Flow measurement at the pump head of centrifugal pumps: Comparison of ultrasonic transit time and ultrasonic Doppler systems

被引:6
|
作者
Schima, H
Huber, L
Schmallegger, H
Drost, CJ
Droudt, A
Wieselthaler, G
Losert, U
机构
[1] BOLTZMANN INST CARDIOSURG RES,DEPT CARDIOTHORAC SURG,VIENNA,AUSTRIA
[2] UNIV VIENNA,CTR BIOMED ENGN,VIENNA,AUSTRIA
[3] TRANSON SYST INC,ITHACA,NY
关键词
rotary pump; monitoring; ultrasound; flow; transit time;
D O I
10.1111/j.1525-1594.1997.tb03748.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Determination of blood flow is essential for monitoring rotary blood pumps. However, accurate measurement directly adjacent to the pump housing is difficult because of the highly irregular flow profiles near the fast spinning rotor. Therefore, a specially adapted flow probe based on the ultrasound transit time (USTT) principle was designed to evaluate the flow in centrifugal blood pumps. The probe can be directly mounted at the housing and creates 2 crossed measuring ultrasound beams. The mean value, Q(m), of the 2 output signals corresponds to the blood flow and the difference, Q(d), correlates to the vorticity of the flow profile in the pump outflow tract. In vitro measurements obtained an accuracy for mean flow values of better than +/-0.6 L/min in extreme working points and for vorticity values even as high as Q(d) = 3.5 L/min. Because of vorticity, however, the output signal contained considerable noise, and that required the application of a 10 Hz filter. Positioning of the ultrasound (US) beams parallel to the axial direction of the pump was superior to radial positioning. Additional measurement of the flow profile demonstrated that a large vorticity occurred (up to Q(d) equal to 3.5 L/min), and this vorticity was highly dependent upon the afterload of the pump. In vivo experiments demonstrated the reliability of the method. We concluded that USTT flow measurement can determine blood flow immediately adjacent to the pump housing with sufficient accuracy, and these measurements are superior to those from US-Doppler systems (which cannot handle the vorticity accurately enough) and electromagnetic devices (which lack zero stability).
引用
收藏
页码:808 / 815
页数:8
相关论文
共 50 条
  • [21] Accuracy in real time ultrasonic applications and transit-time flow meters
    Eren, H
    WHERE INSTRUMENTATION IS GOING - CONFERENCE PROCEEDINGS, VOLS 1 AND 2, 1998, : 568 - 572
  • [22] DESIGN AND EVALUATION OF PIEZOELECTRIC SENSORS FOR THE MEASUREMENT OF BLOOD FLOW IN CORONARY IMPLANTS BY THE ULTRASONIC TRANSIT TIME METHOD
    Carrillo, E.
    Jimenez, A.
    Lopez, R.
    Portelles, J. J.
    Desdin, L. F.
    REVISTA CUBANA DE FISICA, 2020, 37 (02): : 116 - 124
  • [23] SIGNAL TRANSMISSION BY ULTRASONIC FLUID-FLOW MEASUREMENT IN FOLLOWING THE TRANSIT-TIME DIFFERENCE PROCEDURE
    LECHNER, H
    ACUSTICA, 1983, 54 (01): : 33 - 38
  • [24] ANALYSIS OF TRANSIT-TIME EFFECTS ON DOPPLER FLOW MEASUREMENT
    NEWHOUSE, VL
    BENDICK, PJ
    VARNER, LW
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1976, 23 (05) : 381 - 387
  • [25] INTEGRATED ULTRASONIC PULSED DOPPLER SYSTEM FOR MEASUREMENT OF BLOOD-FLOW
    ROONEY, JA
    HEYSER, RC
    AVIATION SPACE AND ENVIRONMENTAL MEDICINE, 1985, 56 (05): : 493 - 493
  • [26] COMPARATIVE MEASUREMENT OF BLOOD-FLOW WITH ELECTROMAGNETIC AND ULTRASONIC DOPPLER TECHNIQUE
    PASCH, T
    VONDEREM.J
    BAUER, RD
    ZEITSCHRIFT FUR KREISLAUFFORSCHUNG, 1972, 61 (06): : 485 - &
  • [27] An Extracting Algorithm of Ultrasonic Doppler Shift for Multiphase Fluid Flow Measurement
    Cao Xin-liang
    Cui Wei
    Wang Meng-de
    CURRENT DEVELOPMENT OF MECHANICAL ENGINEERING AND ENERGY, PTS 1 AND 2, 2014, 494-495 : 817 - 820
  • [28] Flow rate measurement using Ultrasonic Doppler Method with cavitation bubbles
    Koike, Y
    Tsuyoshi, T
    Kikura, H
    Aritomi, M
    Mori, M
    2002 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2002, : 531 - 534
  • [29] MEASUREMENT OF BLOOD-FLOW WITH ULTRASONIC PULSED DOPPLER-EFFECT
    GREENE, ER
    REVISTA MEDICA DE CHILE, 1983, 111 (05) : 495 - 504
  • [30] Installation Effects of a Clamp-On Transit Time Ultrasonic Flow Meter
    Kumar, K.
    Farande, K. U.
    Sahoo, G. C.
    INTERNATIONAL JOURNAL OF FLUID MECHANICS RESEARCH, 2011, 38 (06) : 489 - 498