Topographical Analysis of Non-Glaucomatous Myopic Optic Discs Using a Confocal Scanning Laser Ophthalmoscope (TopSS)

被引:2
|
作者
Oh, Seung Hoon [1 ]
Chung, Sung Kun [2 ]
Lee, Na Young [1 ]
机构
[1] Catholic Univ Korea, Incheon St Marys Hosp, Dept Ophthalmol & Visual Sci, Inchon, South Korea
[2] Catholic Univ Korea, Coll Med, Yeoui Do St Marys Hosp, Dept Ophthalmol & Visual Sci, Seoul 137701, South Korea
关键词
Confocal scanning laser ophthalmoscope; glaucoma; HRT; myopia; TopSS; OPEN-ANGLE GLAUCOMA; HEIDELBERG RETINA TOMOGRAPH; NERVE HEAD; COHERENCE TOMOGRAPHY; OCULAR HYPERTENSION; EYES; THICKNESS; PHOTOGRAPHY; PROGRESSION; DAMAGE;
D O I
10.3109/08820538.2014.912339
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objectives: To show normative data of optic discs and the mechanism of glaucoma in people with myopia. Design: Cross-sectional study. Participants: This study investigated 89 Korean adults with myopia but without glaucoma. Methods: Patients were divided into three groups according to the refractive error: low, moderate, and high; and axial length: normal or below normal length, moderately long, and extremely long. Optic disc variables were obtained by confocal scanning laser ophthalmoscope and compared among groups. Results: The optic disc parameters have a correlation between the refractive error and the optic disc parameters such as average depth, volume below, and half-depth volume. Those parameters also decreased as the axial length increased. The thickness of the volume above decreased significantly as the axial length increased, but a similar relationship was not evident with the refractive error change. In addition, the optic disc parameters were analyzed with respect to the 12 clockwise directions. Conclusions: Analyses of optic disc parameters provided by TopSS revealed the height of the disc decreased as the myopic refractive error and/or axial length increased. The RNFL bundle became compacted in the thinner disc of the myopic population. This could be an explanation for the fragility of the RNFL in the myopic population. The 12 radial section analyses revealed the shallow cupping at the temporal side in the high-myopic, very-long-axis group. The neuroretinal rim (NRR) height significantly decreased at the superior and inferior sides. These findings suggest that the RNFL bundle should be under high mechanical strain in these sectors.
引用
收藏
页码:397 / 409
页数:13
相关论文
共 50 条
  • [1] Ability of a confocal scanning laser ophthalmoscope (TopSS) to detect early glaucomatous visual field defect
    Ahn, BS
    Kee, C
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2000, 84 (08) : 852 - 855
  • [2] Use of a confocal laser scanning ophthalmoscope to detect glaucomatous cupping of the optic disc
    Cucevic, V
    Brooks, AMV
    Strang, NT
    Klein, AG
    Nugent, KA
    [J]. AUSTRALIAN AND NEW ZEALAND JOURNAL OF OPHTHALMOLOGY, 1997, 25 (03): : 217 - 220
  • [3] CONFOCAL LASER SCANNING OPHTHALMOSCOPE - REPRODUCIBILITY OF OPTIC-NERVE HEAD TOPOGRAPHIC MEASUREMENTS WITH THE CONFOCAL LASER SCANNING OPHTHALMOSCOPE
    CIOFFI, GA
    ROBIN, AL
    EASTMAN, RD
    PERELL, HF
    SAFARAZI, FA
    KELMAN, SE
    [J]. OPHTHALMOLOGY, 1993, 100 (01) : 57 - 62
  • [4] Analysis of optic nerve head by confocal scanning laser ophthalmoscope in ocular hypertension
    Kria, L
    Beltaief, O
    Farah, H
    Anane, R
    Erraies, K
    Karray, F
    Boussen, IM
    Chkir, R
    Zhioua, R
    Ouertani, AM
    [J]. JOURNAL FRANCAIS D OPHTALMOLOGIE, 2004, 27 (09): : 995 - 999
  • [5] Assessment of localised fluid at the macula using the TopSS scanning laser ophthalmoscope.
    Mahomed, I
    Waldock, A
    Markham, RH
    Grey, RHB
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 2818 - 2818
  • [6] Optic disk analysis by confocal scanning laser ophthalmoscope in rat optic nerve crush model using aneurysm clip
    Park, K
    Kang, JH
    Kim, YJ
    Kim, DM
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 : U1163 - U1163
  • [7] Neuroretinal rim in non-glaucomatous large optic nerve heads: a comparison of confocal scanning laser tomography and spectral domain optical coherence tomography
    Enders, Philip
    Schaub, Friederike
    Hermann, Manuel M.
    Cursiefen, Claus
    Heindl, Ludwig M.
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2017, 101 (02) : 138 - 142
  • [8] Effect of glaucomatous damage on repeatability of confocal scanning laser ophthalmoscope, scanning laser polarimetry, and optical coherence tomography
    Ortega, Julio E. DeLeon
    Sakata, Lisandro M.
    Kakati, Bobby
    McGwin, Gerald, Jr.
    Monheit, Blythe E.
    Arthur, Stella N.
    Girkin, Christopher A.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (03) : 1156 - 1163
  • [9] Glaucomatous optic disc changes in myopic eyes greater than -5 diopters using TopSS(TM)
    Ochiai, H
    Liu, X
    Dong, J
    Chihara, E
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1997, 38 (04) : 3868 - 3868
  • [10] Ability of MultiColor scanning laser ophthalmoscope to detect non-glaucomatous retinal nerve fiber layer defects in eyes with retinal diseases
    Terasaki, Hiroto
    Sonoda, Shozo
    Kakiuchi, Naoko
    Shiihara, Hideki
    Yamashita, Takehiro
    Sakamoto, Taiji
    [J]. BMC OPHTHALMOLOGY, 2018, 18