
LENS POWER CALCULATION; OPTICAL COHERENCE TOMOGRAPHY; PREDICTION ACCURACY; REPRODUCIBILITY; BIOMETER; TOPOGRAPHY; AGREEMENT; POSTERIOR; CAMERA; CURVATURE;
D O I
10.1038/s41433-020-01245-8
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objectives We aimed (1) to assess the repeatability of Total Keratometry (TK) and standard keratometry (K) measurements, as provided by the IOLMaster 700 (Carl Zeiss Meditec), and (2) to compare the corneal astigmatism measured by TK to the total corneal astigmatism (TCA) measured by a Scheimpflug camera (Pentacam AXL, Oculus). Methods Two groups of patients were prospectively enrolled: Group A included previously unoperated eyes undergoing cataract surgery, and Group B eyes with previous myopic corneal excimer laser surgery. TK and K were measured three times by the same examiner. Repeatability was assessed based on the within-subject standard deviation (S-w), test-retest variability, coefficient of variation and intraclass correlation coefficient (ICC). In Group A, TCA was measured once and compared to TK astigmatism. Vector analysis was performed according to N AE ser. Results In Group A (69 eyes) the mean K and TK were, respectively, 43.14 +/- 1.37 D and 43.18 +/- 1.37 D. In Group B (51 eyes) the mean K and TK were, respectively, 40.14 +/- 2.20 D and 39.71 +/- 2.35 D. The repeatability of the average K and TK was high (S-w < 0.10D). All measurements revealed an ICC > 0.9. For most measurements the variance of K and TK did not show any statistically significant difference either within groups or between groups. Vectors KP(45) were significantly different between TK astigmatism and TCA. Conclusions TK measurements offer high repeatability in unoperated and post-excimer laser surgery eyes. TK astigmatism and TCA measurements could not be considered interchangeable.
引用
收藏
页码:307 / 315
页数:9
相关论文
共 50 条
  • [1] Repeatability of total Keratometry and standard Keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpflug imaging
    Giacomo Savini
    Leonardo Taroni
    Domenico Schiano-Lomoriello
    Kenneth J. Hoffer
    [J]. Eye, 2021, 35 : 307 - 315
  • [2] Predicting Postoperative Astigmatism Using Scheimpflug Keratometry (Pentacam) and Automated Keratometry (IOLMaster)
    Park, Choul Yong
    Do, Jae Rock
    Chuck, Roy S.
    [J]. CURRENT EYE RESEARCH, 2012, 37 (12) : 1091 - 1098
  • [3] Comparing Total Keratometry Measurement on the IOLMaster 700 With Goggin Nomogram Adjusted Anterior Keratometry
    LaHood, Benjamin R.
    Goggin, Michael
    Beheregaray, Simone
    Andrew, Nicholas H.
    Esterman, Adrian
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2018, 34 (08) : 521 - 526
  • [4] Assessing Total Keratometry Astigmatism, Simulated Keratometry, and Total Corneal Topographic Astigmatism Against Manifest Refractive Cylinder
    Alpins, Noel
    Ong, James K. Y.
    Stamatelatos, George
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2021, 37 (03) : 198 - 201
  • [5] Comparison of refractive outcomes using Scheimpflug Holladay equivalent keratometry or IOLMaster 700 keratometry for IOL power calculation
    Mustafa Aksoy
    Leyla Asena
    Sirel Gür Güngör
    Ali Küçüködük
    Ahmet Akman
    [J]. International Ophthalmology, 2021, 41 : 2205 - 2212
  • [6] Comparison of refractive outcomes using Scheimpflug Holladay equivalent keratometry or IOLMaster 700 keratometry for IOL power calculation
    Aksoy, Mustafa
    Asena, Leyla
    Gungor, Sirel Gur
    Kucukoduk, Ali
    Akman, Ahmet
    [J]. INTERNATIONAL OPHTHALMOLOGY, 2021, 41 (06) : 2205 - 2212
  • [7] Agreement of Total Keratometry and Posterior Keratometry Among IOLMaster 700, CASIA2, and Pentacam
    Jin, Aixia
    Han, Xiaotong
    Zhang, Jiaqing
    Qiu, Xiaozhang
    Zhang, Yifan
    Qu, Bo
    Tan, Xuhua
    Luo, Lixia
    [J]. TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2023, 12 (03):
  • [8] Comparison of corneal astigmatism and axis location in cataract patients measured by total corneal power, automated keratometry, and simulated keratometry
    Srivannaboon, Sabong
    Soeharnila
    Chirapapaisan, Chareenun
    Chonpimai, Pratuangsri
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (12): : 2088 - 2093
  • [9] Comparison of Standard and Total Keratometry Astigmatism Measured with three Different Technologies
    Carreras, Humberto
    Garcia, Angel G.
    Pinero, David P.
    [J]. OPEN OPHTHALMOLOGY JOURNAL, 2020, 14 : 59 - 66
  • [10] Anterior Surface-Based Keratometry Compared With Scheimpflug Tomography-Based Total Corneal Astigmatism
    Tonn, Bastian
    Klaproth, Oliver Klaus
    Kohnen, Thomas
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (01) : 291 - 298