Evaluation and second check of a commercial Monte Carlo dose engine for small-field apertures in pencil beam scanning proton therapy

被引:9
|
作者
Holmes, Jason [1 ]
Shen, Jiajian [1 ]
Shan, Jie [1 ]
Patrick, Christopher L. [1 ]
Wong, William W. [1 ]
Foote, Robert L. [2 ]
Patel, Samir H. [1 ]
Bues, Martin [1 ]
Liu, Wei [1 ]
机构
[1] Mayo Clin, Dept Radiat Oncol, 5777 E Mayo Blvd, Phoenix, AZ 85054 USA
[2] Mayo Clin, Dept Radiat Oncol, Rochester, MN USA
关键词
Monte-Carlo; patient-specific apertures; pencil beam scanning; PATIENT-SPECIFIC APERTURE; SPOT SIZE; VERIFICATION; SIMULATION; SYSTEM; IMPACT;
D O I
10.1002/mp.15604
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To evaluate the accuracy of the RayStation Monte Carlo dose engine (RayStation MC) in modeling small-field block apertures in proton pencil beam scanning. Furthermore, we evaluate the suitability of MCsquare as a second check for RayStation MC. Methods We have enhanced MCsquare to model block apertures. To test the accuracy of both RayStation MC and the newly enhanced MCsquare, we compare the dose predictions of each to in-water dose measurements obtained using diode detectors and radiochromic film. Nine brass apertures with openings of 1, 2, 3, 4, and 5 cm and either 2 cm or 4 cm thickness were used in the irradiation of a water phantom. Two measurement setups were used, one with a range shifter and 119.7 MeV proton beam energy and the other with no range shifter and 147 MeV proton beam energy. To further test the validity of RayStation MC and MCsquare in modeling block apertures and to evaluate MCsquare as a second check tool, 10 small-field (average target volume 8.3 cm(3)) patient treatment plans were calculated by each dose engine followed by a statistical comparison. Results Comparing to the absolute dose measurements in water, RayStation MC differed by 1.2% +/- 1.0% while MCsquare differed by -1.8% +/- 3.7% in the plateau region of a pristine Bragg peak. Compared to the in-water film measurements, RayStation MC and MCsquare both performed well with an average 2D-3D gamma passing rate of 99.4% and 99.7% (3%/3 mm), respectively. A t-test comparing the agreement with the film measurements between RayStation MC and MCsquare suggested that the relative spatial dose distributions calculated by MCsquare and RayStation MC were statistically indistinguishable. Directly comparing the dose calculations between MCsquare and RayStation MC over 10 patients resulted in an average 3D-3D gamma passing rates of 98.5% (3%/3 mm) and 94.1% (2%/2 mm), respectively. Conclusion The validity of RayStation MC algorithm for use with patient-specific apertures has been expanded to include small apertures. MCsquare has been enhanced to model apertures and was found to be an adequate second check of RayStation MC in this scenario.
引用
收藏
页码:3497 / 3506
页数:10
相关论文
共 50 条
  • [1] Validation of a Commercial Treatment Planning System for Small-Field Patient-Specific Apertures in Pencil Beam Scanning Proton Therapy
    Holmes, J.
    Shen, J.
    Shan, J.
    Wong, W.
    Foote, R.
    Patel, S.
    Bues, M.
    Liu, W.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [2] Monte Carlo Dose Verification of Pencil Beam Scanning Proton Therapy
    Giantsoudi, D.
    Clasie, B.
    Grassberger, C.
    Dowdell, S.
    Depauw, N.
    Madden, T.
    Kooy, H.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [3] Evaluation of Lateral Penumbra for Proton Pencil Beam Scanning with and Without Brass Apertures Using a Commercial Monte Carlo Dose Calculation Algorithm
    Maes, D.
    Regmi, R.
    Egan, A.
    Nevitt, Z.
    Leuro, E.
    Wong, T.
    Bowen, S.
    Saini, J.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E270 - E270
  • [4] Validation of pencil beam scanning proton therapy with multi-leaf collimator calculated by a commercial Monte Carlo dose engine
    Tominaga, Yuki
    Sakurai, Yusuke
    Miyata, Junya
    Harada, Shuichi
    Akagi, Takashi
    Oita, Masataka
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12):
  • [5] Neutron Dose Equivalent Evaluation for Pencil Beam Scanning Proton Therapy with Apertures
    Geng, C.
    Schuemann, J.
    Moteabbed, M.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3466 - 3466
  • [6] Clinical validation of a GPU-based Monte Carlo dose engine of a commercial treatment planning system for pencil beam scanning proton therapy
    Fracchiolla, Francesco
    Engwall, Erik
    Janson, Martin
    Tamm, Fredrik
    Lorentini, Stefano
    Fellin, Francesco
    Bertolini, Mattia
    Algranati, Carlo
    Righetto, Roberto
    Farace, Paolo
    Amichetti, Maurizio
    Schwarz, Marco
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 88 : 226 - 234
  • [7] Clinical validation of Monte Carlo dose calculation for pencil beam scanning proton therapy.
    Widesott, L.
    Lorentini, S.
    Fracchiolla, F.
    Farace, P.
    Schwarz, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S45 - S46
  • [8] Improvements in pencil beam scanning proton therapy dose calculation accuracy in brain tumor cases with a commercial Monte Carlo algorithm
    Widesott, Lamberto
    Lorentini, Stefano
    Fracchiolla, Francesco
    Farace, Paolo
    Schwarz, Marco
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (14):
  • [9] A Monte Carlo study on the collimation of pencil beam scanning proton therapy beams
    Charlwood, Frances C.
    Aitkenhead, Adam H.
    Mackay, Ranald I.
    [J]. MEDICAL PHYSICS, 2016, 43 (03) : 1462 - 1472
  • [10] Monte Carlo model of a gantry for proton pencil beam scanning
    Winterhalter, Carla
    Safai, Sairos
    Grassberger, Clemens
    Oxley, David
    Weber, Damien C.
    Lomax, Tony
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (11) : 837 - 837