Analytical performance of the Gem® Premier™ 4000-a comparison study

被引:0
|
作者
Vlah, Snjezana Hrabric [1 ]
Dvornik, Stefica [1 ]
Grdovic, Dino [1 ]
机构
[1] Rijeka Clin Hosp Ctr, Dept Clin Lab Diagnost, Rijeka 51000, Croatia
关键词
point of care testing; Gem (R) Premier (TM) 4000; hemoglobin; POINT-OF-CARE; BLOOD-GAS; ELECTROLYTE; ANALYZER;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objective: The aim of this study was to compare a new point-of-care testing (POCT) Gem (R) Premier (TM) 4000 analyzer with reference analyzers. Special attention was given to the technology of measuring the concentration of hemoglobin and hematocrit on POCT analyzers. Materials and methods:Analytical performance of the Gem (R) Premier (TM) 4000 was compared with Ciba Corning 865, Gem (R) Premier (TM) 3000 and Dimension RxL analyzers for pH, pCO(2), pO(2), electrolytes, metabolites and for hemoglobin/hematocrit. Within-run and between-run imprecisions have been determined for the Gem (R) Premier (TM) 4000 analyzer. In the period of 30 days, 200 whole blood samples were analyzed. Results: Satisfactory results were obtained for within-run imprecision for all study analytes (CV <= 4.89%), except for pO(2) (CV = 14.66%) and for between-run imprecision for all parameters (CV <= 4.41%) except for lactate (Level I :CV = 8%, Level II :CV = 6.59%). The correlation with the comparative methods was satisfactory, correlation coefficients were between 0.9458-0.9948 for most parameters, except for sodium r = 0.8635. For hemoglobin, correlation coefficients were excellent, r = 0.9923(Gem (R) Premier (TM) 4000vs. Ciba Corning 865) and r = 0.9759 (Gem (R) Premier (TM) 4000 vs. Gem (R) Premier (TM) 3000). Passing Bablok regression analysis of analyte concentrations showed good compatibility except for sodium (Gem (R) Premier (TM) 4000 vs. Ciba Corning 865; y = 34.7500 + 0.7500x) and hemoglobin (Gem (R) Premier (TM) 4000vs. Gem (R) Premier (TM) 3000; y = 23.6923 + 0.8308x). Conclusions: Results of analyses obtained by the comparative measurement of analytes on Gem (R) Premier (TM) 4000 and reference analyzers were compatible. In determining hemoglobin concentration, Gem (R) Premier (TM) 4000 is completely comparable with Ciba Corning 865 analyzer, both analyzers using the same co-oximetry method.
引用
收藏
页码:192 / 198
页数:7
相关论文
共 50 条
  • [31] Analytical Study and Performance Comparison of mm-Wave CMOS LNAs
    Malignaggi, Andrea
    Hamidian, Amin
    Shu, Ran
    Kamal, Ali M.
    Boeck, Georg
    2013 8TH EUROPEAN MICROWAVE INTEGRATED CIRCUITS CONFERENCE (EUMIC), 2013, : 260 - 263
  • [32] STUDY ON THE ANALYTICAL PREDICTION OF PREMIER CHIPPING IN THE INVOLUTE GEAR-CUTTING PROCESS
    KIM, JD
    KIM, JG
    LEE, ES
    JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 1994, 41 (01) : 23 - 46
  • [33] GEM POCKETS IN PEGMATITES-A WORLDWIDE STUDY AND COMPARISON
    Lyckberg, Peter
    GEMS & GEMOLOGY, 2011, 47 (02): : 111 - 112
  • [34] Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room
    Jan Zatloukal
    Jiri Pouska
    Jakub Kletecka
    Richard Pradl
    Jan Benes
    Journal of Clinical Monitoring and Computing, 2016, 30 : 949 - 956
  • [35] Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room
    Zatloukal, Jan
    Pouska, Jiri
    Kletecka, Jakub
    Pradl, Richard
    Benes, Jan
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2016, 30 (06) : 949 - 956
  • [36] Simulation study of the performance of quadruple-GEM detectors
    Yue Wang
    Licheng Zhang
    Aera Jung
    Dayong Wang
    Yong Ban
    Radiation Detection Technology and Methods, 2023, 7 : 107 - 116
  • [37] Simulation study of the performance of quadruple-GEM detectors
    Wang, Yue
    Zhang, Licheng
    Jung, Aera
    Wang, Dayong
    Ban, Yong
    RADIATION DETECTION TECHNOLOGY AND METHODS, 2023, 7 (01) : 107 - 116
  • [38] Study of the Performance of an Optically Readout Triple-GEM
    Marafini, Michela
    Patera, Vincenzo
    Pinci, Davide
    Sarti, Alessio
    Sciubba, Adalberto
    Torchia, Natalia Maria
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2018, 65 (01) : 604 - 608
  • [39] COMPARISON OF THE PERFORMANCE OF OPTICAL-SYSTEMS IN FIBER LINKS 4000 KM LONG
    MATERA, F
    SETTEMBRE, M
    FIBER AND INTEGRATED OPTICS, 1995, 14 (02) : 109 - 119
  • [40] Performance of BlackglasTM composites in 4000-hour oxidation study
    AlliedSignal Inc, Des Plaines, United States
    Ceram Eng Sci Proc, 4 (411-420):