Exploring Evaluation Variables for Low-Cost Particulate Matter Monitors to Assess Occupational Exposure

被引:10
|
作者
Ruiter, Sander [1 ]
Kuijpers, Eelco [1 ]
Saunders, John [2 ]
Snawder, John [3 ]
Warren, Nick [2 ]
Gorce, Jean-Philippe [2 ]
Blom, Marcus [1 ]
Krone, Tanja [1 ]
Bard, Delphine [2 ]
Pronk, Anjoeka [1 ]
Cauda, Emanuele [3 ]
机构
[1] Netherlands Org Appl Sci Res TNO, NL-3584 CB Utrecht, Netherlands
[2] HSE Sci & Res Ctr, Harpur Hill, Buxton SK17 9JN, England
[3] NIOSH, Ctr Dis Control & Prevent, 1090 Tusculum Ave, Cincinnati, OH 45226 USA
关键词
low-cost monitors; wearables; sensors; evaluation; occupational; exposure monitoring; particulate matter; AIR; CALIBRATION; SENSORS; AMBIENT;
D O I
10.3390/ijerph17228602
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
(1) Background: Small, lightweight, low-cost optical particulate matter (PM) monitors are becoming popular in the field of occupational exposure monitoring, because these devices allow for real-time static measurements to be collected at multiple locations throughout a work site as well as being used as wearables providing personal exposure estimates. Prior to deployment, devices should be evaluated to optimize and quantify measurement accuracy. However, this can turn out to be difficult, as no standardized methods are yet available and different deployments may require different evaluation procedures. To gain insight in the relevance of different variables that may affect the monitor readings, six PM monitors were selected based on current availability and evaluated in the laboratory; (2) Methods: Existing strategies that were judged appropriate for the evaluation of PM monitors were reviewed and seven evaluation variables were selected, namely the type of dust, within- and between-device variations, nature of the power supply, temperature, relative humidity, and exposure pattern (peak and constant). Each variable was tested and analyzed individually and, if found to affect the readings significantly, included in a final correction model specific to each monitor. Finally, the accuracy for each monitor after correction was calculated; (3) Results: The reference materials and exposure patterns were found to be main factors needing correction for most monitors. One PM monitor was found to be sufficiently accurate at concentrations up to 2000 mu g/m(3) PM2.5, with other monitors appropriate at lower concentrations. The average accuracy increased by up to three-fold compared to when the correction model did not include evaluation variables; (4) Conclusions: Laboratory evaluation and readings correction can greatly increase the accuracy of PM monitors and set boundaries for appropriate use. However, this requires identifying the relevant evaluation variables, which are heavily reliant on how the monitors are used in the workplace. This, together with the lack of current consensus on standardized procedures, shows the need for harmonized PM monitor evaluation methods for occupational exposure monitoring.
引用
收藏
页码:1 / 18
页数:18
相关论文
共 50 条
  • [31] Evaluation of Personal Protective Equipment Using Low-cost Aerosol Monitors
    Baldelli, Alberto
    Poznikoff, Andrew
    Heieis, Kevin
    Purdy, Robert
    AEROSOL AND AIR QUALITY RESEARCH, 2024, 24 (06)
  • [32] Calibration of low-cost particulate matter sensors for coal dust monitoring
    Amoah, Nana A.
    Xu, Guang
    Kumar, Ashish Ranjan
    Wang, Yang
    SCIENCE OF THE TOTAL ENVIRONMENT, 2023, 859
  • [33] Investigation of Low-Cost and Optical Particulate Matter Sensors for Ambient Monitoring
    Rogulski, Mariusz
    Badyda, Artur
    ATMOSPHERE, 2020, 11 (10)
  • [34] A Review of Low-Cost Particulate Matter Sensors from the Developers' Perspectives
    Alfano, Brigida
    Barretta, Luigi
    Del Giudice, Antonio
    De Vito, Saverio
    Di Francia, Girolamo
    Esposito, Elena
    Formisano, Fabrizio
    Massera, Ettore
    Miglietta, Maria Lucia
    Polichetti, Tiziana
    SENSORS, 2020, 20 (23) : 1 - 56
  • [35] Opportunities for Low-Cost Particulate Matter Sensors in Filter Emission Measurements
    Schwarz, Almuth D.
    Meyer, Joerg
    Dittler, Achim
    CHEMICAL ENGINEERING & TECHNOLOGY, 2018, 41 (09) : 1826 - 1832
  • [36] Suitability of Low-Cost Particulate Matter Sensors for Measurements in Ventilation Systems
    Marval, Jesus
    Medina, Luis
    Vallejo, Juan
    Tronville, Paolo
    IAQ 2020: INDOOR ENVIRONMENTAL QUALITY PERFORMANCE APPROACHES, PT 2, 2022,
  • [37] Field and Laboratory Evaluations of the Low-Cost Plantower Particulate Matter Sensor
    Zamora, Misti Levy
    Xiong, Fulizi
    Gentner, Drew
    Kerkez, Branko
    Kohrman-Glaser, Joseph
    Koehler, Kirsten
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2019, 53 (02) : 838 - 849
  • [38] A Review of Literature on the Usage of Low-Cost Sensors to Measure Particulate Matter
    Raysoni, Amit U.
    Pinakana, Sai Deepak
    Mendez, Esmeralda
    Wladyka, Dawid
    Sepielak, Katarzyna
    Temby, Owen
    EARTH, 2023, 4 (01): : 168 - 186
  • [39] Low-Cost Particulate Matter Sensors for Monitoring Residential Wood Burning
    Hassani, Amirhossein
    Schneider, Philipp
    Vogt, Matthias
    Castell, Nuria
    ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2023, 57 (40) : 15162 - 15172
  • [40] Application of low-cost particulate matter sensors for air quality monitoring and exposure assessment in underground mines: A review
    Nana A.Amoah
    Guang Xu
    Yang Wang
    Jiayu Li
    Yongming Zou
    Baisheng Nie
    InternationalJournalofMinerals,MetallurgyandMaterials, 2022, (08) : 1475 - 1490