Bone stress for a mini-implant close to the roots of adjacent teeth-3D finite element analysis

被引:59
|
作者
Motoyoshi, M. [1 ]
Ueno, S. [1 ]
Okazaki, K. [1 ]
Shimizu, N. [1 ]
机构
[1] Nihon Univ, Sch Dent, Dept Orthodont, Div Clin Res,Dent Res Ctr,Chiyoda Ku, Tokyo 1018310, Japan
关键词
temporary anchorage devices; anchorage; finite element method; bone resorption; POSTERIOR REGION; PLACEMENT TORQUE; TITANIUM SCREWS; ANCHORAGE; THICKNESS; STABILITY; FAILURE;
D O I
10.1016/j.ijom.2009.02.011
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
This study aimed to evaluate stress in the bone when an orthodontic mini-implant is close to the roots of adjacent teeth using finite element models (FEMs), and to investigate the causes of the high implant failure rate in the mandible. Four FEMs were used: the implant touches nothing; the implant touches the surface of the periodontal membrane; part of the screw thread is embedded in the periodontal membrane; and the implant touches the root. The effect of cortical bone thickness was evaluated using values of 1, 2 and 3 mm. Maximum stress value and stress distribution on the bone elements was determined. Maximum stress on the bone increased when the mini-implant was close to the root. When the implant touched the root, stress increased to 140 MPa or more, and bone resorption could be predicted. Stress was higher for a cortical bone thickness of 2 mm than for other thicknesses. Cortical bone 2 mm thick had a higher risk for bone resorption. A mandible with an average cortical bone thickness of 2 mm may have a higher risk for implant loosening than a maxilla with the same degree of root proximity, which may be related to the lower success rate in the mandible.
引用
收藏
页码:363 / 368
页数:6
相关论文
共 50 条
  • [1] Factorial analysis of variables affecting bone stress adjacent to the orthodontic anchorage mini-implant with finite element analysis
    Lin, Ting-Sheng
    Tsai, Feng-De
    Chen, Chih-Yu
    Lin, Li-Wen
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2013, 143 (02) : 182 - 189
  • [2] Effects of washer on the stress distribution of mini-implant A finite element analysis
    Jang, Hyoung-Jun
    Kwon, Soon-Yong
    Kim, Seong-Hun
    Park, Young-Guk
    Kim, Su-Jung
    ANGLE ORTHODONTIST, 2012, 82 (01) : 137 - 144
  • [3] A Three-Dimensional Finite Element Analysis of the Stress Distribution Around the Bone Mini-Implant Interface Based on the Mini-Implant Angle of Insertion, Diameter, and Length
    Sarika, K.
    Kumaran, Navaneethakrishnan Kurunji
    Seralathan, Sakthidaran
    Sathishkumar, Rathinavel Kumar
    Preethi, Somasundaram Kanmani
    JOURNAL OF PHARMACY AND BIOALLIED SCIENCES, 2023, 15 : S535 - S539
  • [4] Finite element analysis of mono- and bicortical mini-implant stability
    Holberg, Christof
    Winterhalder, Philipp
    Rudzki-Janson, Ingrid
    Wichelhaus, Andrea
    EUROPEAN JOURNAL OF ORTHODONTICS, 2014, 36 (05) : 550 - 556
  • [5] Three-dimensional finite element analysis of the stability of mini-implants close to the roots of adjacent teeth upon application of bite force
    Zhou, Guanjun
    Zhang, Xiaoli
    Qie, Hui
    Li, Chenxi
    Lu, Lin
    Shan, Lihua
    DENTAL MATERIALS JOURNAL, 2018, 37 (05) : 851 - 857
  • [6] Biomechanical behaviour of implant prostheses and adjacent teeth according to bone quality: A finite element analysis
    Wang, Lan
    Fu, Zhi-Hui
    Hu, Zhi-Hui
    Li, Min
    Qiu, Li-Hua
    Gao, Zhi
    EUROPEAN JOURNAL OF ORAL SCIENCES, 2022, 130 (03)
  • [7] Effects of monocortical and bicortical mini-implant anchorage on bone-borne palatal expansion using finite element analysis
    Lee, Robert J.
    Moon, Won
    Hong, Christine
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2017, 151 (05) : 887 - 897
  • [8] Stress distribution patterns in mini-implant and bone in the infra-zygomatic crest region at different angulations: A finite element study
    Paul, Paulin
    Mathur, Anirudh K.
    Chitra, Prasad
    JOURNAL OF THE WORLD FEDERATION OF ORTHODONTISTS, 2021, 10 (01) : 29 - 34
  • [9] Finite element analysis of mini implant biomechanics on peri-implant bone
    Hisam, Muhammad Juzaili
    Zaman, Izzuddin
    Shari, Safian
    Kurniawan, Denni
    Nor, Fethma M.
    DIGITAL MANUFACTURING TRANSFORMING INDUSTRY TOWARDS SUSTAINABLE GROWTH, 2019, 30 : 308 - 314
  • [10] Analysis on the stress of the bone surrounding mini-implant with different diameters and lengths under torque
    Lu, Yingjuan
    Chang, Shaohai
    Ye, Jiantao
    Ye, Yushan
    Yu, Yansong
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2015, 26 : S541 - S545