flow injection analysis;
ion-selective electrode;
bromide;
therapeutic drug monitoring;
epilepsy;
D O I:
10.1016/S0731-7085(96)02017-1
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
A flow injection system was constructed using a bromide-selective electrode and used to determine serum bromide in patients with epilepsy. A 10-mu l serum sample was injected into a carrier stream flowing at 0.12 mi min(-1). Potential changes and bromide concentrations were linearly related in the range 3-50 mM. The lower limit of detection for serum bromide was 1 mM and this electrode sensitivity spanned the entire concentration range required for bromide therapy (9-24 mM). The results compared favourably with those obtained by colorimetry. (C) 1997 Elsevier Science B.V.