Analytical performance of the Gem® Premier™ 4000-a comparison study

被引:0
|
作者
Vlah, Snjezana Hrabric [1 ]
Dvornik, Stefica [1 ]
Grdovic, Dino [1 ]
机构
[1] Rijeka Clin Hosp Ctr, Dept Clin Lab Diagnost, Rijeka 51000, Croatia
关键词
point of care testing; Gem (R) Premier (TM) 4000; hemoglobin; POINT-OF-CARE; BLOOD-GAS; ELECTROLYTE; ANALYZER;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objective: The aim of this study was to compare a new point-of-care testing (POCT) Gem (R) Premier (TM) 4000 analyzer with reference analyzers. Special attention was given to the technology of measuring the concentration of hemoglobin and hematocrit on POCT analyzers. Materials and methods:Analytical performance of the Gem (R) Premier (TM) 4000 was compared with Ciba Corning 865, Gem (R) Premier (TM) 3000 and Dimension RxL analyzers for pH, pCO(2), pO(2), electrolytes, metabolites and for hemoglobin/hematocrit. Within-run and between-run imprecisions have been determined for the Gem (R) Premier (TM) 4000 analyzer. In the period of 30 days, 200 whole blood samples were analyzed. Results: Satisfactory results were obtained for within-run imprecision for all study analytes (CV <= 4.89%), except for pO(2) (CV = 14.66%) and for between-run imprecision for all parameters (CV <= 4.41%) except for lactate (Level I :CV = 8%, Level II :CV = 6.59%). The correlation with the comparative methods was satisfactory, correlation coefficients were between 0.9458-0.9948 for most parameters, except for sodium r = 0.8635. For hemoglobin, correlation coefficients were excellent, r = 0.9923(Gem (R) Premier (TM) 4000vs. Ciba Corning 865) and r = 0.9759 (Gem (R) Premier (TM) 4000 vs. Gem (R) Premier (TM) 3000). Passing Bablok regression analysis of analyte concentrations showed good compatibility except for sodium (Gem (R) Premier (TM) 4000 vs. Ciba Corning 865; y = 34.7500 + 0.7500x) and hemoglobin (Gem (R) Premier (TM) 4000vs. Gem (R) Premier (TM) 3000; y = 23.6923 + 0.8308x). Conclusions: Results of analyses obtained by the comparative measurement of analytes on Gem (R) Premier (TM) 4000 and reference analyzers were compatible. In determining hemoglobin concentration, Gem (R) Premier (TM) 4000 is completely comparable with Ciba Corning 865 analyzer, both analyzers using the same co-oximetry method.
引用
收藏
页码:192 / 198
页数:7
相关论文
共 50 条
  • [1] Analytical performance evaluation of the new GEM® Premier™ 5000 analyzer in comparison to the GEM® Premier™ 4000 and the RapidPoint® 405 systems
    Mion, Monica Maria
    Bragato, Giada
    Zaninotto, Martina
    Alessandroni, Jhessica
    Bernardini, Sergio
    Plebani, Mario
    CLINICA CHIMICA ACTA, 2018, 486 : 313 - 319
  • [2] THEORETICAL-STUDY OF 4000-A SYSTEM OF CLCF
    SO, SP
    JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS II, 1979, 75 : 820 - 821
  • [3] Analytical Performance Evaluation of the New GEM® Premier™ 5000 in Comparison to the Epoc® Blood Gas Analyzer in Horses
    Sandersen, Charlotte
    Dmitrovic, Petra
    Dupont, Julien
    Cesarini, Carla
    Guyot, Hugues
    Serteyn, Didier
    Kirsch, Katharina
    VETERINARY SCIENCES, 2023, 10 (02)
  • [4] Clinical evaluation of the GEM® premier 4000 critical care analyzer
    Blanco, P.
    Romero, C.
    Rodriguez, C.
    Molina, M. J.
    Mantecas, J.
    Cervera, I.
    CLINICAL CHEMISTRY, 2007, 53 (06) : A177 - A177
  • [5] Total bilirubin assay on GEM premier 4000 critical care analyzer
    DeAbreu, M.
    Pamidi, P.
    Mansouri, S.
    Kim, D.
    Strempek, G.
    CLINICAL CHEMISTRY, 2007, 53 (06) : A214 - A214
  • [6] Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®
    Li Wang
    Arianne Y. K. Albert
    Benjamin Jung
    Keyvan Hadad
    Martha E. Lyon
    Melanie Basso
    BMC Pediatrics, 17
  • [7] Limitations and opportunities of whole blood bilirubin measurements by GEM premier 4000®
    Wang, Li
    Albert, Arianne Y. K.
    Jung, Benjamin
    Hadad, Keyvan
    Lyon, Martha E.
    Basso, Melanie
    BMC PEDIATRICS, 2017, 17
  • [8] Analytical properties of the GEM Premier 3000 analyzer evaluated
    Vukelic, Nada
    Futac, Daljia Papic
    Topic, Elizabeta
    BIOCHEMIA MEDICA, 2007, 17 (02) : 231 - 241
  • [9] Sulfhemoglobin dtetecion and correction to CO-Oximetry measurements on GEM premier 4000
    Pamidi, P.
    Mansouri, S.
    DeAbreu, M.
    Kim, D.
    CLINICAL CHEMISTRY, 2007, 53 (06) : A221 - A221
  • [10] Development of a creatinine biosensor for the GEM Premier™ 4000 critical care analyzer.
    Shin, M
    Xu, C
    Conant, J
    Mansouri, S
    D'Orazio, P
    CLINICAL CHEMISTRY, 2004, 50 (06) : A170 - A170