Evaluation of interarm blood pressure differences using the Microlife WatchBP Office in a clinical setting

被引:8
|
作者
Krogager, Christoffer [1 ,4 ,5 ]
Laugesen, Esben [1 ,2 ,5 ]
Rossen, Niklas B. [1 ]
Poulsen, Per L. [1 ]
Erlandsen, Mogens [3 ]
Hansen, Klavs W. [4 ]
机构
[1] Aarhus Univ Hosp, Dept Endocrinol & Internal Med, Noerrebrogade 44, DK-8000 Aarhus C, Denmark
[2] Aarhus Univ, Dept Clin Med, Aarhus, Denmark
[3] Aarhus Univ, Dept Publ Hlth, Sect Biostat, Aarhus, Denmark
[4] Silkeborg Reg Hosp, Diagnost Ctr, Silkeborg, Denmark
[5] Odense Univ Hosp, Danish Diabet Acad, Odense, Denmark
关键词
blood pressure; blood pressure measurements; interarm difference; GUIDELINES; HYPERTENSION; MANAGEMENT; SOCIETY; PREVALENCE; DEVICE;
D O I
10.1097/MBP.0000000000000246
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective The aim of this study was to evaluate the usefulness of Microlife WatchBP Office and the effect of increasing the number of measurements in the clinical evaluation of systolic interarm difference (IAD). Patients and methods Office blood pressure was measured simultaneously on both arms in 339 patients (85% diabetic) using the Microlife WatchBP Office, a fully automatic, oscillometric device. The patients included were all scheduled for ambulatory blood pressure measurement at the outpatient clinic of endocrinology at Silkeborg Regional Hospital, Denmark. Two successive sets of three individual measurements were made. A statistical analysis of variance was carried out on the measurements. Results In the first set of measurements, the mean IAD was-0.3 mmHg and the prevalence of IAD greater than or equal to 10 mmHg was 9.1%. Only 7.6% of the patients with an IAD less than 10 mmHg in the first set of measurements had an IAD greater than or equal to 10 mmHg in the second set of measurements. The 95% limits of agreement for the mean IAD for a single set of three measurements were +/- 13.16 mmHg. The probability of detecting an IAD more than 10 mmHg only increased slightly with an increasing number of measurements. Conclusion A single set of triplicate measurements using Microlife WatchBP is an acceptable method for evaluating IAD as more measurements do little to improve the probability of detecting an IAD more than 10 mmHg because of high intraindividual variation. Copyright (C) 2017 Wolters Kluwer Health, Inc. All rights reserved.
引用
收藏
页码:161 / 165
页数:5
相关论文
共 50 条
  • [1] Validation of the Microlife WatchBP Office professional device for office blood pressure measurement according to the International Protocol
    Stergiou, G.
    Tzamouranis, D.
    Protogerou, A.
    Nasothimiou, E.
    Kapralos, C.
    [J]. JOURNAL OF HYPERTENSION, 2008, 26 : S481 - S481
  • [2] Automated device that complies with current guidelines for office blood pressure measurement: design and pilot application study of the Microlife WatchBP Office device
    Stergiou, George S.
    Lin, Che-Wei
    Lin, Chia-Ming
    Chang, Shih-Lung
    Protogerou, Athanasios D.
    Tzamouranis, Dimitris
    Nasothimiou, Efthimia
    Tan, Ty-Minh
    [J]. BLOOD PRESSURE MONITORING, 2008, 13 (04) : 231 - 235
  • [3] An automated device that complies with current guidelines for office blood pressure measurement: Design and pilot application study of the Microlife WatchBP Office device
    Stergiou, G.
    Lin, C. -W.
    Lin, C. -M.
    Chang, S. -L.
    Protogerou, A.
    Nasothimiou, E.
    Tzamouranis, D.
    Tan, T. -M.
    [J]. JOURNAL OF HYPERTENSION, 2008, 26 : S495 - S495
  • [4] Practical Suitability of a Stand-Alone Oscillometric Central Blood Pressure Monitor: A Review of the Microlife WatchBP Office Central
    Verberk, Willem J.
    Cheng, Hao-Min
    Huang, Li-Chih
    Lin, Chia-Ming
    Teng, Yao-Pin
    Chen, Chen-Huan
    [J]. PULSE, 2015, 3 (3-4) : 205 - 216
  • [5] An evaluation of interarm differences in blood pressure using both simultaneous and sequential measurements
    Altunkan, S.
    Genc, Y.
    Iliman, N.
    Kayaturk, N.
    Altunkan, E.
    [J]. JOURNAL OF HYPERTENSION, 2007, 25 : S304 - S304
  • [6] Measurement Accuracy of a Stand-Alone Oscillometric Central Blood Pressure Monitor: A Validation Report for Microlife WatchBP Office Central
    Cheng, Hao-Min
    Sung, Shih-Hsien
    Shih, Yuan-Ta
    Chuang, Shao-Yuan
    Yu, Wen-Chung
    Chen, Chen-Huan
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 2013, 26 (01) : 42 - 50
  • [7] Blood pressure - Assessment of interarm differences
    Singer, AJ
    Hollander, JE
    [J]. ARCHIVES OF INTERNAL MEDICINE, 1996, 156 (17) : 2005 - 2008
  • [8] Validation of the Microlife WatchBP Home blood pressure device in pregnancy for medium and large arm circumferences
    Clark, Katherine
    Snowball, Olivia
    Nzelu, Diane
    Kay, Polly
    Kametas, Nikos A.
    [J]. BLOOD PRESSURE MONITORING, 2018, 23 (03) : 171 - 174
  • [9] Interarm differences in blood pressure. Are they reproducible?
    Svraka, M. S.
    Hornnes, N. H.
    Kruuse, C. R. K.
    [J]. CEREBROVASCULAR DISEASES, 2013, 35 : 496 - 496
  • [10] Measuring interarm blood pressure differences in diabetes
    Parati, Gianfranco
    Zanchetti, Alberto
    [J]. NATURE REVIEWS ENDOCRINOLOGY, 2014, 10 (07) : 387 - 388