Development and Evaluation of a Pixel Detector System for Pencil Beam Scanning Proton Therapy

被引:3
|
作者
Eichin, Michael [1 ]
Actis, O. [1 ]
Grossmann, M. [1 ]
Konig, S. [1 ]
Mayor, A. [1 ]
Meer, D. [1 ]
机构
[1] Paul Scherrer Inst, CH-5232 Villigen, Switzerland
关键词
Biomedical applications of radiation; field-programmable gate arrays (FPGAs); position sensitive particle detector; printed circuits; PERFORMANCE; DOSIMETRY;
D O I
10.1109/TNS.2019.2918635
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
At the Paul Scherrer Institute (PSI) in Switzerland, cancer patients are treated with protons. The PSI Gantries work with Pencil Beam Scanning (PBS) technology. For beam profile characterization, a new 2-D planar pixel detector based on Printed Circuit Board technology was developed. The main challenge of the design was the limited amount of 256 readout channels while several thousands of detector signals had to be processed. With PBS, only a few pixels will be hit by the proton beam and give a significant signal, while most of them will not capture any signal. On the detector design, the active area was divided into pixel arrays with equal size, while one array is such big to cover the whole size of the proton beam. By a so-called channel recycling technique, the same pixel number of all pixel arrays were connected together to one readout channel. The proof of concept was verified with a small detector. In the next step, the design was changed such, getting a full-scale detector for Gantry 2. The readout electronics board itself is also a development performed at PSI. A core component of the design is the ADAS1128 chip from ANALOG Devices featuring 128 input channels. The pixel detector is fully integrated into the therapy control system (TCS) through readout electronics with a field-programmable gate array (FPGA) technology. The FPGA technology allows to readout the detector synchronized to the beam delivery. This paper summarizes the features of the pixel detector design, the readout electronics, and shows the first experimental results.
引用
收藏
页码:1273 / 1279
页数:7
相关论文
共 50 条
  • [41] Ion Recombination for Ionization Chamber Dosimetry in a Pencil Beam Scanning Proton Therapy Beam
    Sorriaux, J.
    Takoukam, P.
    Bertrand, D.
    Baumer, C.
    Lee, J.
    Vynckier, S.
    Sterpin, E.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [42] Toxicity outcomes of hypofractionated pencil beam scanning proton beam therapy for spinal Chordomas
    Burela, N.
    Chilukuri, S.
    Nangia, S.
    Patro, K.
    Mp, N.
    Gaikwad, U.
    Sundar, S.
    Thimma, R.
    Rajendran, A.
    Sulaiman, A.
    Sharma, D.
    Panda, P. K.
    Jalali, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1166 - S1166
  • [43] Development of beam monitoring system for proton pencil beam scanning using fiber-optic radiation sensor
    Son, Jaeman
    Koo, Jihye
    Moon, Sunyoung
    Yoon, Myonggeun
    Jeong, Jonghwi
    Kim, Sun-Young
    Lim, Youngkyung
    Lee, Se Byeong
    Shin, Dongho
    Kim, Meyoung
    Kim, Dongwook
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2017, 71 (07) : 438 - 443
  • [44] Development of beam monitoring system for proton pencil beam scanning using fiber-optic radiation sensor
    Jaeman Son
    Jihye Koo
    Sunyoung Moon
    Myonggeun Yoon
    Jonghwi Jeong
    Sun-Young Kim
    Youngkyung Lim
    Se Byeong Lee
    Dongho Shin
    Meyoung Kim
    Dongwook Kim
    [J]. Journal of the Korean Physical Society, 2017, 71 : 438 - 443
  • [45] Simulation and experimental benchmarking of a proton pencil beam scanning nozzle model for development of MR-integrated proton therapy
    Oborn, Bradley M.
    Semioshkina, Ekaterina
    van der Kraaij, Erik
    Hoffmann, Aswin L.
    [J]. MEDICAL PHYSICS, 2024, : 6196 - 6205
  • [46] Time-resolved dose rate measurements in pencil beam scanning proton FLASH therapy with a fiber-coupled scintillator detector system
    Kanouta, Eleni
    Poulsen, Per Rugaard
    Kertzscher, Gustavo
    Sitarz, Mateusz Krzysztof
    Sorensen, Brita Singers
    Johansen, Jacob Graversen
    [J]. MEDICAL PHYSICS, 2023, 50 (04) : 2450 - 2462
  • [47] Single Imager Proton Radiography with a Pencil-Beam Scanning System
    Harms, J.
    Maloney, L.
    Lin, Y.
    Liu, T.
    Erickson, A.
    Zhang, R.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E659 - E660
  • [48] A simulator of proton pencil beam scanning delivery
    Poulsen, P. R.
    Nystrom, H.
    Skyt, P. S.
    Jensen, M. F.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S124 - S125
  • [49] Adaptation of a Clinical Proton Pencil Beam Scanning System for FLASH Experiments
    Erickson, D. P. J.
    Saini, J.
    Cao, N.
    Ford, E. C.
    Emery, R.
    Kranz, M.
    Goff, P. H.
    Meyer, J.
    Wong, T.
    Bloch, C.
    Stewart, R. D.
    Sandison, G. A.
    Morimoto, A.
    DeLonais-Dick, A.
    Shaver, B.
    Rengan, R.
    Zeng, J.
    Schwarz, M.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E664 - E664
  • [50] IBA PROTON PENCIL BEAM SCANNING : VALIDATION