SIGNAL TO NOISE RATIO OPTIMIZATION FOR A CMUT BASED MEDICAL ULTRASOUND IMAGING SYSTEM

被引:1
|
作者
Zangabad, Reza Pakdaman [1 ]
Bozkurt, Ayhan [2 ]
Yaralioglu, Goksenin [3 ]
机构
[1] Erasmus Univ, Med Ctr, Dept Biomed Engn, Rotterdam, Netherlands
[2] Sabanci Univ, Dept Elect Engn, Istanbul, Turkey
[3] Ozyegin Univ, Dept Elect Engn, Istanbul, Turkey
关键词
RADIATION IMPEDANCE; CIRCUIT MODEL; TRANSDUCERS; MEMBRANE; PERFORMANCE; CELL;
D O I
10.1109/ULTSYM.2015.0470
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
CMUTs offers key performance benefits compared to their piezoelectric counterparts. However, CMUTs are not widely adopted in commercial ultrasound imaging systems due to low signal to noise ratio (SNR) in the obtained images. Tunable parameters of the CMUT include the membrane material, radius, thickness, gap height, electrode size, and bias voltage. The aim of this work is to optimize these parameters to provide a solution for low SNR problem of CMUTs. In medical imaging, the device will be used in pulse echo mode for which both good receive sensitivity and high output pressure is required. Optimizing these performance metrics will end up in conflicting design parameters, such as a high gap for large pressure output whereas a small gap for better receive sensitivity. In this work, we develop a methodology for the investigation of the round-trip behavior of the transducer and try to co-optimize transmit and receive characteristics of the device to maximize the SNR of the received signal by tuning the bias voltage.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Signal-to-noise optimization of medical imaging systems
    Cunningham, IA
    Shaw, R
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1999, 16 (03): : 621 - 632
  • [2] MEMS-Based Transducers (CMUT) and Integrated Electronics for Medical Ultrasound Imaging
    Savoia, Alessandro S.
    Caliano, Giosue
    [J]. SENSORS, 2018, 431 : 421 - 429
  • [3] A 50-MHz CMUT Probe for Medical Ultrasound Imaging
    Zhuang, Steve
    Zhao, Danhua
    Chen, Lei
    Zhai, Liang
    [J]. 2018 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2018,
  • [4] Multiplane wave imaging increases signal-to-noise ratio in ultrafast ultrasound imaging
    Tiran, Elodie
    Deffieux, Thomas
    Correia, Mafalda
    Maresca, David
    Osmanski, Bruno-Felix
    Sieu, Lim-Anna
    Bergel, Antoine
    Cohen, Ivan
    Pernot, Mathieu
    Tanter, Mickael
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (21): : 8549 - 8566
  • [5] Advancements on Silicon Ultrasound Probes (CMUT) for Medical Imaging Applications
    Caliano, Giosue
    Savoia, Alessandro S.
    [J]. APPLICATIONS IN ELECTRONICS PERVADING INDUSTRY, ENVIRONMENT AND SOCIETY, APPLEPIES 2014, 2016, 351 : 51 - 57
  • [6] A Commercialized High Frequency CMUT Probe for Medical Ultrasound Imaging
    Zhao, Danhua
    Zhuang, Steve
    Daigle, Ron
    [J]. 2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [7] Medical imaging with capacitive micromachined ultrasound transducer (cMUT) arrays
    Mills, DM
    [J]. 2004 IEEE Ultrasonics Symposium, Vols 1-3, 2004, : 384 - 390
  • [8] DOUBLE PULSE TRANSMISSION - SIGNAL-TO-NOISE RATIO IMPROVEMENT IN ULTRASOUND IMAGING
    Trots, Ihor
    Nowicki, Andrzej
    Lewandowski, Marcin
    Secomski, Wojciech
    Litniewski, Jerzy
    [J]. ARCHIVES OF ACOUSTICS, 2008, 33 (04) : 593 - 601
  • [9] Converting Coherence to Signal-to-noise Ratio for Enhancement of Adaptive Ultrasound Imaging
    Hasegawa, Hideyuki
    Nagaoka, Ryo
    [J]. ULTRASONIC IMAGING, 2020, 42 (01) : 27 - 40
  • [10] Performance Optimization of a High Frequency CMUT Probe for Medical Imaging
    Savoia, Alessandro
    Caliano, Giosue
    Mauti, Barbara
    Pappalardo, Massimo
    [J]. 2011 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2011, : 600 - 603