Unipolar electrogram characteristics predictive of successful radiofrequency catheter ablation of accessory pathways

被引:11
|
作者
Barlow, MA [1 ]
Klein, GJ [1 ]
Simpson, CS [1 ]
Murgatroyd, FD [1 ]
Yee, R [1 ]
Krahn, AD [1 ]
Skanes, AC [1 ]
机构
[1] Univ Western Ontario, Div Cardiol, London, ON N6A 5A5, Canada
关键词
unipolar; electrogram; radiofrequency; ablation; accessory pathway;
D O I
10.1111/j.1540-8167.2000.tb00313.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Unipolar Electrogram in Ablation, Introduction: The purpose of this study was to determine the characteristics of the unipolar electrogram that are most helpful in predicting successful radiofrequency ablation of accessory pathways. Methods and Results: The unipolar electrogram was analyzed at 185 ablation sites in 53 patients; 94 attempts were directed at the site of earliest atrial activation ("atrial group") and 91 at the site of earliest ventricular activation ("ventricular group"). The electrogram was analyzed for several features, including pattern ("QS" or "initial R"), Unipolar pattern: Overall, a "QS" pattern was seen at 55% of unsuccessful, 75% of temporarily successful, and 90% of permanently successful sites. For the atrial group, the respective frequencies were 53%, 77%, and 92%, and for the ventricular group, 57%, 73%, and 86%, The difference in pattern distribution between unsuccessful and permanently successful sites was significant for all groups: overall, P < 0.0001; atrial group, P = 0.0005; ventricular group, P = 0.02. Absence of a "QS" pattern (i.e., "initial R") predicted a 92% chance of unsuccessful ablation, Additional features: Activation times were significantly shorter at permanently successful than at unsuccessful (P < 0.0001) or temporarily successful sites (P = 0.0002), No significant differences were found in atrial or ventricular amplitudes or in AN ratios. Intrinsic deflection slew was lower at temporarily successful sites (P = 0.03 vs all other sites), Conclusion: Ablation at sites revealing an "initial R" pattern (i.e., absent "QS") is very unlikely to be successful. Activation time is shorter at successful sites. These features are equally applicable when mapping the atrial potential as when mapping the ventricular potential.
引用
收藏
页码:146 / 154
页数:9
相关论文
共 50 条
  • [31] RADIOFREQUENCY CATHETER ABLATION OF ACCESSORY PATHWAYS - A CONTEMPORARY REVIEW
    HAISSAGUERRE, M
    GAITA, F
    MARCUS, FI
    CLEMENTY, J
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1994, 5 (06) : 532 - 552
  • [32] RADIOFREQUENCY CATHETER ABLATION OF ACCESSORY PATHWAYS IN 58 DOGS
    Webber, D.
    Wright, K.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2013, 27 (03) : 627 - 628
  • [33] RADIOFREQUENCY CURRENT CATHETER ABLATION OF ACCESSORY ATRIOVENTRICULAR PATHWAYS
    KUCK, KH
    SCHLUTER, M
    GEIGER, M
    SIEBELS, J
    DUCKECK, W
    LANCET, 1991, 337 (8757): : 1557 - 1561
  • [34] Unipolar electrogram-guided radiofrequency catheter ablation in paroxysmal atrial fibrillation: electrogram patterns and outcomes
    Guohua Fu
    Bin He
    Binhao Wang
    Jing Liu
    Yibo Yu
    Xianfeng Du
    Mingjun Feng
    Fang Gao
    He Jin
    Renyuan Fang
    Huimin Chu
    Journal of Interventional Cardiac Electrophysiology, 2019, 54 : 203 - 208
  • [35] CATHETER ABLATION OF ATRIOVENTRICULAR ACCESSORY PATHWAYS BY RADIOFREQUENCY CURRENT
    WANG, LX
    HU, DY
    DING, YS
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 1993, 42 (02) : 155 - 160
  • [36] Unipolar electrogram-guided radiofrequency catheter ablation in paroxysmal atrial fibrillation: electrogram patterns and outcomes
    Fu, Guohua
    He, Bin
    Wang, Binhao
    Liu, Jing
    Yu, Yibo
    Du, Xianfeng
    Feng, Mingjun
    Gao, Fang
    Jin, He
    Fang, Renyuan
    Chu, Huimin
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2019, 54 (03) : 203 - 208
  • [37] ATRIAL ELECTROGRAM CHARACTERISTICS ASSOCIATED WITH SUCCESSFUL RADIOFREQUENCY CATHETER ABLATION OF THE SLOW AV NODAL PATHWAY
    BURROUGHS, JM
    ROME, MP
    SCHAAL, SF
    NELSON, SD
    LOVE, CJ
    CIRCULATION, 1993, 88 (04) : 491 - 491
  • [38] ATRIAL ELECTROGRAM CHARACTERISTICS ASSOCIATED WITH SUCCESSFUL RADIOFREQUENCY CATHETER ABLATION OF THE SLOW AV NODAL PATHWAY
    BURROUGHS, JM
    ROME, MP
    SCHAAL, SF
    NELSON, SD
    LOVE, CJ
    CLINICAL RESEARCH, 1993, 41 (03): : A629 - A629
  • [39] Radiofrequency catheter ablation of septal accessory pathways within the triangle of Koch: Importance of energy titration testing other than the local electrogram characteristics for identifying the successful target site
    Lin, JL
    Huang, SKS
    Lai, LP
    Cheng, TF
    Tseng, YZ
    Lien, WP
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (10): : 1909 - 1917
  • [40] RECURRENT TACHYCARDIA AND SYMPTOMS FOLLOWING SUCCESSFUL RADIOFREQUENCY CATHETER ABLATION OF ACCESSORY PATHWAYS - INCIDENCE AND MECHANISMS
    WADAS, MA
    OLSHANSKY, B
    BAERMAN, JM
    KALL, JG
    KOPP, DE
    WILBER, DJ
    CIRCULATION, 1992, 86 (04) : 783 - 783