Comparison and Combination of Dual-Energy- and Iterative-Based Metal Artefact Reduction on Hip Prosthesis and Dental Implants

被引:78
|
作者
Bongers, Malte N. [1 ]
Schabel, Christoph [1 ]
Thomas, Christoph [2 ]
Raupach, Rainer [3 ]
Notohamiprodjo, Mike [1 ]
Nikolaou, Konstantin [1 ]
Bamberg, Fabian [1 ]
机构
[1] Univ Hosp Tubingen, Dept Diagnost & Intervent Radiol, D-72076 Tubingen, Germany
[2] Univ Hosp Dusseldorf, Dept Diagnost & Intervent Radiol, D-40225 Dusseldorf, Germany
[3] Siemens AG, Healthcare Sect, D-91301 Forchheim, Germany
来源
PLOS ONE | 2015年 / 10卷 / 11期
关键词
COMPUTED-TOMOGRAPHY; ORTHOPEDIC IMPLANTS; CT; RECONSTRUCTIONS; IMAGES;
D O I
10.1371/journal.pone.0143584
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose To compare and combine dual-energy based and iterative metal artefact reduction on hip prosthesis and dental implants in CT. Material and Methods A total of 46 patients (women: 50%, mean age: 63+/-15years) with dental implants or hip prostheses (n = 30/20) were included and examined with a second-generation Dual Source Scanner. 120kV equivalent mixed-images were derived from reconstructions of the 100/Sn140kV source images using no metal artefact reduction (NOMAR) and iterative metal artefact reduction (IMAR). We then generated monoenergetic extrapolations at 130keV from source images without IMAR (DEMAR) or from source images with IMAR, (IMAR + DEMAR). The degree of metal artefact was quantified for NOMAR, IMAR, DEMAR and IMAR+DEMAR using a Fourier-based method and subjectively rated on a five point Likert scale by two independent readers. Results In subjects with hip prosthesis, DEMAR and IMAR resulted in significantly reduced artefacts compared to standard reconstructions (33% vs. 56%; for DEMAR and IMAR; respectively, p<0.005), but the degree of artefact reduction was significantly higher for IMAR (all p<0.005). In contrast, in subjects with dental implants only IMAR showed a significant reduction of artefacts whereas DEMAR did not (71%, vs. 8% p<0.01 and p = 0.1; respectively). Furthermore, the combination of IMAR with DEMAR resulted in additionally reduced artefacts (Hip prosthesis: 47%, dental implants 18%; both p<0.0001). Conclusion IMAR allows for significantly higher reduction of metal artefacts caused by hip prostheses and dental implants, compared to a dual energy based method. The combination of DE-source images with IMAR and subsequent monoenergetic extrapolation provides an incremental benefit compared to both single methods.
引用
收藏
页数:12
相关论文
共 33 条
  • [21] Metal implants on CT: comparison of iterative reconstruction algorithms for reduction of metal artifacts with single energy and spectral CT scanning in a phantom model
    Fang, Jieming
    Zhang, Da
    Wilcox, Carol
    Heidinger, Benedikt
    Raptopoulos, Vassilios
    Brook, Alexander
    Brook, Olga R.
    ABDOMINAL RADIOLOGY, 2017, 42 (03) : 742 - 748
  • [22] Metal Artifact Reduction in Pelvic Computed Tomography With Hip Prostheses Comparison of Virtual Monoenergetic Extrapolations From Dual-Energy Computed Tomography and an Iterative Metal Artifact Reduction Algorithmin a Phantom Study
    Higashigaito, Kai
    Angst, Florian
    Runge, Val M.
    Alkadhi, Hatem
    Donati, Olivio F.
    INVESTIGATIVE RADIOLOGY, 2015, 50 (12) : 828 - 834
  • [23] CT metal artefact reduction for internal fixation of the proximal humerus: Value of mono-energetic extrapolation from dual-energy and iterative reconstructions
    Winklhofer, S.
    Benninger, E.
    Spross, C.
    Morsbach, F.
    Rahm, S.
    Ross, S.
    Jost, B.
    Thali, M. J.
    Stolzmann, P.
    Alkadhi, H.
    Guggenberger, R.
    CLINICAL RADIOLOGY, 2014, 69 (05) : E199 - E206
  • [24] Dual-Energy and Iterative Metal Artifact Reduction for Reducing Artifacts Due to Metallic Hardware: A Loosening Hip Phantom Study
    Kim, Yeun Jeong
    Cha, Jang Gyu
    Kim, Hunji
    Yi, Ji Sook
    Kim, Hyun-Joo
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2019, 212 (05) : 1106 - 1111
  • [25] Combination of Iterative Metal Artifact Reduction and Virtual Monoenergetic Reconstruction Using Split-Filter Dual-Energy CT in Patients With Dental Artifact on Head and Neck CT
    Schmidt, Alexander Marcel Alfred
    Grunz, Jan-Peter
    Petritsch, Bernhard
    Gruschwitz, Philipp
    Knarr, Jonas
    Huflage, Henner
    Bley, Thorsten Alexander
    Kosmala, Aleksander
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 218 (04) : 716 - 727
  • [26] Comparison of metal artifact reduction using single-energy CT and dual-energy CT with various metallic implants in cadavers
    Barreto, Izabella
    Pepin, Eric
    Davis, Ivan
    Dean, Cooper
    Massini, Tara
    Rees, John
    Olguin, Catherine
    Quails, Nathan
    Correa, Nathalie
    Rill, Lynn
    Arreola, Manuel
    EUROPEAN JOURNAL OF RADIOLOGY, 2020, 133
  • [27] Metal implants on abdominal CT: does split-filter dual-energy CT provide additional value over iterative metal artifact reduction?
    Hildegard M. Wichtmann
    Kai R. Laukamp
    Sebastian Manneck
    Konrad Appelt
    Bram Stieltjes
    Daniel T. Boll
    Matthias R. Benz
    Markus M. Obmann
    Abdominal Radiology, 2023, 48 : 424 - 435
  • [28] Metal implants on abdominal CT: does split-filter dual-energy CT provide additional value over iterative metal artifact reduction?
    Wichtmann, Hildegard M.
    Laukamp, Kai R.
    Manneck, Sebastian
    Appelt, Konrad
    Stieltjes, Bram
    Boll, Daniel T.
    Benz, Matthias R.
    Obmann, Markus M.
    ABDOMINAL RADIOLOGY, 2023, 48 (01) : 424 - 435
  • [30] Analysis of metal artifact reduction tools for dental hardware in CT scans of the oral cavity: kVp, iterative reconstruction, dual-energy CT, metal artifact reduction software: does it make a difference?
    An De Crop
    Jan Casselman
    Tom Van Hoof
    Melissa Dierens
    Elke Vereecke
    Nicolas Bossu
    Jaime Pamplona
    Katharina D’Herde
    Hubert Thierens
    Klaus Bacher
    Neuroradiology, 2015, 57 : 841 - 849