Flow measurement at the pump head of centrifugal pumps: Comparison of ultrasonic transit time and ultrasonic Doppler systems

被引:6
|
作者
Schima, H
Huber, L
Schmallegger, H
Drost, CJ
Droudt, A
Wieselthaler, G
Losert, U
机构
[1] BOLTZMANN INST CARDIOSURG RES,DEPT CARDIOTHORAC SURG,VIENNA,AUSTRIA
[2] UNIV VIENNA,CTR BIOMED ENGN,VIENNA,AUSTRIA
[3] TRANSON SYST INC,ITHACA,NY
关键词
rotary pump; monitoring; ultrasound; flow; transit time;
D O I
10.1111/j.1525-1594.1997.tb03748.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Determination of blood flow is essential for monitoring rotary blood pumps. However, accurate measurement directly adjacent to the pump housing is difficult because of the highly irregular flow profiles near the fast spinning rotor. Therefore, a specially adapted flow probe based on the ultrasound transit time (USTT) principle was designed to evaluate the flow in centrifugal blood pumps. The probe can be directly mounted at the housing and creates 2 crossed measuring ultrasound beams. The mean value, Q(m), of the 2 output signals corresponds to the blood flow and the difference, Q(d), correlates to the vorticity of the flow profile in the pump outflow tract. In vitro measurements obtained an accuracy for mean flow values of better than +/-0.6 L/min in extreme working points and for vorticity values even as high as Q(d) = 3.5 L/min. Because of vorticity, however, the output signal contained considerable noise, and that required the application of a 10 Hz filter. Positioning of the ultrasound (US) beams parallel to the axial direction of the pump was superior to radial positioning. Additional measurement of the flow profile demonstrated that a large vorticity occurred (up to Q(d) equal to 3.5 L/min), and this vorticity was highly dependent upon the afterload of the pump. In vivo experiments demonstrated the reliability of the method. We concluded that USTT flow measurement can determine blood flow immediately adjacent to the pump housing with sufficient accuracy, and these measurements are superior to those from US-Doppler systems (which cannot handle the vorticity accurately enough) and electromagnetic devices (which lack zero stability).
引用
收藏
页码:808 / 815
页数:8
相关论文
共 50 条
  • [1] Optimized Ultrasonic Transit Time Measurement and Discharge Curve in Comparison
    Lambeck, Torsten
    WASSERWIRTSCHAFT, 2021, 111 (7-8) : 50 - 53
  • [2] DOPPLER ULTRASONIC MEASUREMENT OF CAROTID FLOW
    MULLER, HR
    RADUE, EW
    SAIA, A
    PALLOTTI, C
    ULTRASOUND IN MEDICINE AND BIOLOGY, 1983, 9 (01): : L91 - L95
  • [3] A Novel Mathematical Model for Transit-time Ultrasonic Flow Measurement
    Kang, Lei
    Feeney, Andrew
    Somerset, Will
    Su, Riliang
    Lines, David
    Ramadas, Sivaram Nishal
    Dixon, Steve
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 1590 - 1593
  • [4] MEASUREMENT OF TIDAL FLOW USING A TRANSIT-TIME ULTRASONIC BREATH ANALYZER
    WILLIAMS, EM
    BURROUGH, SLM
    MCPEAK, H
    ANAESTHESIA, 1995, 50 (05) : 427 - 432
  • [5] Intraoperative evaluation of coronary anastomosis by transit-time ultrasonic flow measurement
    Morota, T
    Duhaylongsod, FG
    Burfeind, WR
    Huang, CT
    ANNALS OF THORACIC SURGERY, 2002, 73 (05): : 1446 - 1450
  • [6] Flow acoustics modelling and implications for ultrasonic flow measurement based on the transit-time method
    Willatzen, M
    ULTRASONICS, 2004, 41 (10) : 805 - 810
  • [7] ATRAUMATIC MEASUREMENT OF BLOOD PRESSURE BY DOPPLER ULTRASONIC FLOW
    FRANKLIN, D
    PIERSON, K
    KAZAMIAS, TM
    CIRCULATION, 1969, 40 (4S3) : II83 - &
  • [8] USE OF THE "ULTRASONIC TRANSIT TIME" TECHNIQUE FOR THE MEASUREMENT OF BLOOD FLOW IN VESSELS: RESULTS OBTAINED
    Jimenez, A.
    Moreno, E.
    Carrillo, E.
    Torres, D.
    REVISTA CUBANA DE FISICA, 2011, 28 (01): : 4 - 10
  • [9] Application of ultrasonic doppler method for bubbly flow measurement using two ultrasonic frequencies
    Murakawa, H
    Kikura, H
    Aritomi, M
    EXPERIMENTAL THERMAL AND FLUID SCIENCE, 2005, 29 (07) : 843 - 850
  • [10] Noninvasive measurement of cerebrospinal fluid flow using an ultrasonic transit time flow sensor: a preliminary study
    Pennell, Thomas
    Yi, Juneyoung L.
    Kaufman, Bruce A.
    Krishnamurthy, Satish
    JOURNAL OF NEUROSURGERY-PEDIATRICS, 2016, 17 (03) : 270 - 277